Set the hold conditions of the measured value.
The hold function holds (retain) any value from the measured values during the specific time (sampling period), such as the maximum or minimum value.
Setting the Hold Mode
Set the hold mode of the measured value.
Item Setting item Setting value Description
Hold Hold mode Peak Holds the "maximum value" of the sampling period. Output changes after sampling, and this state is held until the next sampling ends.
Bottom Holds the "minimum value" of the sampling period. Output changes after sampling, and this state is held until the next sampling ends.
Peak to Peak Holds the "difference between the maximum value and the minimum value" of the sampling period. Mainly select this when detecting vibration, for example.
Output changes after sampling, and this state is held until the next sampling ends.
Auto Peak Holds the "maximum value" of the measurement result. The output changes every time the maximum value is updated.
Auto Bottom Holds the "minimum value" of the measurement result. The output changes every time the minimum value is updated.
Current
Setting the Output Conditions
ZW User's Manual 99
4
Settings for Function Explorer pane : [Bank] | [(Bank Data Name)] (double click)
Edit pane : [Task Settings] icon ( )
Task Settings window : [Hold]
Hold Hold mode Auto Peak to
Peak
Holds the difference between the maximum and minimum values of the measurement result. The output changes every time the maximum or the minimum value is updated.
Average Holds the "average of the measured value" of the sampling period. Output changes after sampling, and this state is held until the next sampling ends.
Sampling Holds the measured value the moment that control enters the sampling period. Output changes at the start of sampling, and this state is held until the next sampling is started.
Through (default value)
The hold function is not used. The measured value is always output.
1
Select the hold mode from [Hold mode].Setting items for required parameters are displayed depending on the selected hold mode.
Setting Triggers p.101
Setting a Trigger Delay p.104
The hold mode can also be set by the operating keys on the Sensor Controller.
Setting the Hold Mode p.318
Item Setting item Setting value Description
0
Setting the Output Conditions
100 ZW User's Manual
Hold clearing conditions
Held values can be cleared by the following operation.
• When FUN (adjustment) mode/RUN (operation) mode is switched
• When hold RESET is entered (32-pole extension connector, key)
• Hold RESET command is entered.
(Example) Peak
(Example) Auto Peak
If an abnormal measured value is obtained during sampling, the hold value is cleared. To not clear the hold value even if an abnormal measured value is obtained, set "KEEP" as the non-measurement setting.
Setting operation when measurement is not possible p.148
When the hold value is reset, the clamp level is output until the next hold value is applied.
The mode is switched between the FUN mode and RUN mode
FUN mode RUN mode
Sampling Sampling
Hold value Measurement value Clamp level
When RESET is executed
(32-pole extension connector/key input/command).
The mode is switched between the FUN mode and RUN mode
FUN mode RUN mode
Hold value Measurement value Clamp level
When RESET is executed
(32-pole extension connector/key input/command).
Important
Setting the Output Conditions
ZW User's Manual 101
4
Settings for Function Setting TriggersSet how measurement start to end timing is to be input.
Item Setting item Setting value Description Trigger
Set-tings
Trigger method External input Enter the trigger for the start of sampling in "TIMING input" for the 32-pole terminal block.
The period that the signal input to "TIMING input" is ON is the sampling period.
When a delay time is set, input OFF and sampling period end are not synchronized. End timing is after the specified sampling period has elapsed.
Selfup trigger The period in which the measured value exceeds the preset self-trigger level is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
When a delay time is set, the timing that the measured value falls below the self-trigger level and the sampling period end are not synchronized. End timing is after the sampling period has elapsed.
Selfdown trigger The period in which the measured value falls below the preset self-trigger level is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
When a delay time is set, the timing that the measured value exceeds the self-trigger level and the sampling period end are not synchronized. End timing is after the sampling period has elapsed.
Trigger level -999.999999 to 999.999999 [mm]
(default value: 0)
Set the self-trigger level when [SELF-UP] or [SELF-DOWN] are selected at [Trigger method].
Trigger hysteresis
0 to 999.999999 [mm]
(default value: 0.05% of measuring range)
Set the hysteresis width for the self-trigger when UP] or [SELF-DOWN] are selected at [Trigger method].
ON
Setting the Output Conditions
102 ZW User's Manual
Explorer pane : [Bank] | [(Bank Data Name)] (double click)
Edit pane : [Task Settings] icon ( )
Task Settings window : [Hold]
Set "Trigger hysteresis" according to the fluctuation of the measured value near the trigger level. Hysteresis is applied simultaneously with start of measurement to prevent chattering of the TIMING input.
1
Set the hold mode.Setting the Hold Mode p.98
2
Select the trigger method from [Trigger settings] - [Trigger method].3
Enter [Trigger level] and [Trigger hysteresis]when [Selfup trigger] or [Selfdown trigger]
are selected at [Trigger method].
The trigger can also be set by the operating keys on the Sensor Controller.
Setting Triggers p.319 Important
Note
Setting the Output Conditions
ZW User's Manual 103
4
Settings for Function Performing Hold with a Key InputThe TIMING/RESET inputs can be held on the Sensor Controller by setting as follows.
Setting the Hold Mode p.98
Explorer pane : [System] (double-click)
Edit pane : [Sensor Settings] icon ( )
Item Setting item Setting value Description
Sensor settings Timing/Reset key input OFF (default value)/ON Enables TIMING and RESET key inputs from the Sensor Controller.
1
Set the operating mode to the FUN mode.3-3 Switching operation modes p.61
2
Select ON/OFF from [Key input mode(Timing and Reset)].The timing/reset key input can also be set by the operating keys on the Sensor Controller.
Performing Hold with a Key Input p.320 Note
Setting the Output Conditions
104 ZW User's Manual
Setting a Trigger Delay
Set this to ignore measure values following TIMING input to avoid the influence of bounding or mechanical vibration when a device is started up.
The delay time (the delay between timing input and the start of sampling) and the sampling time can be set.
Explorer pane : [Bank] | [(Bank Data Name)] (double click)
Edit pane : [Task Settings] icon ( )
Task Settings window : [Hold]
Item Setting item Setting value Description
Trigger delay Trigger delay mode OFF (default value)/ON Turns the trigger delay ON/OFF.
Delay time 1 (default value) to 5,000 [ms]
Set the time from TIMING input ON up to start of sampling.
Sampling time 1 to 5,000 [ms]
(default value: 100ms)
Set to period during which sampling is to be performed.
1
Set the hold mode and trigger.Setting the Hold Mode p.98
Setting Triggers p.101
2
Select ON/OFF from [Trigger delay] - [Trigger delay mode].3
Enter [Delay time] and [Sampling time].Set so that the "delay time + sampling time" is shorter than the TIMING input interval. When the TIMING input turns ON again before the "delay time + sampling time" elapses, the TIMING input that is input later is ignored and is not reflected in sampling.
The trigger delay can also be set by the operating keys on the Sensor Controller.
Setting a Trigger Delay p.321 ON OFF
Measured values in this time period is included in the sampling.
Delay time Sampling time TIMING input
Sampling period
Measured value
Important
Note
Setting the Output Conditions