Chapter 3: Electrical Characteristics of Back-Gate TFT Gas Sensor
3.2 The Effects of Measurement Environment on Poly-Si TFTs
Typical characteristics of the planar TFTs with various film thickness measured
under different test environments are shown in Figures 3-3(a)~(d). In these
measurements, the device was first tested in normal ambient which had a moisture
level of 42%, and then in the closed test chamber (see the description in Chap. 2)
which was pumped down to 0.2 Torr. After the second measurement performed in
vacuum, the vacuum chamber was injected with high purity N2 (99.999%) to increase
the pressure to around 1 atm, and then the third measurement was performed. Finally
the chamber was open to expose the device to the air again. We then repeated the
measurements to check if the transfer characteristics of the device recovered to the
original characteristics.
In the figures we can find that there is a big change in the Id-Vg curves as the test
environment is switched from atmosphere to vacuum. When measured in vacuum, the
drain current became low and insensitive to the variation of gate voltage. As N2 was
injected into the chamber, the Id-Vg curve remained the same as that measured in
vacuum.
When the device was put back to the normal air, the I-V characteristics would
recover, but the extent depended strongly on the film thickness. As can be seen in
Figure 3-3(a), the TFT with 70Å-thick channel almost return to the original
characteristics while the other devices with thicker channel do not t recover
completely (Figures 3-3(b) ~3-3(d)).
Note that the high purity N2 is very dry and contains less moisture, unlike the
normal air which has a relative humidity of 42 %. Thus we postulate that the above
difference in I-V characteristics is due to the existence of moisture and the interaction
between moisture and the poly-Si films. Note the moisture contains H-related species
(such as H or OH) which can help passivate the defects contained in the poly-Si [19].
The schematic illustration about the passivation of defects in the grain boundaries
with H from the moisture in the ambient is shown in Figure 3-4 (a). In the figure and
following discussion, for simplicity, we assume H is the main species responsible for
the passivation of the defects. Note that the fresh devices should contain specific
amount of H throughout the film since the last step in device fabrication was a
de-ionized water rinse after the removal of photoresist with H2SO4/H2O2. According
to the water passivation effect [14], some H species should remain in the films even
when the device was dried. At vacuum, the amount of active defects in the poly-Si
increases because the vacuum tends to drive out the H from the grain boundaries and
destruct the bonding, such as Si-H, leaving the defects unpassivated (see Figure 3-4).
The amount of dangling bonds increases massively and affects the electrical
characteristics considerably. As the high-purity N2 is injected into the chamber such
characteristics are retained due to the dry ambient. After returning to the air, the Id-Vg
curve is restored due to the diffusion of H from the moisture contained in the normal
air. However, the concentration of H decreases with increasing depth due to the
diffusion process. The device with70 Å-thick channel restores soon due to its shallow
channel (Figure 3-5(a)). However, for devices with thicker channel it needs longer
time to recover completely. The gated channel (bottom path shown in Figure 3-2) thus
retains a high amount of un-passivated defects, as shown in Figure 3-5(b). This
explains why the characteristics cannot recover completely in Figures 3-3(b)~(d) as
the test environment returns to the normal air.
3.2.2 Id-time Measurements
In this Id-time measurement, the setup of measurement environment is the same
as that described in last sub-section. The measurement scheme is stated in Sec. 2.3,
and the Vg is set at 0.5V. In Figure 3-6, the Id of a TFT with 300Å channel thickness
initially increased as measured under normal atmosphere. To check this phenomenon
we compared the transfer curves of the device before and 60 sec after the above
Id-time measurement. The results are shown in Figure 3-7. We can see that the
threshold voltage becomes smaller while the SS remains unchanged after the Id-time
measurement. To confirm and figure out the mechanism, we made a series of
measurements performed on one another device with the results are shown in Figure
3-8. First, we measured the transfer curve of the device with 300Å channel thickness
under normal air (curve (1) in Figure 3-8). And then we made the Id-Time
measurement (Vg = 3.7V > 0V ) (curve (1) in Figure 3-9) on the device for 2100 sec,
and then the transfer curve was measured again (curve (2) in Figure 3-8). As the trend
shown in Figure 3-7, the threshold voltage of the curve (2) in Figure 3-8 shifts
leftward. And then we made Id-Vg measurement with gate voltage sweeping from 0
to -5V, as shown in Figure 3-10. After that we measured the curve 3) in Figure 3-8
and found the threshold voltage shifts slightly rightward with respect to the curve (2)
shown in the same figure. Before measuring the curve (4) in Figure 3-8, we made
another Id-Time measurement with Vg=-3.7V (< 0V) (curve (2) in Figure 3-9). The
curve (4) shifts leftward and becomes close to the original curve (1) in Figure 3-8.
From the above experimental results we postulate the instability in the transfer
curves of the test devices as measured in the normal ambient is due to the action of
mobile ions, like sodium or potassium ions, presenting in the oxide. This is a well
known issue for old MOSFET technology [20]. Since there is no passivation dielectric
covering the devices, these contamination species are likely to appear in the test
samples. As a positive gate voltage is applied for a sufficient long time, these ions
tend to accumulate near the oxide/channel interface and the threshold voltage is thus
decreased (see Figure 3-11(a)). In opposite situation as the gate bias is negative, these
ions would be attracted toward and accumulate near the oxide/gate interface, leading
to an increase in threshold voltage.
Now let’s return to discuss the results shown in Figure 3-6. While the air was
pumped out to vacuum the text chamber, the Id dropped drastically which is
consistent with the results shown in Figure 3-3(b). After the measurement
environment turned back to atmosphere, the current increased again. The air has
obvious passivation to the device.