• 沒有找到結果。

Clinical evaluation of the eBsensor hand-held blood glucose monitoring system

N/A
N/A
Protected

Academic year: 2021

Share "Clinical evaluation of the eBsensor hand-held blood glucose monitoring system"

Copied!
4
0
0

加載中.... (立即查看全文)

全文

(1)

Clinical evaluation of the eBsensor hand-held

blood glucose monitoring system

Chien-Chih Chen

a,b

, Der-Yan Tai

a

, Ping Ho

a

, Yiu-Kay Lai

b

, Jui-Jane Lin

a

, Simon J.T. Mao

c

,

Chia-Chi Lin

d

, Yu-Chen Hsieh

d

, Hsiang-Cheng Chiang

d

, Ken-Shwo Dai

c,d,

aWei-Gong Memorial Hospital, Miaoli, Taiwan

bThe Institude of Biotechnology, College of Life Science, National Tsing Hua University, Hsinchu, Taiwan cDepartment of Biological Science and Technology, National Chiao Tung University, Hsinchu, Taiwan

dVisgeneer Inc., Hsinchu, Taiwan

Received 11 September 2006; received in revised form 20 September 2006; accepted 21 September 2006 Available online 3 October 2006

Abstract

Background: Home glucose monitoring system is increasingly recognized as an important tool for glycemic control. We evaluated the clinical

performance of the eBsensor glucose monitoring system.

Methods: Fingertip capillary blood glucose concentrations from 282 subjects were measured using eBsensor glucose monitoring system and

compared against predicate devices and the Yellow Springs Instruments (YSI) 2300 blood glucose analyzer. Accuracy and precision of the

eBsensor glucose monitoring system were assessed using several methods. The comparative study between the eBsensor and 2 currently marketed

monitoring systems was performed.

Results: The 282 eBsensor readings covered a wide range from 2.6 to 24.4 mmol/l. Deming regression and Pearson correlation analyses showed a

linear relationship between the eBsensor readings and the YSI reference method (eBsensor = 0.9496 YSI + 0.4127 mmol/l; r = 0.98). Error Grid

analysis demonstrated that 100% of the eBsensor readings in clinically acceptable zones A and B. The CVs for the 6 lots of strips were within the

satisfactory interval (b6%). The comparative study showed that the eBsensor readings correlated well with the OneTouch Ultra values (r=0.97)

and the Glucocard II values (r = 0.97).

Conclusions: eBsensor is a reliable glucose monitoring system which provides high accurate and precise glucose readings over a wide range of

glucose concentrations.

© 2006 Elsevier B.V. All rights reserved.

Keywords: Diabetes; Glucose monitoring system; Clarke Error Grid analysis; Glucose meter; Strip; Self-monitoring; Accuracy; Precision; Electrochemical

1. Introduction

Diabetes is a chronic disease associated with serious health

complications including visual disability, kidney failure,

neurop-athy, and heart disease. The global number of people with diabetes

was estimated to be 171 million in 2000 and 366 million in 2030

[1]

. The total number of diabetes-associated deaths was estimated

to be 2.9 million worldwide in 2000

[2]

. Diabetes was ranked as

the 6th leading cause of death in the U.S. in 2003

[3]

. Previous

studies have shown that maintaining a normal or near-normal

glucose level significantly prevents or delays some health

complications for people with diabetes

[4–7]

. The development

of blood glucose monitoring system permitted diabetic patients to

monitor their glucose levels at home. This system was

consequently recommended as an important component of

routine care for diabetic patients to control their blood glucose

levels

[8

–10]

.

There are many types of blood glucose monitoring systems

currently available on the market. The clinical performance of

many of these glucose monitoring systems has been reported

[11–16]

. Recently, eBsensor, a new 10-second hand-held blood

glucose monitoring system (Visgeneer Inc., Hsinchu, Taiwan),

Clinica Chimica Acta 377 (2007) 170–173

www.elsevier.com/locate/clinchim

⁎ Corresponding author. Visgeneer Inc., 3F.-2, No.83, Sec. 1, Gongdao 5th Rd., Hsinchu City 300, Taiwan. Tel.: +886 3 516 0111x1000; fax: +886 3 516 0161.

E-mail address:president@visgeneer.com.tw(K.-S. Dai).

0009-8981/$ - see front matter © 2006 Elsevier B.V. All rights reserved. doi:10.1016/j.cca.2006.09.023

(2)

has been introduced into the market. The objective of the

present study was designed to evaluate the clinical performance

of the eBsensor glucose monitoring system.

2. Materials and methods

2.1. Subjects and blood glucose measurements

This study comprised 282 patients with diabetes, who were all in the morning fasting state, attending the outpatient clinic of Wei-Gong Memorial Hospital (Miaoli, Taiwan). Of these participants, fingertip capillary blood glucose concentrations were obtained using eBsensor blood glucose monitoring system and 2 currently marketed monitoring systems, the Glucocard II (Arkray) and the OneTouch Ultra (LifeScan). All measurements were performed in the Department of Laboratory Medicine of Wei-Gong Memorial Hospital by the same trained technician according to the manufacturer's instructions to avoid errors made by patients. Immediately after fingerstick measurement, venous blood sample from each of the diabetic patients was drawn by a nurse. Plasma glucose concentrations of these samples determined by the Yellow Springs Instruments (YSI) 2300 blood glucose analyzer were served as the reference values. The hematocrits of the participants ranged from 30–55% were within the operative specification of the eBsensor. To evaluate the precision of the eBsensor glucose monitoring system, 6ty measurements at each of the 6 clinically relevant blood glucose ranges (2.8–3.1, 4.9–5.3, 7.8–8.7, 13.2–14.2, 18.8–19.9, and 24.3–25.4 mmol/l) were performed for the 6 lots of test strips using 6 eBsensor meters. To produce the 6 different glucose concentrations for

each lot of strips, venous whole blood samples from healthy volunteers were spiked before analysis. Analyses for this part of the study were performed over 2 weeks. All study protocols were reviewed and approved by the ethical committee of Wei-Gong Memorial Hospital.

2.2. Data analysis

The clinical accuracy of eBsensor was assessed by comparing the eBsensor readings with the YSI reference values using Pearson correlation analysis, Deming regression analysis, Clarke Error Grid analysis [17,18], and the percentage of readings within a defined range of deviation from the reference values. Using Pearson correlation and Deming regression analysis, the eBsensor readings versus reference values and 2 currently marketed monitoring systems (comparative study) were performed. For Clarke Error Grid analysis, the x-axis and the y-axis were defined as the YSI reference values and the eBsensor readings, respectively. The resulting graphic display of the Clarke Error Grid analysis was divided into five zones: 1) zone A: clinically accurate; 2) zone B: error greater than ± 20% but would lead to benign or no treatment; 3) zone C: overcorrection of the true glucose value; 4) zone D: dangerous failure to detect and treat; 5) zone E: erroneous treatment. The percentage of readings within a defined deviation range (b5%, 5–10%, 10–15%, 15–20%, and N20%) of the reference values was analyzed for each of the glucose range (≤5.5 mmol/l, 5.6– 11.1 mmol/l, and≥11.2 mmol/l). Precision of the eBsensor glucose monitoring system was assessed using coefficients of variation (CVs) calculated from the 60 measurements at each of the 6 different glucose concentrations.

3. Results

The present study included 282 patients, among whom 119

(42.2%) were females and 163 (57.8%) were males. The age of

Fig. 1. Error Grid analysis of the eBsensor monitoring system. eBsensor readings were plotted against the YSI reference method.

Table 1

Number and percentage of eBsensor readings within a defined interval according to the different glucose concentration ranges

Percentage deviation from the reference value

Glycemic ranges b5% 5–10% 10–15% 15–20% N20% Total ≤5.5 mmol/L 20 (36.3%) 14 (25.5%) 15 (27.3%) 3 (5.5%) 3 (5.5%) 55 5.6–11.1 mmol/L 92 (50.5%) 49 (26.9%) 23 (12.6%) 14 (7.7%) 4 (2.2%) 182 ≥11.2 mmol/L 17 (37.8%) 16 (35.6%) 8 (17.8%) 4 (8.9%) 0 (0.0%) 45 Total 129 79 46 21 7 282 Data are number or (%).

Table 2

The CVs (%) for measurements in series for six different glucose concentration ranges (mmol/L) Lots of strips 2.8–3.1 4.9–5.3 7.8–8.7 13.2–14.2 18.8–19.9 24.3–25.4 Lot 1 4.32 3.81 4.25 3.52 3.4 3.44 Lot 2 5.56 4.89 3.35 3.34 3.83 3.93 Lot 3 4.74 4.4 4.32 3.51 2.62 3.14 Lot 4 5.24 4.0 3.86 3.42 3.43 4.6 Lot 5 5.84 5.11 4.86 5.92 4.47 5.26 Lot 6 4.17 5.43 5.02 4.37 3.77 3.12 171 C.-C. Chen et al. / Clinica Chimica Acta 377 (2007) 170–173

(3)

the participants was from 18 to 87 y. The hematocrits of these

participants were within the operative specification (30–55%) of

the eBsensor. The 282 eBsensor readings covered a wide range

of glucose concentrations from 2.6 to 24.4 mmol/l (55 readings

(19.5%)

≤5.5 mmol/l, 182 readings (64.5%) 5.6–11.1 mmol/l,

34 readings (12.1%) 11.2–16.6 mmol/l, 9 readings (3.2%) 16.7–

22.2 mmol/l, and 2 readings (0.7%) 22.3–27.8 mmol/l). Over the

range of glucose readings, Pearson correlation and Deming

regression analyses showed that eBsensor correlated well with

the YSI values with the equation of eBsensor = 0.9496 YSI

+ 0.4127 mmol/l and the 95% confidence intervals (CI) for the

slope and intercept were 0.9262 to 0.9731 and 0.2015 mmol/l to

0.6239 mmol/l, respectively. Pearson correlation analysis

showed that the correlation coefficient (r) was 0.98. Error Grid

analysis showed that 275 readings (97.5%) fell within zone A

and 7 readings (2.5%) fell within zone B (

Fig. 1

). All readings

falling on zone B were near the A/B border. There were no

readings in the C, D, or E zones.

Of the eBsensor readings, 97.5%, 90.1%, 73.8% and 45.7%

were within ± 20%, ± 15%, ± 10%, and ± 5%, respectively, of the

reference values. The total readings were divided into three

glucose ranges (≤ 5.5 mmol/l, 5.6–11.1 mmol/l, and

≥11.2 mmol/l) for further analysis (

Table 1

). In each of the

glucose ranges, the percentage of readings deviating from the

reference values within the 10% interval was found to be 61.8%

(

≤ 5.5 mmol/l), 77.4% (5.6–11.1 mmol/l), and 73.4%

(

≥11.2 mmol/l). The precision of the eBsensor glucose

monitoring system is shown in

Table 2

. Of the 6 test glucose

ranges (2.8–3.1 mmol/l, 4.9–5.3 mmol/l, 7.8–8.7 mmol/l,

13.2–14.2 mmol/l, 18.8–19.9 mmol/l, and 24.3–25.4 mmol/l),

CVs for lot 1 to lot 6 were ranged from 3.4% to 4.32%, from

3.34% to 5.56%, from 2.62% to 4.74%, from 3.42% to 5.24%,

from 4.47% to 5.92%, and from 3.12% to 5.43%. As compared

to the currently marketed glucose monitoring systems, eBsensor

correlated well with the OneTouch Ultra values

(eBsen-sor = 0.8512 OneTouch Ultra + 0.5787 mmol/l, and r = 0.97;

Fig. 2

a) and with the Glucocard II values (eBsensor = 0.8466

Glucocard II + 0.814 mmol/l, and r = 0.97;

Fig. 2

b).

4. Discussion

eBsensor, a hand-held glucose monitoring system, was

recently introduced into the self-monitoring diagnostic market.

Using electrochemical detection technique, eBsensor was

designed for rapid determination (10 s) of glucose concentrations

over a wide range of glucose values (1.7 to 33.3 mmol/l) from a

small amount (approximately 2.5

μl) of capillary whole blood

sample. In order to determine if the eBsensor is an acceptable

device for measuring blood glucose concentrations, the present

study was aimed to evaluate the clinical performance of the

eBsensor. The 282 eBsensor readings covered a wide range of

glucose concentrations (2.6 to 24.4 mmol/l) suggesting that this

study population was sufficient to represent a larger population of

diabetic patients. Pearson correlation and Deming regression

analyses showed that the eBsensor readings correlated excellently

with the YSI reference values over the range of glucose

concentrations measured. The Error Grid analysis showed that

eBsensor glucose monitoring system had 100% of measurements

in zones A and B suggesting that eBsensor is a clinically

acceptable device for measuring blood glucose concentrations.

The percentage of deviation from the reference value is

another relevant approach to determine the accuracy of the

glucose monitoring system. It was recommended by the

International Organization for Standardization (ISO) that 95%

of the measurements should be within ± 20% of the reference

values for glucose concentrations higher than 4.2 mmol/l and

within ± 0.83 mmol/l for glucose values under 4.2 mmol/l

[19]

.

Our results showed that 97.5% of the eBsensor readings were

within ± 20% interval indicating that eBsensor met the accuracy

criteria proposed by ISO. According to the recommendation of

the American Diabetes Association (ADA), the acceptable

range of deviation from laboratory reference values was limited

within ± 5%

[20]

. Although the clinical performance of the

glucose monitoring systems has been reported regularly

[11–

16]

, none of the glucose monitoring systems was shown to be

satisfied based on the ADA stringent criterion

[21–23]

. Another

analysis approach has been proposed to group the glucose

monitoring systems into good, acceptable, and unacceptable for

Fig. 2. Linear regression analysis of the eBsensor readings against the currently marketed monitoring systems. (a) OneTouch Ultra and (b) the Glucocard II.

(4)

clinical use

[24]

. The

“good” meters were defined as N60% of

the readings within ± 10% of the reference value. Thus,

eBsensor should be classified as a

“good” meter based on the

result that eBsensor not only had 73.8% of the overall readings

but also had 61.8%, 77.4%, and 73.4% of the readings in three

different glycemic ranges within ± 10% interval.

The CVs of the eBsensor at the 6 different glucose levels

were within the satisfactory interval (b6%) suggesting that

eBsensor monitoring system provided precise measurements.

Furthermore, 2 currently marketed glucose monitoring systems,

OneTouch Ultra and Glucocard II, were selected for

compar-ison. Correlation and regression analyses showed that the

eBsensor readings correlated well with the OneTouch Ultra and

the Glucocard II values over the range of glucose concentrations

measured. In conclusion, the present study demonstrates that

eBsensor is a reliable glucose monitoring system providing

high accurate and precise glucose readings and can be rated as

“good meter”.

Acknowledgements

We thank our patients for participating in this study. We also

like to thank Wei-Gong Memorial Hospital, National Tsing Hua

University, National Chiao Tung University, and Visgeneer Inc.,

who supported the present study.

References

[1] Wild S, Roglic G, Green A, Sicree R, King H. Global prevalence of diabetes: estimates for the year 2000 and projections for 2030. Diabetes Care 2004;27:1047–53.

[2] Roglic G, Unwin N, Bennett PH, et al. The burden of mortality attributable to diabetes: realistic estimates for the year 2000. Diabetes Care 2005;28:2130–5.

[3] Hoyert DL, Heron MP, Murphy SL, Kung HC. Deaths: final data for 2003. Natl Vital Stat Rep 2006;54:1–120.

[4] DCCT Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86.

[5] UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). Lancet 1998;352:837–53.

[6] American Diabetes Association:Standards of medical care in diabetes (Position Statement). Diabetes Care 2004;27(Suppl 1):S15–35. [7] World Health Organization. The Global Strategy on Diet, Physical Activity

and Health:Fact Sheet– Diabetes; 2003.

[8] Sonksen PH, Judd SL, Lowy C. Home monitoring of blood-glucose. Method for improving diabetic control. Lancet 1978;1:729–32. [9] Ikeda Y, Tsuruoka A. Self-monitoring of blood glucose, as a means of

self-management. Diabetes Res Clin Pract 1994:S269–71 [Suppl].

[10] Franciosi M, Pellegrini F, De Berardis G, et al. The impact of blood glucose self-monitoring on metabolic control and quality of life in type 2 diabetic patients: an urgent need for better educational strategies. Diabetes Care 2001;24:1870–7.

[11] Tate PF, Clements CA, Walters JE. Accuracy of home blood glucose monitors. Diabetes Care 1992;15:536–8.

[12] Trajanoski Z, Brunner GA, Gfrerer RJ, Wach P, Pieber TR. Accuracy of home blood glucose meters during hypoglycemia. Diabetes Care 1996;19:1412–5.

[13] Rheney CC, Kirk JK. Performance of three blood glucose meters. Ann Pharmacother 2000;34:317–21.

[14] Alto WA, Meyer D, Schneid J, Bryson P, Kindig J. Assuring the accuracy of home glucose monitoring. J Am Board Fam Pract 2002;15:1–6. [15] Dai KS, Tai DY, Ho P, et al. Accuracy of the EasyTouch blood glucose

self-monitoring system: a study of 516 cases. Clin Chim Acta 2004;349:135–41.

[16] Cohen M, Boyle E, Delaney C, Shaw J. A comparison of blood glucose meters in Australia. Diabetes Res Clin Pract 2006;71(2):113–8. [17] Cox DJ, Clarke WL, Gonder-Frederick L, et al. Accuracy of perceiving

blood glucose in IDDM. Diabetes Care 1985;8:529–36.

[18] Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987;10:622–8.

[19] ISO/CD 15197: Determination of Performance Criteria for In Vitro Blood Glucose Monitoring Systems for Management of Human Diabetes Mellitus. Geneva: ISO Central Secretariat; 2001.

[20] American Diabetes Association. Self-monitoring of blood glucose (consensus statement). Diabetes Care 1996;19(Suppl 1):S62–6. [21] Brunner GA, Ellmerer M, Sendlhofer G, et al. Validation of home blood

glucose meters with respect to clinical and analytical approaches. Diabetes Care 1998;21:585–90.

[22] Weitgasser R, Gappmayer B, Pichler M. Newer portable glucose meters-analytical improvement compared with previous generation devices? Clin Chem 1999;45:1821–5.

[23] Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436–72. [24] Poirier JY, Le Prieur N, Campion L, Guilhem I, Allannic H, Maugendre D.

Clinical and statistical evaluation of self-monitoring blood glucose meters. Diabetes Care 1998;21:1919–24.

173 C.-C. Chen et al. / Clinica Chimica Acta 377 (2007) 170–173

數據

Fig. 1. Error Grid analysis of the eBsensor monitoring system. eBsensor readings were plotted against the YSI reference method.
Fig. 2 a) and with the Glucocard II values (eBsensor = 0.8466

參考文獻

相關文件

Wang, Solving pseudomonotone variational inequalities and pseudocon- vex optimization problems using the projection neural network, IEEE Transactions on Neural Networks 17

Define instead the imaginary.. potential, magnetic field, lattice…) Dirac-BdG Hamiltonian:. with small, and matrix

The study explore the relation between ownership structure, board characteristics and financial distress by Logistic regression analysis.. Overall, this paper

Microphone and 600 ohm line conduits shall be mechanically and electrically connected to receptacle boxes and electrically grounded to the audio system ground point.. Lines in

Employee monitoring involves the use of computers to observe, record, and review an employee’s use of a computer. Employee monitoring involves the use of computers to

By using the case study and cross analysis of the results, The Purpose of this research is find out the Business implementing Supply Chain Management system project, Our study

The Performance Evaluation for Horizontal, Vertical and Hybrid Schema in Database Systems.. -A Case Study of Wireless Broadband

This study aimed to establish the strength models of High-Performance Concrete (HPC) using Nonlinear Regression Analysis (NLRA), Back-Propagation Networks (BPN) and