Medical Engineering & Physics 30 (2008) 1004–1012

In vivo three-dimensional kinematics of the normal knee during

active extension under unloaded and loaded conditions

using single-plane fluoroscopy

Tung-Wu Lu

a,∗

, Tsung-Yuan Tsai

a

, Mei-Ying Kuo

a,b

,

Horng-Chaung Hsu

c

, Hao-Ling Chen

d

aInstitute of Biomedical Engineering, National Taiwan University, No. 1, Section 1, Jen-Ai Road, Taipei 100, Taiwan, ROC bDepartment of Physical Therapy, China Medical University, Taiwan, ROC

cDepartment of Orthopaedic Surgery, China Medical University Hospital, Taichung, Taiwan, ROC dSchool of Occupational Therapy, National Taiwan University, Taiwan, ROC

Received 30 June 2007; received in revised form 27 January 2008; accepted 3 March 2008

Abstract

Measurement of the changes of the three-dimensional (3D) motion and surface kinematics of the knee under different external loading conditions is essential for the understanding and evaluation of the function of the joint, as well as for relevant clinical applications. Knee extension exercise (KEE) has been applied extensively in the rehabilitation programs of patients with various knee disorders. This study measured the 3D knee kinematics of eight normal subjects during active knee extension for unloaded and loaded conditions using a voxel-based method for the registration of fluoroscopic images with CT bone data. The knee kinematics during unloaded conditions were found to be similar to previous findings. A mass of 5 kg at the ankle did not affect the joint angles but significantly altered the lateral contact positions during knee extension, especially at knee flexion angles higher than 75◦, and also reduced the asymmetry of the surface kinematics between the medial and lateral condyles. The results of the current study may be useful for knee replacement design and for developing guidelines for the use of KEEs for the rehabilitation of patients with knee disorders.

© 2008 IPEM. Published by Elsevier Ltd. All rights reserved.

Keywords: Knee; Kinematics; Fluoroscopy; Registration; Joint contact

1. Introduction

Measurement of the three-dimensional (3D) motion and surface kinematics of the knee during movement is essen-tial for the understanding and evaluation of the function of the joint, as well as for relevant clinical purposes, such as the design and implantation of replacement joints[1–3], and the design of a suitable rehabilitation program for par-ticular injuries [4,5]. Knee extension exercise (KEE), a non-weight-bearing, open kinetic chain exercise, has been applied extensively in the rehabilitation programs of patients

with various knee disorders, such as knee OA [6] and

patellofemoral pain syndrome [7], for patients following

Corresponding author. Tel.: +886 2 33653335; fax: +886 2 33653335.

E-mail address:twlu@ntu.edu.tw(T.-W. Lu).

anterior cruciate ligament (ACL) reconstruction[8], and for elderly people[9]to increase quadriceps muscle strength and improve physical function. However, some studies discour-aged the use of KEE in certain conditions such as during the first year following ACL injury and reconstruction (e.g.[10]). It seems that the correct inclusion of KEE in rehabilitation programs requires a better understanding of the mechanics of the knee and effects of external loading during this exercise. Providing a detailed description of the mechanics of the knee during KEE can be challenging because it is difficult to assess directly the forces transmitted in the force-bearing structures in vivo, including the articular surfaces, ligaments and muscles[11–15]. However, since the kinematics of the knee are a result of the complex interactions between the force-bearing structures, changes in the knee kinematics dur-ing KEE under different external loaddur-ing conditions may be

1350-4533/$ – see front matter © 2008 IPEM. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.medengphy.2008.03.001

(2)

used to examine the overall response of these internal struc-tures. An in vitro study has revealed the role of the ligaments and articular surfaces in determining the passive kinemat-ics of the knee[16]. However, the effects of muscle activity in response to external loads on the knee kinematics during movement have not been studied extensively. This is because these conditions are difficult to simulate in vitro[17–19], and the accurate measurement of the in vivo 3D joint kinematics during dynamic movements is not straightforward[20–24]. To the best knowledge of the authors, no data have been reported on the effects of load on the in vivo 3D knee motion and surface kinematics during KEE.

Skin maker-based stereophotogrammetry has been widely used for the measurement of the in vivo 3D kinematics of human joints. However, errors due to undesired motion between markers and the underlying bones, called skin move-ment artifacts, are difficult to eliminate. The use of skin marker-based methods to describe joint surface kinemat-ics during motion is also difficult because of the lack of joint surface information. Methods based on the registration of two-dimensional (2D) fluoroscopic images and computer models of the components of total knee replacements (TKRs) have recently been developed for accurate 3D kinematic anal-ysis of replaced knee joints[25,26]. However, the application of this method to natural knee kinematics has been limited

[27,28]as human bones are of weaker contrast, and are thus more difficult to identify in fluoroscopic images, than metal-lic replacement components, leading to reduced accuracy. A voxel-based 2D to 3D registration method for more accu-rate measurement of the natural knee kinematics[29]can be a useful tool for the study of in vivo 3D knee motion and surface kinematics during KEE.

The purposes of the study were to measure the in vivo 3D motion and surface kinematics of the normal knee during KEE in unloaded and loaded conditions using single-plane fluoroscopy with a voxel-based 2D to 3D registration method. It was hypothesized that the knee kinematics in unloaded and loaded conditions during KEE would be different. It is hoped that these accurate and detailed descriptions of the knee kinematics during this exercise will be helpful for future clinical applications.

2. Materials and methods 2.1. Subjects

Eight normal subjects (age: 20.7± 0.9 years; mass:

64.8± 5.5 kg; height: 168.3 ± 6.8 cm), who were free from any disease of the neuromuskuloskeletal system, volunteered to participate in this study. Each subject gave his/her writ-ten informed consent before the experiment, clearance of which was carried out and approved by the Institutional Ethics Committee.

Each subject received a computed tomography (CT, PQ-5000, Picker International) scan over a range from 15 cm

superior to the knee joint line to 15 cm inferior, with a slice thickness of 1 mm and a pixel size of 0.625 mm× 0.625 mm. The CT data were segmented by applying a threshold filter to the slices, and the external soft tissues were then digitally removed. The remaining CT data of the femur, tibia, fibula and patella were then used to obtain volumetric models of the individual bones. These subject-specific volumetric models preserved the scanned details of the bones, including their external surfaces and internal structures, and were necessary for subsequent registration with fluoroscopic images. With several million voxels, volumetric models are expected to describe the bones, and their positions and orientations at an accuracy higher than the voxel size that is the best point accuracy in CT[30].

2.2. Experimental procedure

Each subject sat on a chair and performed isolated active knee extension tests in unloaded and loaded conditions under dynamic fluoroscopic (Angiography, Advantx LCA, GE, France) surveillance,Fig. 1. Subjects were asked to extend their right knees from a fully flexed to a fully extended posi-tion at an average speed of about 70◦/s with the assistance of a metronome. In the loaded tests, a 5-kg mass was fixed to the ankle to impose a maximum moment of about 30 Nm at the knee, which is a commonly prescribed rehabilitation activity aimed at training of the quadriceps[13]. The fluoroscopic video was converted to digital images at a sampling rate of 30 frame/s using an image grabber (PCI bus frame grabber, Foresight, USA).

2.3. Fluoroscopy correction and calibration

For the registration of the fluoroscopic images with the CT bone models, the fluoroscopy system was represented

Fig. 1. Schematic diagram showing a subject performing active knee exten-sion under dynamic fluoroscopic surveillance. An extra 5 kg mass would be fixed on the ankle for the loaded test conditions.

(3)

by a perspective projection image formation model, with a point X-ray source and an image plane, i.e., the pla-nar phosphor. All the fluoroscopic images were corrected for distortions through a calibration procedure based on

a modified polynomial method [31] using images of a

reseau plate (21× 21 lead markers at a grid interval of 25 mm) placed on the intensifier tube. The mean (standard deviation) of the residual distortion errors after calibra-tion was 0.07 (0.03) mm. The posicalibra-tion of the X-ray focus was determined by minimizing the sum of the projection errors of 160 lead markers distributed in a star shape on a calibration plate (marker-to-marker distance = 12.5 mm) positioned 30 cm away from the reseau plate. The mean (standard deviation) of the projection errors was 0.13 (0.05) mm.

2.4. Registration procedure

With the perspective projection model of the fluoroscopy system, the registration of the voxel-based CT-to-fluoroscopy images was performed by comparing systematically the flu-oroscopic image with simulated 2D synthetic X-ray images of the bone model called a digitally reconstructed radiograph

(DRR)[32], Fig. 2. A DRR was generated by casting

X-rays through the volumetric CT data of a bone model and projecting onto the image plane to form an image resem-bling a radiograph with detailed internal information of the bony structures [33]. The 3D poses of the bones at each image frame were obtained by searching for the pose of the bone model using an optimization procedure to produce the DRR which best matched the fluoroscopic image. A computer

Fig. 2. The computer graphics user interface (GUI) program to assist the visualization of the registration and reconstruction of the 3D bone kinematics. The fluoroscopy image can be reconstructed and, with the imported CT data of the bone(s), digitally reconstructed radiographs (DRRs) are generated on the image plane for automatic registration through an optimization procedure. A surface rendered registered knee joint is shown from (a) the front and (b) the oblique view, and the corresponding DRRs are given in (c) and (d).

(4)

Fig. 3. The mean internal/external rotation (I/E) and abduction/adduction (Abd/Add) angles, and posterior/anterior (P/A) and medial/lateral (M/L) translations with respect to the knee flexion angle during (a and b) unloaded and (c and d) loaded conditions. Shaded areas indicate one standard deviation.

graphics user interface (GUI) was developed to assist with the visualization of the registration and reconstruction of the 3D bone kinematics (Fig. 3). The registration method had previ-ously been evaluated for its accuracy using a cadaveric knee, giving means and standard deviations of the knee kinematics errors of 0.24± 0.77 mm, 0.41 ± 3.06 mm and 0.59 ± 1.13◦ for in-plane translation, out-of-plane translation, and all rota-tions, respectively, over 11 flexion positions ranging from 0◦ to 100◦[29].

2.5. Data analysis

The registered poses of the femur and tibia were used to calculate the 3D kinematics of the knee, including joint rotations and translations, as well as contact points and slip ratios of the articular surfaces, described as follows.

Joint angles were calculated as the relative rotations between the local coordinate systems of the femur and tibia. Since the CT scans of the bones were performed at full knee extension, the local coordinate systems of both bones were taken to be coincident with each other and originated at the most posterior point of the tibial attachment of the ACL. This definition indicated that the knee joint had zero rotational angles and translations at the fully extended position, which matched clinical definitions. The positive Z-axis was defined to be parallel to the inter-epicondylar axis and directed to the right. The positive X-axis was normal to the plane defined by the Z-axis and the longitudinal axis of the femur and directed anteriorly. The positive Y-axis was orthogonal to both the

Z- and X-axes, following the right-hand rule, and directed

superiorly. The rotational angles were then obtained using a cardanic sequence of Z–Y–X [16]. The translations were defined as the displacements between the origins of the two local coordinate systems and were expressed relative to the tibial coordinate system.

Contact points of the medial and lateral femoral condyles on the corresponding compartments of the tibial plateau and the slip ratios between the articular surfaces were used to describe the surface kinematics of the tibiofemoral joint. For the description of the contact patterns, the origin of the tibial coordinate system was set at the center of an ellipse that best fitted the tibial plateau using points on the articulating sur-faces. Contact points on one articular surface were estimated as the centroid of the contact area defined by the points that were within a distance of 2 mm from the corresponding points on the other articular surface. This may be considered as a first approximation of possible cartilage deformations, and the resulting contact point can be related to the center of the contact pressure. Previous studies have used the lowest point of the femur, assuming a flat tibial plateau, and the point with minimal distance between the articular surfaces to determine the contact points [28,34,35]. Both methods assume point contact between the articular surfaces, which may be sensi-tive to rough surfaces associated with the noise from the CT data.

The slip ratio was defined as the ratio of the length of the femoral articulation between two successive contact points and the length of the tibial articulation between the

(5)

corre-Table 1

The means (standard deviations) of the total displacements of the medial and lateral contact points in the anterior–posterior and medial–lateral directions during unloaded and loaded conditions

Anterior–posterior Medial–lateral

Unloaded (mm) Loaded (mm) p-Value Unloaded (mm) Loaded (mm) p-Value

Medial 11.2 (4.0) 11.5 (3.9) 0.890 5.0 (3.4) 5.8 (2.4) 0.637

Lateral 16.5 (2.1) 12.5 (3.3) 0.034 6.6 (1.8) 4.9 (1.2) 0.079

sponding contact points[36]. A slip ratio of 1 indicates pure rolling and 0 indicates pure sliding. The slip ratios were calculated at 15◦ intervals, up to 120◦ knee flexion. Inter-polation of the contact points was used when the data points did not correspond to the intervals. To quantify the bilat-eral asymmetry of medial and latbilat-eral condylar slip ratios, a symmetry index (SI) was used and defined as follows

[37]:

SI (%)= (medial slip ratio− lateral slip ratio) (medial slip ratio+ lateral slip ratio)/2× 100 An SI value of 0 indicates perfect bilateral symmetry.

2.6. Statistical analysis

For all calculated variables, paired t-tests were performed to test for differences between conditions using SPSS Version 11.0 (SPSS Inc., Chicago, USA). The significance level was set atα = 0.05.

3. Results

No statistically significant differences in the aver-age speeds were found between the two test conditions (p = 0.5367). During unloaded active knee extension, the tibia rotated externally by about 8.8◦from 114◦to 30◦knee flex-ion, and by about 10◦ during the terminal 30◦ of flexion (Fig. 3a). The knee remained at about 5◦adduction before returning to neutral during the terminal 30◦ (Fig. 3a). The reference point on the femur translated relative to the tibia by a mean of 29.6 mm anteriorly, 19.9 mm distally, and 4.9 mm laterally as the knee extended (Fig. 3b). Under the loaded condition, the angular and translational motion curves were statistically the same as those during the unloaded condition, although the mean joint rotation curves appear to be different for flexion angles greater than 75◦(Fig. 3c and d).

Contact points on the medial and lateral tibial plateau moved anteriorly during knee extension during both load-ing conditions (Fig. 4a and b). The total anterior–posterior

Fig. 4. The mean positions of the contact points on the tibial plateau under (a) unloaded and (b) loaded conditions at intervals of 30◦, up to 120◦knee flexion. Comparisons of the contact positions between unloaded (thin, square markers) and loaded (thick, circular markers) conditions are shown in (c) for the medial and (d) for the lateral condyle. Error bars indicate one standard deviation (*p-value < 0.05).

(6)

Fig. 5. Slip ratios of the medial (thick, square markers) and lateral (thin, circular markers) condyles during (a) unloaded and (b) loaded conditions. Error bars indicate one standard deviation (*p-value < 0.05).

and medial–lateral displacements of the contacts in both tib-ial compartments were not significantly different between loading conditions, except for the total anterior–posterior dis-placements of the lateral contact during the loaded condition being significantly smaller (Table 1,Fig. 4a and b). At 75◦, 90◦, 105◦and 120◦of knee flexion, the lateral contact points under the loaded condition were more anterior than during the unloaded condition (Fig. 4d), while the medial contact points were not significantly different between loading conditions (Fig. 4c).

Slip ratios for both medial and lateral condyles at 15◦ intervals are shown in Fig. 5. During unloaded extension, the slip ratios of the medial compartment were all sig-nificantly greater than those of the lateral, except for the 105◦–120◦interval (Fig. 5a). The slip ratios of the medial compartment were less than 4 for flexion angles greater

than 60◦ and increased to a maximum mean value of 7.6

for flexion angles less than 60◦, while the corresponding slip ratios for the lateral compartment were 2.5 and 4.9, respectively (Fig. 5a). Under the loaded condition, the slip ratios of the medial compartment were not significantly changed, but those of the lateral side were increased through-out the whole range of extension and became statistically the same as those of the medial side. This change was also reflected in the SI values of the medial and lateral slip ratios. Under the loaded condition, the asymmetry of the slip ratios between the two compartments was largely reduced, with significance in the range between about 40◦ and 80◦ (Fig. 6).

4. Discussion

During unloaded knee extension, the tibia rotated exter-nally with decreasing flexion angle, and about half of the total range of external rotation (about 10◦) occurred during the terminal 30◦(Fig. 3a), corresponding to the so-called screw-home mechanism[16,18,38,39]. Patterns of the angular and translational displacements of the knee were also similar to

those during passive motion reported in a cadaveric study

[16]. A longer path for the lateral contact on the convex tibial plateau was found in the current study (Table 1), in agreement with previous reports[28,34]. The longer path contributed to the tibial external rotation during knee extension but, apart from articular surfaces, ligaments also play an essential role in guiding the motion of the femur relative to the tibia as shown by Wilson et al.[16]. Although contact movement paths pro-vided some information about the surface kinematics of the joint, the slip ratio, proposed by O’Connor et al.[36], com-bined the information of the contacts on the tibial and femoral articular surfaces and provided more insight into the relative motion between these articular surfaces. During active knee extension, the slip ratios varied with flexion angle, and those of the medial condyles were generally greater than those of the lateral, but nowhere is the slip ratio equal to unity, corre-sponding to pure rolling. With a 4-bar linkage model of the knee, O’Connor et al.[36]reproduced the backward rolling and forward sliding motion of the femur on the tibia during passive knee flexion, and the slip ratio was used to quantify the relative motion of the femoral and tibial articulations. The

Fig. 6. SI of the medal and lateral slip ratios against knee flexion angles between unloaded (thin, square markers) and loaded (thick, circular markers) conditions (*p-value < 0.05).

(7)

slip ratios obtained in the current study were in agreement with those reported by O’Connor et al., except that higher values were found for low flexion angles in the current study. This may have to do with the involvement of muscles in the current study.

During loaded knee extension, the patterns of the joint angles and translations were statistically similar to those dur-ing the unloaded condition, while the contact positions in the lateral compartment were significantly different for flex-ion angles higher than 75◦ (Fig. 4d). These results appear to suggest different effects of load on the joint kinematics. However, since the bone-embedded coordinate systems were consistently attached to bones for all the subjects, the results may be due to the fact that the surface kinematics are deter-mined by both the rigid-body kinematics and the articular surface geometry. Under the constraints of the articular sur-faces, changes of contact positions under loading appeared to be more consistent between subjects than the rigid-body kinematics.

The change in the lateral contact reduced the total dis-placement of the contact path (Table 1) and led to the increase of the lateral slip ratios (Fig. 5b), which further decreased the asymmetry of the slip ratios between the two compart-ments (Fig. 6). Between 75◦and 90◦flexion, as the femur was horizontal and fixed relative to the chair, the lower leg and the weight were in front of the knee joint, generating an extensor moment at the knee. Since the patellar tendon is directed posteriorly at these angles [40], it pulled the tibia posteriorly while balancing the external moment. At flexion angles greater than 90◦, the weight and the lower leg would cause a flexor moment at the knee, requiring the contractions of the hamstrings, which would also pull the tibia posteri-orly. These posterior shear forces in response to the external weight would lead to the more anterior movement of the con-tact points on the tibial plateau during the loaded condition. The reason why the contact position changes occurred in the lateral compartment may be related to the net effects of the muscle and ligament forces on the external tibial rotation, including the patellar tendon and posterior cruciate ligament (PCL) forces. The shape differences of the two tibial plateaus also contributed to the observed phenomenon. While the ante-rior displacement of the contact in the anteante-rior part of the concave medial plateau will increase the articular resistance to posterior shear forces and reduce the loads in the PCL, that in the posterior part of the convex lateral plateau will have the opposite effects[41].

The anterior shift of the contact points may lead to a greater load in the PCL or lateral collateral ligament because these two ligaments provide the posterior and lateral stabil-ity of the knee joint. These reasons suggest that KEEs at higher flexion angles may not help in protecting an injured PCL or a PCL graft, but they may be safe for training in ACL-deficient patients. External loads appeared to reduce the relative motions of the femoral and tibial articular sur-faces between medial and lateral compartments, as revealed by the reduced SI of slip ratios. These changes in joint surface

kinematics may be relevant to the rehabilitation of patients with knee OA. The 3D in vivo data on the knee motion and surface kinematics in the current study may also be helpful for theoretical estimation of the forces transmitted in the knee joint during active KEEs for the rehabilita-tion of patients with knee injuries [11]. These data may also provide a useful reference for future knee replacement design.

The voxel-based method used in the current study for the registration of the fluoroscopic images with the CT bone mod-els was shown to be highly accurate, with sub-millimeter accuracy in in-plane (i.e., anterior–posterior) translations. Therefore, the changes on the surface kinematics could be obtained. Although not as accurate as “quasi-static” mea-surements using roentgen stereophotogrammetric analysis (RSA), CT or MRI, the method has the advantage of mea-suring more natural knee motions in 3D, dynamic, and weight-bearing conditions. Further study may include inves-tigation of the effects of higher loads on the knee kinematics, and the estimation of the forces transmitted by the force-bearing structures by combining the accurate in vivo knee kinematic data with an appropriate knee model. A more advanced fluoroscopy system may be needed for the study of faster knee motions.

5. Conclusions

In vivo 3D knee motion and surface kinematics during

active knee extension under unloaded and loaded conditions were measured using single-plane fluoroscopy with CT bone models. The knee kinematics during the unloaded condition were similar to previous findings. A weight at the ankle did not affect the joint angles but did significantly alter the lateral contact positions during knee extension, especially at knee flexion angles higher than 75◦, and also reduced the asymme-try of the surface kinematics between the medial and lateral condyles. The results may be useful for knee replacement design and for developing guidelines for the use of KEEs for rehabilitation purposes.

Acknowledgments

The authors gratefully acknowledge the financial support from the National Science Council of Taiwan (NSC-94-2213-E-002-112). We are also grateful for the staff of the Medical Imaging Department at China Medical University.

Conflict of interest

We, Tung-Wu Lu, Tsung-Yuan Tsai, Mei-Ying Kuo, Horng-Chaung Hsu and Hao-Ling Chen, declare that we have no proprietary, financial, professional or other personal interest of any nature or kind in any product, service and/or

(8)

company that could be construed as influencing the posi-tion presented in, or the view of, the manuscript entitled, “In

vivo three-dimensional kinematics of the normal knee during

active extension under unloaded and loaded conditions using single-plane fluoroscopy.”

References

[1] Coughlin KM, Incavo SJ, Doohen RR, Gamada K, Banks S, Beynnon BD. Kneeling kinematics after total knee arthroplasty: anterior–posterior contact position of a standard and a high-flex tibial insert design. J Arthroplasty 2007;22:160–5.

[2] Muratoglu OK, Rubash HE, Bragdon CR, Burroughs BR, Huang A, Harris WH. Simulated normal gait wear testing of a highly cross-linked polyethylene tibial insert. J Arthroplasty 2007;22:435–44.

[3] Kessler O, Durselen L, Banks S, Mannel H, Marin F. Sagittal curvature of total knee replacements predicts in vivo kinematics. Clin Biomech 2007;22:52–8.

[4] Schmalz T, Blumentritt S, Wagner R, Junge R. Evaluation with biome-chanical gait analysis of various treatment methods after rupture of the anterior cruciate ligament. Sportverletz Sportschaden 1998;12: 131–7.

[5] DeVita P, Hortobagyi T, Barrier J. Gait biomechanics are not normal after anterior cruciate ligament reconstruction and accelerated rehabil-itation. Med Sci Sports Exerc 1998;30:1481–8.

[6] Huang M-H, Lin Y-S, Yang R-C, Lee C-L. A comparison of various therapeutic exercises on the functional status of patients with knee osteoarthritis. Semin Arthritis Rheum 2003;32:398–406.

[7] Witvrouw E, Danneels L, Van Tiggelen D, Willems TM, Cam-bier D. Open versus closed kinetic chain exercises in patellofemoral pain: a 5-year prospective randomized study. Am J Sports Med 2004;32:1122–30.

[8] Snyder-Mackler L, Delitto A, Bailey SL, Stralka SW. Strength of the quadriceps femoris muscle and functional recovery after reconstruction of the anterior cruciate ligament. A prospective, randomized clinical trial of electrical stimulation. J Bone Joint Surg Am 1995;77:1166– 73.

[9] Ferri A, Scaglioni G, Pousson M, Capodaglio P, Van Hoecke J, Narici MV. Strength and power changes of the human plantar flexors and knee extensors in response to resistance training in old age. Acta Physiol Scand 2003;177:69–78.

[10] Bynum EB, Barrack RL, Alexander AH. Open versus closed chain kinetic exercises after anterior cruciate ligament reconstruction. A prospective randomized study. Am J Sports Med 1995;23:401–6 [see comment].

[11] Toutoungi DE, Lu TW, Leardini A, Catani F, O’Connor JJ. Cruciate ligament forces in the human knee during rehabilitation exercises. Clin Biomech 2000;15:176–87.

[12] Lu TW, O’Connor JJ, Taylor SJ, Walker PS. Validation of a lower limb model with in vivo femoral forces telemetered from two subjects. J Biomech 1998;31:63–9.

[13] Fleming BC, Beynnon BD, Renstrom PA, Peura GD, Nichols CE, John-son RJ. The strain behavior of the anterior cruciate ligament during bicycling. An in vivo study. Am J Sports Med 1998;26:109–18. [14] Komi PV. Relevance of in vivo force measurements to human

biome-chanics. J Biomech 1990;23(Suppl. 1):23–34.

[15] An KN, Berglund L, Cooney WP, Chao EY, Kovacevic N. Direct in

vivo tendon force measurement system. J Biomech 1990;23:1269–71.

[16] Wilson DR, Feikes JD, Zavatsky AB, O’Connor JJ. The components of passive knee movement are coupled to flexion angle. J Biomech 2000;33:465–73.

[17] Torzilli PA, Deng X, Warren RF. The effect of joint-compressive load and quadriceps muscle force on knee motion in the intact and anterior cruciate ligament-sectioned knee. Am J Sports Med 1994;22:105–12.

[18] Shoemaker SC, Adams D, Daniel DM, Woo SL. Quadriceps/anterior cruciate graft interaction. An in vitro study of joint kinematics and anterior cruciate ligament graft tension. Clin Orthop 1993:379– 90.

[19] More RC, Karras BT, Neiman R, Fritschy D, Woo SL, Daniel DM. Hamstrings—an anterior cruciate ligament protagonist. An in vitro study. Am J Sports Med 1993;21:231–7.

[20] Bergmann G, Graichen F, Rohlmann A. Hip joint loading during walk-ing and runnwalk-ing, measured in two patients. J Biomech 1993;26:969–90. [21] Davy DT, Kotzar GM, Brown RH, Heiple KG, Goldberg VM, Heiple Jr KG, et al. Telemetric force measurements across the hip after total arthroplasty. J Bone Joint Surg Am 1988;70:45–50.

[22] English TA, Kilvington M. In vivo records of hip loads using a femoral implant with telemetric output (a preliminary report). J Biomed Eng 1979;1:111–5.

[23] Rydell NW. Forces acting on the femoral head-prosthesis. A study on strain gauge supplied prostheses in living persons. Acta Orthop Scand 1966;37(Suppl. 88):1–132.

[24] Taylor ME, Tanner KE, Freeman MER, Yettram AL. Stress and strain distribution within the intact femur: compression or bending? Med Eng Phys 1996;18:122–31.

[25] Banks SA, Hodge WA. Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng 1996;43:638–49.

[26] Mahfouz MR, Hoff WA, Komistek RD, Dennis DA. A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 2003;22:1561–74.

[27] Komistek RD, Dennis DA, Mahfouz M. In vivo fluoroscopic analysis of the normal human knee. Clin Orthop 2003;410:69–81.

[28] Dennis DA, Mahfouz MR, Komistek RD, Hoff WA. In vivo determina-tion of normal and anterior cruciate ligament-deficient knee kinematics. J Biomech 2005;38:241–53.

[29] Tsai T-Y, Lu T-W, Kuo M-Y, Hsu H-C, Kuo C-C. A voxel-based 2-D/3-D registration method for measuring natural knee kinematics with single-plane fluoroscopy. IEEE Trans Med Imaging; submitted for pub-lication.

[30] Zannoni C, Zannoni C, Viceconti M, Pierotti L, Cappello AACA. Eval-uation of CT accuracy in orthopaedic implants geometry reconstruction. In: Proceedings of the Engineering in Medicine and Biology Society, 1997 Proceedings of the 19th Annual International Conference of the IEEE. 1997.

[31] Baltzopoulos V. A video fluoroscopy method for optical distortion correction and measurement of knee-joint kinematics. Clin Biomech 1995;10:85–92.

[32] Weese J, Penney GP, Desmedt P, Buzug TM, Hill DL, Hawkes DJ, et al. Voxel-based 2-D/3-D registration of fluoroscopy images and CT scans for image-guided surgery. IEEE Trans Inf Technol Biomed 1997;1:284–93.

[33] Penney GP, Weese J, Little JA, Desmedt P, Hill DL, Hawkes DJ, et al. A comparison of similarity measures for use in 2-D–3-D medical image registration. IEEE Trans Med Imaging 1998;17:586–95.

[34] Li G, DeFrate LE, Park SE, Gill TJ, Rubash HE. In vivo articular cartilage contact kinematics of the knee: an investigation using dual-orthogonal fluoroscopy and magnetic resonance image-based computer models. Am J Sports Med 2005;33:102–7.

[35] Li G, Moses JM, Papannagari R, Pathare NP, DeFrate LE, Gill TJ. Anterior cruciate ligament deficiency alters the in vivo motion of the tibiofemoral cartilage contact points in both the anteroposterior and mediolateral directions. J Bone Joint Surg Am 2006;88:1826–34. [36] O’Connor J, Shercliff T, FitzPatrick D, Bradley J, Daniel D, Biden E,

et al. Geometry of the knee. In: Daniel D, Akeson WH, O’Connor JJ, editors. Knee ligaments: structure, function, injury, and repair. Raven Press; 1990. p. 163–200.

[37] Lu T-W, Lin H-C, Hsu H-C. Influence of functional bracing on the kinet-ics of anterior cruciate ligament-injured knees during level walking. Clin Biomech 2006;21:517–24.

(9)

[38] Markolf KL, Mensch JS, Amstutz HC. Stiffness and laxity of the knee—the contributions of the supporting structures. A quantitative

in vitro study. J Bone Joint Surg Am 1976;58:583–94.

[39] Trent PS, Walker PS, Wolf B. Ligament length patterns, strength, and rotational axes of the knee joint. Clin Orthop 1976;117:263– 70.

[40] Lu TW, O’Connor JJ. Lines of action and moment arms of the major force-bearing structures crossing the human knee joint: comparison between theory and experiment. J Anat 1996;189:575–85.

[41] Chan SC, Seedhom BB. The effect of the geometry of the tibia on prediction of the cruciate ligament forces: a theoretical analysis. Proc Inst Mech Eng [H] 1995;209:17–30.

數據

Fig. 1. Schematic diagram showing a subject performing active knee exten- exten-sion under dynamic fluoroscopic surveillance
Fig. 2. The computer graphics user interface (GUI) program to assist the visualization of the registration and reconstruction of the 3D bone kinematics
Fig. 3. The mean internal/external rotation (I/E) and abduction/adduction (Abd/Add) angles, and posterior/anterior (P/A) and medial/lateral (M/L) translations with respect to the knee flexion angle during (a and b) unloaded and (c and d) loaded conditions
Fig. 4. The mean positions of the contact points on the tibial plateau under (a) unloaded and (b) loaded conditions at intervals of 30 ◦ , up to 120 ◦ knee flexion.
+2

參考文獻

相關文件

You are given the wavelength and total energy of a light pulse and asked to find the number of photons it

 Promote project learning, mathematical modeling, and problem-based learning to strengthen the ability to integrate and apply knowledge and skills, and make. calculated

Wang, Solving pseudomonotone variational inequalities and pseudocon- vex optimization problems using the projection neural network, IEEE Transactions on Neural Networks 17

Hope theory: A member of the positive psychology family. Lopez (Eds.), Handbook of positive

volume suppressed mass: (TeV) 2 /M P ∼ 10 −4 eV → mm range can be experimentally tested for any number of extra dimensions - Light U(1) gauge bosons: no derivative couplings. =&gt;

Define instead the imaginary.. potential, magnetic field, lattice…) Dirac-BdG Hamiltonian:. with small, and matrix

incapable to extract any quantities from QCD, nor to tackle the most interesting physics, namely, the spontaneously chiral symmetry breaking and the color confinement.. 

• Formation of massive primordial stars as origin of objects in the early universe. • Supernova explosions might be visible to the most