Default Setup
3. A broken line indicates the waveform display no longer matches the controls
This happens when you stop the acquisition, and change a control setting that the oscilloscope is not able to apply to the displayed waveform. For example, changing the trigger controls on a stopped acquisition causes a broken-line waveform.
Key Points Persistence. The oscilloscope displays persistence waveform data with less intensity than "live" waveform data. With Persistence set to Infinite, record points accumulate until a control is changed.
Option Comments
Off Removes default or old waveforms whenever new waveforms display
Time limit Displays new waveforms at normal intensity and old waveforms at a lower intensity; erases old waveforms when they reach the time limit
Infinite Older waveforms become less bright but always remain visible;
use Infinite persistence to look for infrequent events and to measure long term peak-to-peak noise
XY Format. Use the XY format to analyze phase differences, such as those represented by Lissajous patterns. The format plots the voltage on channel 1 against the voltage on channel 2, where channel 1 is the horizontal axis and
Reference
channel 2 is the vertical axis. The oscilloscope uses the untriggered Sample acquisition mode and displays data as dots. The sampling rate is fixed at 1 MS/s.
NOTE. The oscilloscope can capture a waveform in normal YT mode at any sampling rate. You can view the same waveform in XY mode. To do so, stop the acquisition and change the display format to XY.
In XY format, the controls operate as follows:
The channel 1 Vertical Scale and Vertical Position controls set the horizontal scale and position.
The channel 2 Vertical Scale and Vertical Position controls continue to set vertical scale and position.
The following functions do not work in XY display format:
Autoset (resets display format to YT) Autorange
Automatic measurements Cursors
Reference or Math waveforms Save/Recall ► Save All Time base controls Trigger controls
Help
Push the Help button to display the Help menu. The topics cover all the menu options and controls of the oscilloscope. (See page xiv, Help System.)
Horizontal
You can use the horizontal controls to set up two views of a waveform, each with their own horizontal scale and position. The horizontal position readout shows the time represented by the center of the screen, using the time of the trigger as zero. When you change the horizontal scale, the waveform will expand or contract around the screen center.
Options Comments
Main The main horizontal time base setting is used to display the waveform
Reference
Options Comments
Window Zone Two cursors define a window zone
Adjust the window zone with the Horizontal Position and Horizontal Scale controls
Window Changes the display to show the waveform segment (expanded to screen width) within the window zone
Set Holdoff Displays the holdoff value; push the option button and use the multipurpose knob to adjust
NOTE. You can push the horizontal option buttons to switch between an entire waveform display and a zoomed part of it.
A readout near the top right of the screen displays the current horizontal position in seconds. An M indicates the Main time base and a W indicates the Window time base. The oscilloscope also indicates horizontal position with an arrow icon at the top of the graticule.
Knobs and Buttons Horizontal Position Knob. Use to control the position of the trigger relative to the center of the screen.
The trigger point can be set to the left or the right of the center of the screen. The maximum number of divisions to the left depends on the Horizontal Scale (time base) setting. For most scales, the maximum is at least 100 divisions. Placing the trigger point off the screen to the left is called Delayed Sweep.
Set To Zero Button. Use to set the horizontal position to zero.
Horizontal Scale Knob (seconds/division). Use to change the horizontal time scale to magnify or compress the waveform.
Key Points Horizontal Scale. If waveform acquisition is stopped (using the Run/Stopor Single button), the Horizontal Scale control expands or compresses the waveform. Use to zoom in on a detail of the waveform.
Scan Mode Display (Roll Mode). When the Horizontal Scale control is set to 100 ms/div or slower and the trigger mode is set to Auto, the oscilloscope enters the Scan acquisition mode. In this mode, the waveform display updates from left to right. There is no trigger or horizontal position control of waveforms during scan mode. (See page 77, Scan Mode Display.)
Window Zone. Use the window zone option to define a segment of a waveform to see in more detail (zoom). The Window time base setting cannot be set slower than the Main time base setting.
Reference
Vertical bars define Window Zone
Main time base displayed Window zone displayed
Window. Expands the Window Zone to cover the entire screen. Use to change between two time bases.
NOTE. When you change between the Main, Window Zone, and Window views, the oscilloscope erases any waveform saved on the screen through persistence.
Persistence is erased with Horizontal menu changes.
Holdoff. Use holdoff to help stabilize the display of complex waveforms. (See page 100, Holdoff.)
Math
Push the Math button to display waveform math operations. Push the Math button again to remove math waveforms. (See page 104, Vertical Controls.)
Options Comments
+, -, ×, FFT Math operations; see the next table
Sources Sources used for the operations; see the next table Position Use the multipurpose knob to set the vertical position of the
resultant Math waveform
Vertical Scale Use the multipurpose knob to set the vertical scale of the resultant Math waveform
The Math Menu includes Sources options for each operation.
Operation Sources option Comments
CH1 + CH2 Channels 1 and 2 are added
together + (addition)
CH3 + CH41 Channels 3 and 4 are added together
Reference
Operation Sources option Comments
CH1 - CH2 The channel 2 waveform is
subtracted from the channel 1 waveform
CH2 - CH1 The channel 1 waveform is
subtracted from the channel 2 waveform
CH3 - CH41 The channel 4 waveform is subtracted from the channel 3 waveform
- (subtraction)
CH4 - CH31 The channel 3 waveform is subtracted from the channel 4 waveform
CH1×CH2 Channels 1 and 2 are multiplied together
× (multiplication)
CH3×CH41 Channels 3 and 4 are multiplied together
FFT (See page 55.)
1 Available only on a 4-channel oscilloscope.
Key Points Waveform Units. The combination of source waveform units determine the resulting units for the Math waveform.
Waveform unit Waveform unit Operation Resulting Math unit
V V + or - V
Push the Measure button to access automatic measurements. There are eleven types of measurements available. You can display up to five at a time.
Push the top option button to display the Measure 1 Menu. You can choose the channel on which to take a measurement in the Source option. You can choose the type of measurement to take in the Type option. Push the Back option button to return to the Measure Menu and display the selected measurements.
Key Points Taking Measurements. You can display up to five automatic measurements at a time. The waveform channel must be on (displayed) to make a measurement.
Reference
Automated measurements cannot be taken on reference waveforms, or while using XY or scan mode. The measurements update about two times per second.
Measurement type Definition
Freq Calculates the frequency of the waveform by measuring the first cycle Period Calculates the time of the first cycle
Mean Calculates the arithmetic mean amplitude over the entire record
Pk-Pk Calculates the absolute difference between the maximum and minimum peaks of the entire waveform
Cyc RMS Calculates a true RMS measurement of the first complete cycle of the waveform
RMS Calculates a true RMS measurement for all 2,500 samples from one frame of the waveform data Cursor RMS Calculates a true RMS measurement of the waveform data from the selected starting to the ending
point
Min Examines the entire 2500 point waveform record and displays the minimum value Max Examines the entire 2500 point waveform record and displays the maximum value Rise Time Measures the time between 10% and 90% of the first rising edge of the waveform Fall Time Measures the time between 90% and 10% of the first falling edge of the waveform
Pos Width Measures the time between the first rising edge and the next falling edge at the waveform 50% level Neg Width Measures the time between the first falling edge and the next rising edge at the waveform 50% level Duty Cyc Measures the ratio of the positive pulse duration to the whole cycle
Phase Calculates the phase angle difference of signals from two different channels, using the rising edge of the first signal compared to the rising edge of the second signal
Delay Calculates the time difference from two different channels using the rising edge of the first signal compared to the rising edge of the second signal
None Does not take any measurement
When the Save All ► PRINT Button option is set to Prints, you can push the print button to send the screen image to a printer.
You can set up the oscilloscope to send a screen image to your printer through the Utility ► Options ► Printer Setup menu.
Option Setting Comments
Ink Saver On, Off Prints the screen image on a
white background when you select On
Layout1 Portrait, Landscape Printer output orientation
Abort Printing Stops sending the screen image
to the printer
Reference
Option Setting Comments
Paper Size2 Default, L, 2L, Hagaki Postcard, Card Size, 10 x 15 cm, 4" x 6", 8"
x 10", Letter, 11" x 17", A0, A1, A2, A3, A4, A5, A6, A7, A8, A9, B0, B1, B2, B3, B4, B5, B6, B7, B8, B9, 89 mm Roll (L), 127 mm Roll (2L), 100 mm Roll (4"), 210 mm Roll (A4)
Displays settings available on your PictBridge compatible printer
Image Size2 Default, 2.5 x 3.25 in, L (3.5 x 5 in), 4 x 6 in, 2L (5 x 7 in), 8 x 10 in, 4L (7 x 10 in), E, Card, Hagaki card, 6 x 8 cm, 7 x 10 cm, 9 x 13 cm, 10 x 15 cm, 13 x 18 cm, 15 x 21 cm, 18 x 24 cm, A4, Letter Paper Type2 Default, Plain, Photo, Fast Photo
Print Quality2 Default, Normal, Draft, Fine Date Print2 Default, Off, On
ID Print2 Default, Off, On
1 The printer may override your selection for best fit.
2 If your selection is not supported by the printer, the oscilloscope uses the Default setting.
The alternative function of the print button is to save data to a USB flash drive.
(See page 63, USB Flash Drive and Device Ports.)
The oscilloscope was designed to print to any PictBridge compatible printer.
Refer to the product documentation for your printer to determine if the printer is PictBridge compatible.