Three dimensional electrode array for cell lysis via electroporation

Kuan-Ying Lu

a

, Andrew M. Wo

a,

, Ying-Jie Lo

a

, Ken-Chao Chen

a

,

Cheng-Ming Lin

a

, Chii-Rong Yang

b

aInstitute of Applied Mechanics, National Taiwan University, Taipei 106, Taiwan bDepartment of Mechatronic Technology, National Taiwan Normal University, Taipei 106, Taiwan

Received 7 March 2006; received in revised form 10 June 2006; accepted 9 August 2006

Abstract

Microfabricated devices for cell lysis have demonstrated many advantages over conventional approaches. Among various design of microdevices that employ electroporation for cytolysis, most utilize Ag/AgCl wires or 2D planar electrodes. Although, simple in fabrication the electric field generated by 2D electrodes decays exponentially, resulting in rather non-uniform forcing on the cell membrane. This paper investigates the effect of electric field generated by 3D cylindrical electrodes to perform cell lysis via electroporation in a microfluidic platform, and compared with that by 2D design. Computational results of the electric field for both 2D and 3D electrode geometries showed that the 3D configuration demonstrated a significantly higher effective volume ratio—volume which electric field is sufficient for cell lysis to that of net throughflow volume. Hence, the efficacy of performing cell lysis is substantially greater for cells passing through 3D than 2D electrodes. Experimentally, simultaneous multi-pores were observed on leukocytes lysed with 3D electrodes, which is indicative of enhanced uniformity of the electric field generated by 3D design. Additionally, a single row of 3D electrode demonstrated a substantially higher lysing percentage (30%) than that of 2D (8%) under that same flow condition. This work should aid in the design of electrodes in performing cell lysis via electroporation.

© 2006 Elsevier B.V. All rights reserved.

Keywords: Cell lysis; Electrode design; Electroporation; Microfluidics

1. Introduction

Microfluidics has gained wide acceptance in bio/chemical studies in recent years (Bange et al., 2005; Pihl et al., 2005; Stone et al., 2004). As a result, the idea of packaging tradi-tional laboratory instrumentation onto a chip – or lab-on-a-chip – is becoming a reality. Fluid transport and processing, sens-ing and/or actuation, and analysis elements in an integrated microfluidics platform are being accomplished with microflu-idic components such as micropump, microvalve, micromixer, and microelectrodes (Isik et al., 2005) to perform specific tasks. The focus of this work is to utilize microelectrodes to perform cell lysis within a chip environment.

There are a number of approaches in performing lysis on the cell phospholipid membrane, with the predominant goal of extracting various elements – e.g. membrane and intracellular protein, RNA, and DNA – to study physiologic and pathologic

Corresponding author. Tel.: +886 2 33665656; fax: +886 2 23639290. E-mail address:andrew@iam.ntu.edu.tw(A.M. Wo).

processes at the cellular level. These approaches can be clas-sified into two groups: chemical and physical means (Graham and Rickwood, 1997). Chemical approaches mostly utilize lysis buffer with surfactant to solubilize the lipid membrane. How-ever, the use of surfactant requires additional treatment to remove itself from the sample. Another main drawback is that although chemical methods are widely used for extraction of DNA, it is not ideal if the prime concern is integrity of protein, e.g. organelle (Lu et al., 2005). Moreover, in the limiting case of lysing a single cell picoliter volume of lysis solution in a microfluidic environment is needed (Irimia et al., 2004).

On the other hand, physical means of cell lysis employ a wide variety of approaches. Thermal method, in a reactor, is perhaps the most common (El-Ali et al., 2005; Lee et al., 2005), proba-bly due to compatibility with PCR and DNA analyses. However, this may not be the method of choice for most protein analysis since protein denatures much easier than DNA in elevated tem-perature environment.Marentis et al. (2005)used energy from sound wave from a piezoelectric minisonicator to lyse leukemia HL-60 cells and Bacillus subtilis bacterial spores in microfluidic environment in the absence of chemical denaturants, enzymes or 0956-5663/$ – see front matter © 2006 Elsevier B.V. All rights reserved.

(2)

microparticles.Di Carlo et al. (2005)used poration of hydrox-ide ion through cell membrane for cell lysis.Marmottant and Hilgenfeldt (2003)demonstrated a novel approach in which con-trolled, oscillating microbubbles are sufficient to rupture lipid membranes.

Electroporation, another physical approach, is widely used to create transient pore through the membrane to deliver dyes, drugs, proteins, peptides, RNA, and DNA into cells (Neumann et al., 1989; Tsong, 1991). The technique applies an electric field which raises the transmembrane potential above a thresh-old (0.2–1.2 V) sufficient to disrupt the lipid by-layer and causes pores to form. In recent years, marriage of this technique with microfluidics to attain the same end has been widely accepted (Huang and Rubinsky, 2003; Lin and Huang, 2001; Lin et al., 2004; Shin et al., 2004), and was reviewed byOlofsson et al. (2003). Electroporation has been applied in microfluidic plat-form primarily to lyse the membrane permanently (Lee and Tai, 1999; Lu et al., 2005; Xia et al., 2005).

The design of electrodes is of crucial import in lysing cell, since the electrodes define the electric field. The majority of the effort thus far of utilizes simply with Ag/AgCl wires (Gao et al., 2004; Khine et al., 2005; Lee and Tai, 1999; Xia et al., 2005) or microfabricated planar electrode design (Lee and Tai, 1999). The drawback for planar electrodes is the exponential decay of the electric field away from the plane of the electrode. With cell dimension typically not much greater than the channel height, a cell would experience large non-uniformity in the electric field during electroporation. Very limited studies utilize 3D electrodes for cytolysis. Lu et al. (2005), for example, successful employed pointed, 3D electrodes with the primary intent of preserving organelle integrity. However, the efficacy of electrode design in lysing cells – not being the focus ofLu et al. (2005)– was not investigated.

This paper aims to quantify the effect of utilizing 3D elec-trodes (Fig. 1a) with the aim of enhancing cell lysis via elec-troporation in a microfluidic platform. Both computation and experimental efforts using leukocytes were undertaken, with comparison made to planar electrode design. Results from this work should enhance design of electrodes to perform cell lysis using electroporation.

2. Electroporation

The phenomenon utilizes electric field to create transient or permanent pore (s) on the cell membrane. The phenomenon occurs when sufficient voltage required for dielectric breakdown of the membrane, about 0.2–1.5 V, is imposed by an external electric field (Khine et al., 2005). This transmembrane potential ϕ for a spherical cell can be expressed as (Neumann et al., 1989), ϕ(t) = 1.5aE cos θ  1− exp  −t τm  , (1)

where a is the radius of cell, E the external electric field,θ the polar angle measured from the center of the cell with respect to the direction of electric field, andτm is the charging time of the cell membrane. When the transmembrane potential is

Fig. 1. Three dimensional cylindrical electrodes. (a) Sketch of leukocytes approaching a singe row of electrode. (b) Micrograph of fabricated device. The diameter of each electrode is 50␮m and spacing between electrodes is 20 ␮m. The height of each electrode is 50␮m.

strong enough, pores will be formed. In general, pores can be of two states: reversible electrical breakdown, with the pores reseal themselves when the external electric field is removed, and irreversible breakdown, where the damage is permanent. The signal of electroporation is usually in the form of pulses in order to avoid damaging DNA or organelles, with the pulse duration between 10 and 1000␮s. The charging time of cell membrane could be shown as (Muller et al., 2001):

τm= aCm  1 σi + 1 2σe  , (2)

where Cm is the membrane capacitance, σi intracellular conductivity, andσeis conductivity of the external medium. If cells are embedded in high conductivity medium, the charging time will be shorter. For most mammalian cells, the charging time is quite short—on the order of several hundreds nano seconds (Muller et al., 2001). Therefore, pulses of sufficiently long duration (100␮s used here), compared to the membrane charging time, behaves as if the pulse excitation is DC. Hence, the reduced form of Eq.(1)can be used instead, namely,

ϕ ≈ 1.5aE cos θ (3)

(3)

3. Computation and experimental aspects 3.1. Computation

Computational effort was undertaken to aid the design of the microdevice in characterizing the electric field prior to fabrication. The commercial code Comsol Multiphysics (Com-sol Group) was used to compute the electric field of planar (pointed and circular) and 3D electrodes (cylinders). The com-putational domain of the 3D configuration covered a region of 290␮m width (cross stream dimension), which contained the extent of four cylinders. The streamwise extent was 250␮m, which included only one row of electrodes. The computation was conducted to a height of 170␮m (actual device chan-nel height is 50␮m) to ensure that sufficient decay in the electric field had indeed taken place. Boundary conditions were: vanishing gradient of the electric potential normal to the surface at all locations except at the electrode surface. On the electrodes, the voltage prescribed was a constant 10 V. The spatial increment of the computational grid was 1␮m uniformly.

We also calculated the effective volume ratio of the three elec-trode geometries under different applied voltages and channel heights. This ratio is defined as the volume of throughflow with transmembrane potential sufficient for cytolysis to the total vol-ume that the fluid occupied. For the 3D electrode configuration the total volume does not include the volume of the cylinders. In other words, this ratio provides an idea of the efficacy of cell lysis as cells flow through the region with electrodes within the microchannel. Using a typical transmembrane potential of 1 V, the threshold electric field strength for cell lysis was esti-mated using Eq.(3)for membrane thickness of approximately 5 nm and resulted at 2000 kV/cm (Tsong, 1991). Using this value and results from computation of electric field for various elec-trode geometries, calculations for the effective volume ratio were made for channel height of 50 and 100␮m.

3.2. Fabrication

Fabrication process of the device consists of fabrication of the planar electrodes, 3D electrodes, and the microchannel. Each component is described in some detail below.

The planar electrodes was made via conventional MEMS approach as follows. First, Au/Cr (200/30 nm) was deposited on glass substrate by electron beam evaporation. Positive photore-sist (PR) (model S1813, Shipley) were spincoated on the metal layer, then exposed to UV light through a patterned transparency mask, followed by development of the PR. Then, the PR was stripped leaving the patterned electrodes. Teflon was applied, as protection for electrodes from corrosion due to applied pulses, using the following three-step procedure. First, a layer of PR was spincoated and, after exposure to UV light through a masked pattern and subsequent development, only the portion above the metal layer was removed. Second, Teflon was spincoated onto the bare metal and the remaining PR (between electrodes). Finally, lift-off process was performed on the PR, leaving the Teflon to protect the metal layer only.

To fabricate the 3D electrodes, first, a thin metal layer with the same area as the base area of the 3D cylindrical elec-trodes was formed for subsequent electroforming of the 3D electrode structure. This metal “seed layer” had identical fab-rication steps as that of depositing bare metal layer on the substrate of the planar electrodes described earlier (via elec-tron beam evaporation). The fourth step fabricated the SU-8 mold in which the 3D electrodes would electroform inside. The recipe for SU-8 structure was as follow. It was spun at 4000 rpm for 30 s, which corresponded to a height of 70␮m even though the height of 3D electrodes was 50␮m. The rea-son for the over-height of the mold was to avoid over-grown of copper during the electroforming process, which would also cause a detrimental short circuit. After spinning, the SU-8 was soft baked at 65◦C for 8 min and 95◦C for 25 min. After exposure for 33 s in 14 mJ/cm2 power of UV light, post-exposure bake was made at 65◦C for 2 min and 95◦C for 7 min, followed by development with agitation aided by ultrasound.

Then, electroforming was employed to fabricate the 3D cop-per electrodes. The process is very similar to electroplating with two differences. First, electroforming is normally used in thick deposition (tens of micron) and, second, the deposited structure can be made to conform to a mold. Hence, the topolog-ical accuracy of the 3D electrodes by electroforming is essen-tially determined by the quality of the SU-8 mold. If a crack underneath the SU-8 structure exists the electroforming solu-tion will permeate into the crack and metal will be deposited, which might cause a detrimental short circuit between adja-cent electrodes. The copper electrodes were fabricated using a sulfate-based copper electrolytic bath. The plating bath con-tained 200 g/l copper sulfate (CuSO4·5H2O), 85 g/l sulfuric acid (H2SO4), and 50 mg/l chloride ion (Cl−) operating at room tem-perature and pH below 1. The content of the bath was kept clean by circulation and agitation by pumping through a filter with 0.1 mm pores. Small quantity of surfactant was added to make the electrolyte more hydrophilic and the electroformed copper brighter. The pre-treatment of the photoresist mold and the pro-cess control of electroplating deserved attention to ensure that metallic microstructures of high quality can be obtained. Metal was deposited at a current density of 1 A/dm2 for about 4 h. Next, the SU-8 mold was removed with Remover PG (Micro Chem Corp.) at 80◦C under ultrasonic agitation. Teflon was also used to protect the electrodes with the same procedure as that of the planar electrodes.Fig. 1b shows the fabricated 3D electrodes.

The microchannel was fabricated with (poly)dimethyl-siloxane (PDMS) using conventional soft lithographic process (Xia and Whitesides, 1998). SU-8 was used in forming the master mold for the microchannel with 50␮m height, 290 ␮m width, and about 1 cm length. Replica molding technique for the PDMS elastomer allowed the microchannel geometry to take shape. Oxygen plasma treatment was applied on both the molded PDMS microchannel and the substrate with electrodes to allow irreversible bonding of the two parts to form the encapsulated device. Finally, inlet and outlet ports were inserted for pumping with syringe.

(4)

Fig. 2. Sketch of the entire experimental setup with the cell lysis microdevice. Syringe was used to pump the leukocytes through the device with 10 V, 100␮s pulse supplied from a function generator. Images were captured by CCD for post-processing.

3.3. Experimental setup

Leukocytes (Chinese Blood Donation Association) were diluted with physiological saline solution in ratio of 1:100 by volume. Then, the cells were injected into the straight microchannel by a syringe pump at the inlet and collected at the outlet. Excitation to the electrodes was applied with 10 V, 100␮s pulse at 4 Hz using a function generator (model 195; Wavetek Universal). Micrograph images were recorded by CCD camera, and stored for data interrogation.Fig. 2 shows the experimental setup.

Trypan blue was used to verify successful cytolysis.Khine et al. (2005)also used Trypan blue dye, and independently with

Fig. 4. Computational results of the effective volume ratio defined as volume of throughflow with transmembrane potential sufficient for cytolysis to the total throughflow volume. Applied voltage of 0–35 V and channel heights of: (a) 50␮m and (b) 100 ␮m are computed. Results show the effective volume ratio for 3D electrodes is larger than that of either pointed or circular planar electrodes considered.

Fig. 3. Computed iso-electric field contour of three electrode configurations: (a) 2D pointed planar, (b) 2D circular planar, and (c) 3D cylindrical electrodes, with mid-height plane (z = 25␮m) contour shown in (d). The applied voltage was 0 V and, at the adjacent electrode, 10 V for all cases. Results show that the electric field decays rapid away form the electrodes for the planar cases, (a and b), but does not decay much with height for the 3D electrode. Results at mid-height of 3D electrode, (d), is representative of that at other heights since (c) confirms that the electric field is almost invariant along the cylindrical axis.

(5)

cytoplasmic dye Calcein AM, to definitively identify the lysed cells. Although there are natural pores on cell membrane, exclu-sion measurement of Trypan blue is routinely used as a test of cell viability since healthy membrane would prevent seepage of the toxin. Once the membrane is made permeable (via elec-troporation in this work) the dye could accumulate within the cell staining the intracellular elements a distinct bluish color. In the event that the membrane disintegrated, location of the color change (turned bluish) would be indicative of that of the intracellular elements.

4. Results and discussion

4.1. Computation of the electric field

Fig. 3shows results of the computed iso-electric field surfaces for the two planar electrodes – pointed (Fig. 3a) and circular (Fig. 3b) – and the 3D electrodes (Fig. 3c and d). BothFig. 3a and b suggest that large spatial gradient exists near the 2D electrodes as indicated by large values (non deep blue colored area) near the edge of the electrodes and decay rapidly away. For the 3D cylindrical electrodes (Fig. 3c), even though there exists some variation near the top and bottom of the electrodes – as can be expected due to edge effect – the electric field predominantly does not decay much over the entire height.Fig. 3d plots the electric field distribution at the mid-height (25␮m) plane of the cylinder (or mid-height plane ofFig. 3c), which is representative of the electric field at other heights except very close to the end regions.

As a cell passes through the area between the cylindrical elec-trodes it enters into a region of approximately 20␮m width of high field strength area, as indicated inFig. 3d. The extend of this region is large enough such that it can easily encompasses the entire leukocyte (10␮m in diameter). With sufficient electric field strength, gradient of transmembrane potential is setup to accomplish cytolysis in a fashion where the cell would experi-ence much more uniform distribution of field strength than that under planar electrodes.

Fig. 4 presents the results of effective volume ratio, as defined previously. Results show this ratio for the 3D elec-trode is larger than planar elecelec-trodes in all cases. Moreover, at a particular applied voltage, lysing effectiveness decreases as the channel height increases for the planar electrodes. This overall trend is indicative of the rapid decay nature of pla-nar design, where the electric field of the 3D design essen-tially remains constant over the entire height. Although the planar electrodes are not much worse off than the 3D design at 50␮m channel height, the difference in lysing effective-ness between the two designs increases as the channel height increases.

Lysing events in the microdevice are presented inFig. 5for the pointed, planar electrodes andFig. 6for 3D electrodes. We adopted two independent approaches in verifying that the mem-brane was indeed lysed. First, when the optical appearance of a cell turns from a normal cell with intact membrane to that of a “ghost cell” this would be indicative that its membrane is no longer acting as a boundary of intracellular elements and thus begins to disintegrate.Lu et al. (2005)also employed this approach, which proved to be effective in deciphering lysed

Fig. 5. Micrographs of evolution of lysing of leukocytes by pointed, planar electrodes by excitation of 10 V, 100␮s pulse: (a–c) demonstration of lysed cells became “ghost” cells, as indicated. This appearance is a result of disintegration of membrane allowing intracellular organelles to escape hence disappearing of the entire cellular structure and (d–f) micrographs of fluid dyed with Trypan Blue which turned strongly bluish when came in contact with intracellular elements. Arrows also confirmed disappearing of the cellular structure.

(6)

Fig. 6. Micrographs of time evolution of leukocytes being lysed by 3-D electrodes under excitation of 10 V, 100␮s pulse (same as that inFig. 5). Graduate disintegration of the cellular structure (becoming “ghost” cells) is indicative of successful cytolysis. Moreover, simultaneous multiple pores on the cell membranes were observed as shown by arrows.

cells. Second, Trypan blue was used, with the aforementioned function (see Section 3.3). Fig. 5 presents a series of events of leukocytes closed to the planar electrodes being lysed. The events inFig. 5a–c showed some cells were being turned into ghost cells, with their membrane gradually opening and fainting within a few seconds. In the events here, the entire duration for lysis was about 1 s. Long time observation of these cells con-firmed that they eventually disappeared.Fig. 5d–f used Trypan blue to locate the intracellular elements. Results suggested the cells closed to the electrode surface, as observed, were being definitively lysed. Staining here was indicative that portions of the intracellular element were located outside the disintegrated membrane. In this case, the entire lysis process occurred in about 16 s. The difference in the duration of lysis between that from Fig. 5a–f was likely due to the difference in height from the elec-trode surface for the two cases, indicative that the exponential decay nature of the electric field causing substantial difference in lysing results.

Lysis of leukocytes with the 3D electrodes demonstrated marked difference from that of the planar geometry. As cells flow past the electrodes, Fig. 6 shows simultaneous multiple pores were observed during the lysis process. This is likely due to that the cell was exposed to a more uniform electric field, as suggested by results ofFig. 3d and discussed earlier. The enhanced uniformity in electric field, as compared to that of produced by planar electrodes, apparently caused multiple loca-tions of membrane to exceed that of the transmembrane potential simultaneously and resulted in that ofFig. 6. The lysis process required about three seconds.

The efficacy of cytolysis was further quantified by compar-ing the lysis percentage of both 2D (pointed) and 3D electrode design. This percentage is defined as the amount of lysed cells per total number of cells passing through a single row of elec-trode. This was done by image post-processing in counting the number of lysed cells for both devices under approximately the same background flow rate. Results show that for 2D electrodes, the lysing percentage was 8.0± 6% for flow rate of 57 ␮m/s. For 3D electrodes, the lysing percentage was 30.5± 5% for flow rate of 50␮m/s. This result suggests that lysing effect of 3D

elec-trodes indeed enable cell lysis much more pronouncedly than that of planar electrodes.

5. Conclusions

Electroporation – a dominant method to transfer DNA and other genetic substance into the cell – has been utilized in this work to investigate the efficacy of cell lysis. The focus is the characterization of electric field generated by electrodes with planar and 3D cylindrical geometries in a microfluidic device. Computational results show that the effective lysis volume ratio of the 3D electrode is higher than that of the planar design. The primary reason being, unlike the planar electrodes, the electric field of the 3D electrodes does not decay rapidly away from the substrate, resulting in much greater effective volume. Exper-imentally, leukocytes were lysed successfully in seconds by both types of electrode designs, and intracellular elements were extracted through cell membrane as observed by increased inten-sity of Trypan blue. Simultaneous multiple pores were observed on leukocyte lysed with 3D electrodes and just single pore with planar electrodes. This is believed to be due to more uniform electric field of 3D geometry than that of planar geometry.

Much improved lysing percentage for 3D electrodes (30%) compared to planar design (8%) carries great design implication. If one desires to increase lysing percentage for 3D design to, say, 90%, one can simply increase the number of rows of electrode that the cells would pass through to three or four. In contrast, many more pairs of planar electrodes are needed to achieve the near-complete lysing process.Lu et al. (2005)used an estimated 30 sets of pointed 3D electrodes to perform cell lysis as a cell progressed downstream—although the actual number of pairs of electrodes to complete lysis was not evident. Over-lysing a cell would conceivably damage the intracellular elements. Since cell lysis is a process with its end goal being extraction of these elements, gene transfection, or the like, the process should be performed to a degree just suffice for the specific application. Hence, this work should aid in the design of electrodes in per-forming cell lysis via electroporation in a particular application.

(7)

Acknowledgements

The authors are grateful for partial financial support from the National Science Council (Grant NSC 94-2212-E-002-054) and the Ministry of Economic Affairs.

References

Bange, A., Halsall, H.B., Heineman, W.R., 2005. Biosens. Bioelectron. 20 (12), 2488–2503.

Di Carlo, D., Ionescu-Zanetti, C., Zhang, Y., Hung, P., Lee, L.P., 2005. Lab Chip 5 (2), 171–178.

El-Ali, J., Gaudet, S., Gunther, A., Sorger, P.K., Jensen, K.F., 2005. Anal. Chem. 77 (11), 3629–3636.

Gao, J., Yin, X.F., Fang, Z.L., 2004. Lab Chip 4 (1), 47–52.

Graham, J.M., Rickwood, D., 1997. Subcellular Fractionation: A Practical Approach. Oxford University Press, New York.

Huang, Y., Rubinsky, B., 2003. Sens. Actuators A Phys. 104 (3), 205–212. Irimia, D., Tompkins, R.G., Toner, M., 2004. Anal. Chem. 76 (20), 6137–6143. Isik, S., Berdondini, L., Oni, J., Blochl, A., Koudelka-Hep, M., Schuhmann, W.,

2005. Biosens. Bioelectron. 20 (8), 1566–1572.

Khine, M., Lau, A., Ionescu-Zanetti, C., Seo, J., Lee, L.P., 2005. Lab Chip 5 (1), 38–43.

Lee, C.Y., Lee, G.B., Lin, J.L., Huang, F.C., Liao, C.S., 2005. J. Micromech. Microeng. 15 (6), 1215–1223.

Lee, S.W., Tai, Y.C., 1999. Sens. Actuators A Phys. 73 (1–2), 74–79. Lin, Y.C., Huang, M.Y., 2001. J. Micromech. Microeng. 11 (5), 542–547. Lin, Y.C., Li, M., Wu, C.C., 2004. Lab Chip 4 (2), 104–108.

Lu, H., Schmidt, M.A., Jensen, K.F., 2005. Lab Chip 5 (1), 23–29.

Marentis, T.C., Kusler, B., Yaralioglu, G.G., Liu, S., Haeggstrom, E.O., Khuri-Yakub, B.T., 2005. Ultrasound Med. Biol. 31 (9), 1265–1277.

Marmottant, P., Hilgenfeldt, S., 2003. Nature 423 (6936), 153–156.

Muller, K.J., Sukhorukov, V.L., Zimmermann, U., 2001. J. Membr. Biol. 184 (2), 161–170.

Neumann, E., Sowers, A.E., Jordan, C.A.E., 1989. Electroporation and Electro-fusion in Cell Biology. Plenum Press, New York.

Olofsson, J., Nolkrantz, K., Ryttsen, F., Lambie, B.A., Weber, S.G., Orwar, O., 2003. Curr. Opin. Biotechnol. 14 (1), 29–34.

Pihl, J., Karlsson, M., Chiu, D.T., 2005. Drug Discov. Today. 10 (20), 1377–1383. Shin, Y.S., Cho, K., Kim, J.K., Lim, S.H., Park, C.H., Lee, K.B., Park, Y., Chung,

C., Han, D.C., Chang, J.K., 2004. Anal. Chem. 76 (23), 7045–7052. Stone, H.A., Stroock, A.D., Ajdari, A., 2004. Ann. Rev. Fluid Mech. 36,

381–411.

Tsong, T.Y., 1991. Biophys. J. 60 (2), 297–306.

Xia, F.Q., Jin, W.R., Yin, X.F., Fang, Z.L., 2005. J. Chromatogr. A 1063 (1–2), 227–233.

數據

Fig. 1. Three dimensional cylindrical electrodes. (a) Sketch of leukocytes approaching a singe row of electrode
Fig. 2. Sketch of the entire experimental setup with the cell lysis microdevice.
Fig. 3 shows results of the computed iso-electric field surfaces for the two planar electrodes – pointed (Fig
Fig. 6. Micrographs of time evolution of leukocytes being lysed by 3-D electrodes under excitation of 10 V, 100 ␮s pulse (same as that in Fig

參考文獻

相關文件

Clear Cell carcinoma of palatal minor salivary gland harboring a novel EWSR1-ATF1 fusion gene: report of a case and review of the literature. Head

bility of the mobile system is the same as the call blocking For example, for a hexagonal planar layout with reuse dis- probability in a cell, Because traffic in

Before along with the evolution of time, the development of science and technology, the progress of cell phone, the cell phone that we used, from the beginning has the basic

In this paper, we study the local models via various techniques and complete the proof of the quantum invariance of Gromov–Witten theory in genus zero under ordinary flops of

Then, we tested the influence of θ for the rate of convergence of Algorithm 4.1, by using this algorithm with α = 15 and four different θ to solve a test ex- ample generated as

• Follow Example 21.5 to calculate the magnitude of the electric field of a single point charge.. Electric-field vector of a

Magnetic fields in a tokamak - the toroidal field is generated by external coils, poloidal by electric current in the

Optim. Humes, The symmetric eigenvalue complementarity problem, Math. Rohn, An algorithm for solving the absolute value equation, Eletron. Seeger and Torki, On eigenvalues induced by