• 沒有找到結果。

Scanning Electron Microscopy (SEM)

The scanning electron microscope (SEM) is a type of electron microscope that

images the sample surface by scanning it with a high-energy beam of electrons in a

raster scan pattern.1-2 With irradiating the sample with electron beam in a vaccum

chamber, secondary electrons (SE), backscattered electrons (BSE), characteristic x-rays

and other signals are generated as indicated in Fig 3.1. 3 The SEM mainly utilizes SE or

BSE signals to form an image. SE are produced near the sample surface, and SE image

obtained upon detecting these electrons reflects the fine topographical structure of the

sample. BSE are beam electrons that are reflected from the sample by elastic scattering.

BSE are often used in analytical SEM along with the spectra made from the

characteristic X-rays, and BSM image therefore reflects the compositional distribution

on the sample surface. Because the intensity of the BSE signal is strongly related to the

atomic number (Z) of the specimen, BSE images can provide information about the

distribution of different elements in the sample. Moreover, x-ray detector can be

equipped to the SEM, so the SEM is also applicable as an x-ray analyzer for

18

determining what elements are included in the sample.

Fig. 3.1 Signals that result from electron beam-specimen interaction.

The typical SEM instrument is made up of electron column, scanning system,

detector(s), display, vacuum system and electronics controls as shown in Fig. 3.2. 4 The

electron column of the SEM consists of an electron gun and two or more

electromagnetic lenses operating in vacuum. An electron beam is thermionically emitted

from an electron gun fitted with a tungsten filament cathode. Tungsten is normally used

19

in thermionic electron guns because it has the highest melting point and lowest vapour

pressure of all metals, thereby allowing it to be heated for electron emission, and

because of its low cost. The electron gun generates free electrons and accelerates these

electrons to energiesin the range 1-40 keV in the SEM. With the the electromagnetic

lenses (condenser and objective lenses), the electron beam generated by the electron gun

is coveraged into a fine beam in a high-vacuum column. Typically the electron beam is

defined by probe diameter, probe current and probe convergence. And by applying a

scan signal to the deflection coils, the electron beam is scanned along the sample

surface in X and Y direction. Objective lens is used to converge the electron beam into a

fine beam and focus it onto the sample surface. When the primary electron beam

interacts with the sample, the electrons lose energy by repeated random scattering and

absorption within a teardrop-shaped volume of the specimen known as the interaction

volume, which extends from less than 100 nm to around 5 μm into the surface. The

energy exchange between the electron beam and the sample results in the reflection of

high-energy electrons by elastic scattering, emission of secondary electrons by inelastic

scattering and the emission of electromagnetic radiation, each of which can be detected

by specialized detectors. A secondary electron detector for detecting signals produced

from the sample could converts signal to electric one. Finally the photomultiplier are

used to amplify the signals, the amplified electrical signal output is displayed as a

20

two-dimensional intensity distribution that can be viewed and photographed on a video

display, which are displayed as variations in brightness.

Fig. 3.2 The structural scheme of a typical scanning electron microscope.

When imaging in the SEM, samples must be electrically conductive at the surface,

and electrically grounded to prevent the accumulation of electrostatic charge at the

surface. Metal objects require cleaning and mounting on a specimen stub for SEM. It

21

tend to charge when scanned by the electron beam for nonconductive specimens, and

this causes scanning faults and other image artifacts especially in secondary electron

imaging mode. Therefore, they are usually coated with an ultrathin coating of

electrically conducting material by sputter coating or evaporation. Additionally, coating

could increase signal/noise ratio for samples which is composed of low atomic number

(Z) atoms. This improvement arises because secondary electron emission for high-Z

materials is enhanced. Figure 3.3 show a photo of the SEM (JEOL, JSM-6500F) image

used in this work.

Fig. 3.3 Photo of scanning electron microscopy (JSM-6500F, JEOL)

22

The image details and resolution in the SEM are determined not by the size of the

electron probe by itself but rather by the size and characteristics of the interaction

volume, which is the area of the sample excited by the electron beam to produce a

signal. When the accelerated beam electrons strike a specimen they penetrate inside it to

depths of about 1 μm and interact both elastically and inelastically with the solid,

forming a limiting interaction volume from which various types of radiation emerge,

including BSE, SE, characteristic and brehmsstrahlung x-rays, and

cathodoluminescence in some materials. The combined effect of elastic and inelastic

scattering controls the penetration of the electron beam into the solid. The resulting

region over which the incident electrons interact with the sample is known as interaction

volume. The interaction volume has several important characteristics, which determine

the nature of imaging in the SEM. The energy deposition rate varies rapidly throughout

the interaction volume, being greatest near the beam impact point. The interaction

volume has a distinct shape as shown in Fig 3.4. For low-Z target it has distinct pear

shape. For intermediate and high-Z number materials the shape is in the form of

hemi-sphere. The interaction volume increases with increasing incident beam energy

and decreases with increasing average atomic number of the specimen. For secondary

electrons the sampling depth is from 5 to 50 nm and diameter equals the diameter of the

area emitting backscattered electrons. BSE are emitted from much larger depths

23

compared to SE. Finally the resolution in the SEM is controlled by the size of the

interaction volume.

The effective interaction volume can be calculated using the electron range (R): 5

 ൌͲǤͲʹ͹͸ܣܧଵǤ଺଻

ߩܼ଴Ǥ଼ଽ ሺߤ݉ሻ

Where A is the atomic weight (in g/mole), Z is the atomic number,Ȱis the density

(in g/cm3), and E0 is the energy of electron beam (in KeV)

Fig. 3.4 The range and spatial resolution of backscattered electrons, secondary electrons, X-rays, and Auger electrons for electrons incident on a solid.

24 3.2 Cathodoluminescence (CL)

When an electron is promoted into the conduction band, the electron and hole

become free and they can move independently in corresponding bands.6 The major

electron-hole recombination pathway between the conduction and valence bands

involve donor and/or acceptor levels, recombination via deep level traps, and

recombination at the surface. Electron beam excitation in general leads to emission by

all the luminescence mechanisms present in the material. Photoluminescence emission

may strongly depend on the excitation hv, which can be used for selective excitation of

particular emission processes. Cathodoluminescence analysis of materials, on the other

hand, can provide depth-resolved information by varying the electron beam energy as

shown in Fig 3.4. In general, electron beam energy of upon to 30 keV can be used.

Cathodolumiscence analysis enables one to assess various properties of the material

with a spatial resolution down to 1 ƶm or less. Spectroscopic CL and monochromatic

imaging can be used in identification and measurement of luminescence center

concentration and distributions, as well as in the determination of the composition of

compound materials.

Here, the CL spectrometer (Gatan, MonoCL3) as shown in Figs. 3.5 is adapted to

the SEM with the energy of electron beam from 1kVto 30 kV. PMT and detectors are

used to gathered the visible and IR emission spectra.

25

Fig. 3.5 Photo of the CL spectrometer.

相關文件