Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vol. 20, No 1,1998

A Modified Fast Laser-Doppler Flowmetry to Measure the Pulsatile Microcirculation

in Rats

Ming-Yie

Jan"2, Hsin. Hsid2, Yuh-Ying Lin Wan$, Wei-

1. Department of Electric Engineering, National Taiwan University, Taipei, Taiwan, R.O.C.

2. Biophysic Lab., Institute of Physics, Academia Sinica, Taipei, Taiwan, R.O.C.

3. Department of Physics, National Taiwan Normal University, Taipei, Taiwan, R.O.C.

Abstract

A fast Laser-Doppler flowmetery (IXDF) modified from a commercial one to measure the pulsatile microcirculation in rats. The time constant (TC) of the integrator in the analogy circuit of the LDF was modified; the frequency response of the modification was simulated by PSPICE to estimate the high 3dB frequency. The coefficients of variance (CV) of 32 2-secod ensembles of flux in the calibration standard measured by the commerical LDF and the modified one were less than 2%. The animal tests also clearly demonstrated the pulsatile microcirculation on the renal cortex in the rat.

Key Words: LDF, Pulsatile Microcirculation, Time Constant

Introduction

The pulsatile flow in microcirculation has been debated for two decades, since the marcohemodynamic point of view assumed that the pulsatility is totally damped. Meanwhile, some distinguish scientists observed the pulsation of microcirculation indirectly by the dual-slit photometric technique in cat omentum [l] and directly by an intravital microscopy system in frog skeletal muscle [2] and in goat skeletal muscle [3]. Due to the illumination and tissue preparation, the pulsatility in a solid organ has not explored by the above methods.

A much more convenient tool, Laser-Doppler flowmetery (LDF), which was frequently used in the study of microcirculation [4] in the last two decades provides a better way to exceed the limitation. According to the Bonner's model [5], LDF can observe the averaged behaviors, flux and velocity, of a group of red blood cells (RBCs) flow into numerously randomized capillaries and arterioles under the cubic zone of the probe. We had use a commercial LDF to explored the pulsatile flux on the renal cortex in rats [6]. Due to the tradeoff of the signal to noise ratio

(SNR)

and the high frequency response, the current commercial LDF can only resolve the frequency up to 20 Hz. In a rat model, the heart rate is about 6 to 8 Hz; LDF can only observe the frequency structure up to the 3rd harmonic.

In this research, we modified the time constant (TC) in

the integrator circuit of a commercial LDF as a fast LDF. The fast LDF satisfied the need to resolve the pulsatile flux

up to the 6th harmonics (48HZ). The coefficients of variance (CV) of 32 2-secod ensembles of flux in the calibration standard measured by the commerical LDF and the modified one were estimated; both were less than 2%. A preliminary animal test was also demonstrated that the modified LDF can clearly observe the pulsatile microcirculation on the renal cortex in the rat.

Methods and Materials

A. Modified LDF

A commercial LDF, MBF3 (Moor Instruments Ltd., England), was used. The TC of the analogy output of MBF3 is about 0.05 second; MBF3 samples the analogy signal with 40Hz sampling rate and then converts it into the analogy output by a D/A converter (DAC). In the modified fast LDF, the TC of the integrator in the analogy circuit was modified. The analogy output was directly connected to a 486PC with a sample & hold card and an A/D board. The frequency responses of the integrators of both systems were simulated by PSPICE to estimate the high 3dB frequency.

B. Experimental Setup

In the commercial one, the flux signals of MBF3 were recorded from the DAC output that has 40Hz data rate. In the modified one, the flux signal was recorded from the analogy output of the integrator. Both were sent to a simultaneous sample & hold card AX753 (AXIOM Technology Co., LTD. Taiwan, R.O.C.) and then were connected to a 486PC with A/D converter board AX5621 (AXIOM Technology) in 1500Hz sampling rate.

C, Calibration and Animal Test

An optical fibre probe (PlOM+P17, plastic fiber 500pm O.D., Moor Instruments Ltd.) were connected to the LDF and then calibrated by the motility flux standard (Moor Instruments Ltd.). The standard contains 0.5% 0.482pm uniform polystyrene spheres in purified water with 0.01% surfactant to reduce conglomeration. The micro-spheres move in the water like the Brownian motion. Using both system to measure the flux in the standard, we acquired 32 2-second ensembles and compared the coefficients of

variance (CV) between them.

In the animal tests, two wistar rats were used to compare the difference between the two systems. The rats were

(2)

anesthetized with Urethane (700mgkg ip) and were placed on a heated pad to keep the body temperature. The blood pressure pulses were recorded by cannulating the I.V. catheter (Becton-Dickinson, USA) eom the caudal artery to the abdominal aorta of the rats with a catheter-tip pressure transducer (PlOEZ, Viggo-Spectramed, USA). The left kidney was exposed from the dorsal side of the rat and fixed with the cotton balls around it. The fatty capsule was separated. The optical fibre probe was put gently on the surface of the renal cortex to avoid the movement artifact introduced by the respiration or other movements.

Results

From the PSPICE simulation, the high 3dB frequency of MBF3 is at 23Hz; the one of the modified LDF is at 48Hz.Thus, the new system can resolve the flux signal in a rat model up to the 6th harmonics. The flux measured in the calibration standard at 24°C using MBF3 was 92.0f1.4 units I-AABP .... R C F / 160

,

,

100 90 8 0 140 120

c

6 0 c9 40 'O h 40

3

60 .3 50

3

30 ir, 20 20

1

, , , , , , , , , { A 0 0 0 0 4 0 8 1.2 1 6 2 Time(sec.)

Fig. 1. The abdominal aortic blood pressure (AABP) and the pulsatile renal cortical flux (RCF) measured by a commercial LDF, MBF3. I-AABP .... RCF/ 140 2o

t

i

2o 0 1 I ' ' , I ' I ' I I 0 0 0.4 0.8 1.2 1.6 2 Time(sec.)

Fig. 2. The abdominal aortic blood pressure (AABP) and the pulsatile renal cortical flux (RCF) measured by a modified

fast LDF with 0.021s time constant.

(CV=0.015); the one using the modified LDF was 102k2.0 units (CV=0.020). Both the CV were less than 2% indicates the measurements are stable. Figure 1 demonstrates the flux measured by MBF3, the stair-like shape dues to the DAC output. The flux measured by the modified LDF is shown on figure 2 that clearly exhibites the renal cortical flux is pulsatile.

Discussion

Alghough the theoretical background of the pulsatile microcirculation has not clearified, some observation had proved the existance of the pulsatile microcirculation in maliman [ 1,3,6] and in frogs [2] with different kind of tools. LDF is a better tool to observe the microcirculation with less preparation than other means. Thus, a fast LDF provides a new way to observe the pulsatile microcirculation in the rat model.

References

[l] M. Intaglietta, W.R. Tompkins, and D.R. Richardson, "Velocity Measurement in the Microvasculature of the Cat Omentum by On-line Method," Microvasc. Res., vol. 2, PP. [2] K. Tyml and A.C. Groom, "Fourier Transform Analysis of Periodic Variations of Red Cell Velocity in Capillaries of Resting Skeletal Muscle in Frog," Microvasc. Res., vol. 10, [3] J.J. Lee, K. Tyml, A. H. Menkis, R.J. Novick and F.N. Mckenzie, "Evaluation of Pulsatile and Nonpulsatile Flow in Capillaries of Goat Skeletal Muscle Using Intravital Microscopy," Microvasc. Res., vol. 48, pp. 316-327, 1994. [4] A. P. Shepherd and P.A. o b e r g ed., Laser-Doppler Blood Flowmetry., Kluwer Academic Publishers, BOSTON,

1990.

[5] R. Bonner and R. Nossal, "Model for Laser Doppler Measurements of Blood Flow in Tissue," Appl. Opt., vol. 20, [6] M.Y. Jan, T.L. Hsu, Y.Y. Lin Wang and W.K. W a g , "A

cyclocorrelogram to analyse the fast Laser-Doppler Flowmetery", Proceeding of World Congress on Medical Physics and Biomedical Engineering, Nice, France, Sep. 14-

19, p.626, 1997.

M.Y. Jan, T.L. Hsu, H. Hsiu, Y.Y. Lin Wang and W.K. Wang, "Using Fast Laser Doppler Flowmetery with an Averaged Periodogram to Analyze the Microcirculation on the Renal Cortex in Rats," Full paper Sumitted to IEEE Trans. on BME. 462-473, 1970. pp. 9-18, 1980. 0 . pp. 2097-2107, 1981. 41 6

數據

Fig.  1.  The  abdominal aortic blood  pressure (AABP) and  the  pulsatile  renal  cortical  flux  (RCF)  measured  by  a  commercial LDF, MBF3

參考文獻

相關文件

In weather maps of atmospheric pressure at a given time as a function of longitude and latitude, the level curves are called isobars and join locations with the same pressure.

The person making the measurement then slowly lowers the applied pressure and listens for blood flow to resume. Blood pressure sure pulsates because of the pumping action of the

• The  ArrayList class is an example of a  collection class. • Starting with version 5.0, Java has added a  new kind of for loop called a for each

Under the pressure of the modern era is often busy with work and financial resources, and sometimes not in fact do not want to clean up the environment, but in a full day of hard

If we recorded the monthly sodium in- take for each individual in a sample and his/her blood pressure, do individuals with higher sodium consumption also have higher blood

 Promote project learning, mathematical modeling, and problem-based learning to strengthen the ability to integrate and apply knowledge and skills, and make. calculated

* School Survey 2017.. 1) Separate examination papers for the compulsory part of the two strands, with common questions set in Papers 1A & 1B for the common topics in

Microphone and 600 ohm line conduits shall be mechanically and electrically connected to receptacle boxes and electrically grounded to the audio system ground point.. Lines in