The following shows the timing charts when communication is performed with external devices.

Basic operation

Explanation of operations

(1) During each measuring cycle, the LEDs are lit and exposure is started.

(2) After the end of exposure, measurement starts.

(3) After the end of measurement, the judgment result is output.

(4) After the judgment result output, the analog output is updated.

Item Min. Max.

T0 Measuring cycle 0.5ms Depends on the set conditions

(0.5 to 10.0 ms)

T1 Exposure time 1 s Max. exposure time

(1 to 5000 s)

T2 Response time of output T0x2 T0x2

T3 Response time of analog output 0.1 ms

(2) (3)

Judgment output HIGH/PASS/LOW) Analog output

(4)

Exposure

T0 (1)

T1 T2 T3

ON OFF

Parallel I/O connection

ZW User's Manual 153

6

Communications with External Devices Hold (peak/bottom/peak to peak/average)

Explanation of operations (1) The TIMING input is turned ON.

(2) During the TIMING input minimum time, when the TIMING input is ON, sampling is started and the BUSY output is turned ON.

(3) The TIMING input is turned OFF.

(4) After the TIMING input turns OFF, sampling is ended and the judgment results are output. The BUSY output is also turned OFF.

(5) After the judgment result output, the analog output is updated.

(6) The RESET input is turned ON. If the RESET input is turned ON during the RESET input minimum time, the measured value is reset.

(7) The judgment result is reset.

(8) After the judgment result reset, the analog output is reset.

(9) The RESET input is turned OFF.

When the setting for non-measurement is "CLAMP", if the sampling value is an abnormal value or an undetermined value *, sampling is not executed. If sampling has been started, it is stopped. The output value is as follows.

• Hold the clamp value.

• To start and continue sampling even if a sampling value is an abnormal value or an undetermined value, set "KEEP"

as the non-measurement setting.

*: After the start of measurement, if measurement results are not obtained the number of times required to take the average, the measurement result is not applied.

Item Min. Max.

T3 Response time of analog output 0.1 ms

T4 TIMING input/RESET input minimum time 3ms+T0

T5 Input response time 2ms+T0 3ms+T0 × 2

(2) (3)

Parallel I/O connection

154 ZW User's Manual

Hold (auto peak/auto bottom/auto peak-to-peak)

Explanation of operations

(1) The peak value is updated and the judgment result is output.

(2) After the judgment result output, the analog output is updated.

(3) The RESET input is turned ON. If the RESET input is turned ON during the RESET input minimum time, the measured value is reset.

(4) The judgment result is reset.The BUSY output is turned OFF.

(5) After the judgment result reset, the analog output is reset.

(6) The RESET input is turned OFF.

(7) The BUSY output is turned ON.

When the setting for non-measurement is "CLAMP", if the sampling value is an abnormal value or an undetermined value *, sampling is not executed. If sampling has been started, it is stopped. The output value is as follows.

• Hold the clamp value.

• The BUSY signal is turned OFF.

• To start and continue sampling even if a sampling value is an abnormal value or an undetermined value, set "KEEP"

as the non-measurement setting.

*: After the start of measurement, if measurement results are not obtained the number of times required to take the average, the measurement result is not applied.

Item Min. Max.

T3 Response time of analog output 0.1 ms

T4 TIMING input/RESET input minimum time 3ms+T0

T5 Input response time 2ms+T0 3ms+T0 × 2

(1) (2)

Parallel I/O connection

ZW User's Manual 155

6

Communications with External Devices Hold (sampling)

Explanation of operations (1) The TIMING input is turned ON.

(2) During the TIMING input minimum time, when the TIMING input is ON, sampling is started and the BUSY output is turned ON. The measurement result is sampled and the judgment result is output.

(3) After the judgment result output, the analog output is updated.

(4) The RESET input is turned ON. If the RESET input is turned ON during the RESET input minimum time, the measured value is reset.

(5) The judgment result and is reset.

(6) After the judgment result output, the analog output is reset.

(7) The RESET input is turned OFF.

When the setting for non-measurement is "CLAMP", if the sampling value is an abnormal value or an undetermined value *, sampling is not executed. The output value is as follows.

• Hold the clamp value.

• The BUSY signal is not turned ON.

*: After the start of measurement, if measurement results are not obtained the number of times required to take the average, the measurement result is not applied.

Item Min. Max.

T3 Response time of analog output 0.1 ms

T4 TIMING input/RESET input minimum time 3ms+T0

T5 Input response time 2ms+T0 3ms+T0 × 2

(1) (2) (3) (4) (5) (6) (7)

Parallel I/O connection

156 ZW User's Manual

Bank Switching

Explanation of operations

(1) The BANK_SEL signal is switched to the bank number to switch to.

(2) After the input response time, the measurement stops and the BUSY output is turned ON, then the bank switching operation is started.

(3) After bank switching ends, the BUSY output is turned OFF and the BANK_ OUT output is switched.

(4) Measurement is restarted and the ENABLE output is turned ON.

LED OFF

Explanation of operations (1) The LED OFF input is turned ON.

(2) After the LED OFF input is turned ON, the LEDs are turned OFF.

(3) After the LEDs are turned OFF, the ENABLE output is turned OFF.

(4) The LED OFF input is turned OFF.

(5) After the LED OFF input is turned OFF, the LEDs are turned ON.

(6) After the LEDs are lit up, measurement is restarted and the ENABLE output is turned ON.

Item Min. Max.

T7 Input response time 200 ms

T8 Bank switching time 100 ms

T9 Measurement start response time 3 × T0 Depends on the set conditions

Item Min. Max.

T9 Measurement start response time 3 × T0 Depends on the set conditions

T10 LED OFF input minimum time 100 s

T11 Time of receiving LED OFF input 100 s

T12 ENABLE response time after LEDs are turned OFF 2 × T0

(1) (2)

Measurement Measurement Clamp Measurement

(4)

Turn ON Turn OFF Turn ON

(6)

Parallel I/O connection

ZW User's Manual 157

6

Communications with External Devices Zero reset

Explanation of operations (1) The ZERO input is turned ON.

(2) After the ZERO input time, the ZERO input is turned OFF.

(3) After the ZERO input is turned OFF, the zero reset is executed and the judgment results reflected in the measurement results are output.

(4) The ZERO input is turned ON.

(5) After at least the cancel time of ZERO input has passed, the zero reset is cancelled.

Operating Mode Switching

Explanation of operations

(1) After the mode is switched from RUN to FUN mode, the BUSY output and ENABLE output are turned OFF.

The judgment outputs all go OFF.

(2) The response time of analog output after the BUSY output is turned ON, the analog output is output clamped.

(3) After the mode is switched from the FUN mode to the RUN mode, the BUSY output is turned OFF.

(4) Measurement is restarted and the ENABLE signal is turned ON, then the measurement results are output.

Item Min. Max.

T7 Input response time 2ms+T0 3ms+T0 × 2

T13 ZERO input time 50 ms 0.8s

T14 ZERO input cancel time 1s

Item Min. Max.

T3 Response time of analog output 0.1 ms

T9 Measurement start response time 3 × T0 Depends on the set conditions

(1) (2) (3) (4)

Parallel I/O connection

158 ZW User's Manual

Internal logging

Explanation of operations

(1) The LOGGING input is turned ON.

(2) After the LOGGING input is turned ON, the internal logging is started.

(3) The LOGGING input is turned OFF.

(4) After the LOGGING input is turned OFF, the internal logging is ended.

Item Min. Max.

T15 LOGGING input minimum time 3ms+T0

T16 Input response time 2ms+T0 3ms+T0 × 2

T16 LOGGING input

Operation status

T16

Logging stopped

state Logging in progress Logging stopped state T15

ON OFF

(1) (2) (3) (4)

EtherCAT Connection

ZW User's Manual 159

6

Communications with External Devices