On: 25 April 2014, At: 11:34 Publisher: Taylor & Francis

Informa Ltd Registered in England and Wales Registered Number: 1072954 Registered office: Mortimer House, 37-41 Mortimer Street, London W1T 3JH, UK

Aerosol Science and Technology

Publication details, including instructions for authors and subscription information: http://www.tandfonline.com/loi/uast20

Laboratory and Field Tests of a Novel Three-Stage

Personal Dust Sampler for Sampling Three Dust

Fractions Simultaneously

Chuen-Jinn Tsai a , Cheng-Sheng Chang a , Sheng-Chieh Chen a , Penny Chen a , Tung-Sheng Shih b , David Y.H. Pui c , Vladimir V. Karasev d , Andrei A. Onischuk d & Shou-Nan Li e a

Institute of Environmental Engineering, National Chiao Tung University , Hsinchu, Taiwan b

Institute of Occupational Safety and Health, Council of Labor Affairs, Executive Yuan , Taipei, Taiwan

c

Particle Technology Laboratory, Mechanical Engineering, University of Minnesota , Minneapolis, Minnesota, USA

d

Siberian Branch of the Russian Academy of Sciences, Institute of Chemical Kinetics and Combustion , 630090, Novosibirsk, Russia

e

Energy and Environment Research Laboratories, Industrial Technology Research Institute (ITRI) , Hsinchu, Taiwan

Published online: 24 Oct 2011.

To cite this article: Chuen-Jinn Tsai , Cheng-Sheng Chang , Sheng-Chieh Chen , Penny Chen , Tung-Sheng Shih , David Y.H. Pui , Vladimir V. Karasev , Andrei A. Onischuk & Shou-Nan Li (2008) Laboratory and Field Tests of a Novel Three-Stage Personal Dust Sampler for Sampling Three Dust Fractions Simultaneously, Aerosol Science and Technology, 42:1, 86-95, DOI: 10.1080/02786820701697556

To link to this article: http://dx.doi.org/10.1080/02786820701697556

PLEASE SCROLL DOWN FOR ARTICLE

Taylor & Francis makes every effort to ensure the accuracy of all the information (the “Content”) contained in the publications on our platform. However, Taylor & Francis, our agents, and our licensors make no

representations or warranties whatsoever as to the accuracy, completeness, or suitability for any purpose of the Content. Any opinions and views expressed in this publication are the opinions and views of the authors, and are not the views of or endorsed by Taylor & Francis. The accuracy of the Content should not be relied upon and should be independently verified with primary sources of information. Taylor and Francis shall not be liable for any losses, actions, claims, proceedings, demands, costs, expenses, damages, and other liabilities whatsoever or howsoever caused arising directly or indirectly in connection with, in relation to or arising out of the use of the Content.

This article may be used for research, teaching, and private study purposes. Any substantial or systematic reproduction, redistribution, reselling, loan, sub-licensing, systematic supply, or distribution in any

form to anyone is expressly forbidden. Terms & Conditions of access and use can be found at http:// www.tandfonline.com/page/terms-and-conditions

(2)

Aerosol Science and Technology, 42:86–95, 2008 CopyrightcAmerican Association for Aerosol Research ISSN: 0278-6826 print / 1521-7388 online

DOI: 10.1080/02786820701697556

Laboratory and Field Tests of a Novel Three-Stage Personal

Dust Sampler for Sampling Three Dust Fractions

Simultaneously

Chuen-Jinn Tsai,

1

Cheng-Sheng Chang,

1

Sheng-Chieh Chen,

1

Penny Chen,

1

Tung-Sheng Shih,

2

David Y.H. Pui,

3

Vladimir V. Karasev,

4

Andrei A. Onischuk,

4

and Shou-Nan Li

5

1Institute of Environmental Engineering, National Chiao Tung University, Hsinchu, Taiwan

2Institute of Occupational Safety and Health, Council of Labor Affairs, Executive Yuan, Taipei, Taiwan 3Particle Technology Laboratory, Mechanical Engineering, University of Minnesota, Minneapolis, Minnesota, USA

4

Siberian Branch of the Russian Academy of Sciences, Institute of Chemical Kinetics and Combustion, 630090, Novosibirsk, Russia

5

Energy and Environment Research Laboratories, Industrial Technology Research Institute (ITRI), Hsinchu, Taiwan

This study designed and calibrated a novel three-stage personal dust sampler for sampling inhalable, thoracic, and respirable dust fractions simultaneously. The sampler has an annular inlet as the first stage for inhalable dust sampling, two impactors in the second and third stages to classify thoracic and respirable dusts, respec-tively, and a final filter. Laboratory calibration tests using monodis-perse liquid and solid particles showed that with 100 ppi (pores per inch) PUF (porous polyurethane foam) substrates and at the flow rate of 3.2 L/min, the sampling efficiency curves of both impactor stages matched with the ISO/CEN/ACGIH thoracic and respirable sampling criteria, respectively. The sampler also agreed with the inhalable criterion for particles smaller than 17 µm while the devi-ation increases with increasing particle diameter with a maximum of 28% for 27 µm particles. It was also found that collection ef-ficiency curve for solid particles was similar to that of liquid par-ticles, indicating that there was no solid particle bounce from the PUF substrates for both impactors.

This study also compared inhalable, thoracic, and respirable dust concentrations measured by the present three-stage personal dust samplers with those of the Respicon samplers at three differ-ent workplaces. Without using a correction factor of 1.5 for the extrathoracic dusts of the Respicon, field results showed that the present three-stage sampler measured three dust fractions compa-rable to those of the Respicon samplers. The inhalable dust concen-trations of the two samplers differed within 5% while the thoracic and respirable concentration differed by less than 22%.

Received 12 May 2007; accepted 20 September 2007.

The authors would like to thank for the financial support of this project the Taiwan Institute of Occupational Safety and Health (IOSH94-A102).

Address correspondence to Chuen-Jinn Tsai, Institute of Environ-mental Engineering, National Chiao Tung University, Hsinchu, Taiwan. E-mail: cjtsai@mail.nctu.edu.tw

INTRODUCTION

Particle size-selective sampling in the workplaces is per-formed to measure the worker’s exposure to particles at dif-ferent regions of the human respiratory system, and to find out if the Threshold Limit Values (TLVs) are exceeded. The ISO (International Organization for Standardization)/CEN (Comit´e Europ´een de Normalisation)/ACGIH (American Conference of Government Industrial Hygienists) proposed three criteria for dust sampling, including the inhalable fraction (IF, the efficiency of entry of particles into the nose or mouth), the thoracic frac-tion (TF, the fracfrac-tion of particles that will pass beyond the larynx and reach the thorax or chest) and the respirable fraction (RF, the fraction of particles which can reach the gas exchange region of the lung). The size ranges of the IF, TF, and RF are defined in the range from 0 to 100µm (cutoff diameter = 80–100 µm), 0 to 30µm (cutoff diameter = 10 µm) and 0 to 10 µm (cutoff diameter= 4 µm), respectively.

Many dust samplers were developed for matching the crite-rion of one of the dust fractions only. Samplers such as the 10 mm nylon cyclone (Seltzer et al. 1971), the SKC Aluminum cyclone (SKC Inc., Eighty Four, PA, USA), the IOSH cyclone (Tsai et al. 1997), and the single nozzle impactor with polyurethane foam substrate (Huang et al. 2005) were developed for respirable dust sampling. Among them, the IOSH cyclone was designed with a larger inner diameter of 18 mm than the 10 mm nylon cyclone to reduce the effect of deposited particles in the cyclone on particle collection efficiency. The GK2.69 cyclone (Kenny and Gussman 1997; Maynard 1999) and IOM thoracic samplers (Aitken et al. 1995) are used for thoracic dust sampling, in which the latter was designed to separate out the extrathoracic fraction using a foam pre-separator, allowing the thoracic fraction to be collected

86

(3)

onto a 37 mm diameter after filter. For inhalable dust sampling, the IOM sampler (SKC Inc., Eighty Four, PA, USA) designed by Mark and Vincent (1986) and the Button sampler (SKC Inc., Eighty Four, PA, USA) designed by Kalatoor et al. (1995) are widely adopted.

However, all of the above samplers can sample one dust frac-tion only. There are few samplers that have been developed to measure the inhalable, thoracic, and respirable fractions simulta-neously. TSI Respicon originally developed by Dunkhorst et al. (1995) and commercialized by TSI (Model 8522, TSI Inc., St. Paul, MN, USA) and Helmut Hund GmbH (Wetzlar, Germany) is one of them. This sampler has two virtual impactors in the first and second stages with cutoffs at 4 and 10 µm, respec-tively, and an after filter in the third stage to collect particles larger than 10µm. The operating total flow rate of the Respi-con is recommended to be 3.11± 2% L/min. For the internal flow rates, it is 2.66± 3% L/min through stage 1, 0.33 ± 5% L/min through stage 2, and 0.11 ± 5% L/min through stage 3. The aerosols collected on the stage 1 filter represent the res-pirable dust, those collected on the stage 1 plus stage 2 represent the thoracic dust, and the sum of all three stages represents the inhalable dust. In order to match the inhalable criterion, TSI reported that a correction factor of 1.5 has to be used for the mass of the third stage sample (extrathoracic dust concentration,

Cexth) as

CI = CT+ Cexth, [1]

and

Cexth = (CI − CT)× 1.5, [2]

where Cexth , CI, and CTare the corrected extrathoracic dust

con-centration, corrected inhalable dust concon-centration, and particle concentration in the thoracic fraction, respectively.

Li et al. (2000) tested six inhalable dust samplers in a small wind tunnel at wind speeds of 0.55 and 1.1 m/s using monodis-perse solid particles with aerodynamic diameter ranging from 5 to 68µm. The six samplers were the Respicon, IOM, a seven-hole, a conical inhalable sampler, a prototype button sampler and a closed-face 37-mm cassette. They found that with the annular slit inlet of the Respicon, the effect of orientations on collection efficiency did not occur. The sampler was found to match with the definition curve of the inhalable fraction quite accurately without applying the factor of 1.5. According to these results, TSI recommended that Respicon users discontinue applying the correction factor of 1.5 when calculating the inhalable fraction (TSI 2001). However, under calm wind condition, the Respicon showed sampling deficiency with increasing diameter for par-ticles greater than 6.0 µm in Feather and Chen (2003) (wind speed<0.015 m/sec), and for particles greater than 30 µm in Koch et al. (1999) (wind speed<0.2 m/sec).

Traditionally, cyclones were used for respirable or thoracic dusts sampling such as the SKC aluminum cyclone and the GK

2.69 cyclone (BGI Inc., Waltham, Mass.), respectively. Their penetration curves are less sharp than those of impactors and can match with the respirable criterion. Cyclones are usually designed for only one cut size while impactors are easily de-signed as a cascade impactor for size-classified particle sampling (Huang et al. 2005). Marple (1978) designed a single stage im-pactor with multiple round nozzles of different sizes. They found that the impactor penetration curve could approximate the res-pirable curve of ACGIH when three different nozzle sizes were used in the impactor design. Kavouras and Koutrakis (2001) used porous polyurethane foams (PUFs) as the impaction sub-strates in the inertial impactors. They found the use of PUFs re-sulted in significant changes in the collection efficiency curves, which were shifted to the left of the curves of the flat-plate im-pactors. The curves were less sharp than those of the flat-plate impactors. Huang et al. (2005) further demonstrated that using the PUF (Foamex Inc., Pennsylvania, USA) as the impaction substrate, a single round nozzle impactor could be used for res-pirable dust sampling. As an extension of this work on the single stage impactor, it is therefore desirable to design a cascade im-pactor which uses PUF as impaction substrates to classify the three dust fractions simultaneously.

In this study, a three-stage impactor was designed and tested for sampling three dust fractions simultaneously. Monodisperse liquid and solid particles were used to evaluate the sampler in a calm air chamber in the laboratory. Field sampling was con-ducted in three different workplaces to compare the three dust concentrations of the three-stage sampler and the Respicon. In addition, the IOSH cyclone was used for the comparison of res-pirable concentration with that of the Respicon and the present sampler.

The Novel Three-Stage Sampler

The novel three-stage sampler uses a fixed sampling flow rate of 3.2 L/min throughout the three stages. As shown in Fig-ure 1(b), the sampler has an annular inlet as the first stage, a thoracic impactor as the second stage and a respirable impactor as the third stage followed by an after filter. The outer diameter and total weight of the sampler are 5.0 cm and 183 g, respectively. Figure 1a shows the top view of the top disc of the annular inlet which has a diameter of 5.0 cm. The disc has three ears used for the purpose of assembling the sampler. In the downstream of the annular inlet, an impaction substrate (PUF1) following a large round nozzle of 1.8 cm in diameter is used to collect part of extrathoracic particles. Inlet of this annular form was found to have high sampling efficiency up to 15µm and 15 km/hr windspeed for an atmospheric PM-10 sampling inlet (Liu and Pui 1981). Figure 1b also shows the schematic diagram of PUF1, PUF2 and PUF3, in which the six holes in each of PUF1 and PUF2 are used to accommodate the nozzles of the thoracic and respirable impactors, respectively.

The thoracic impactor has a cutoff aerodynamic diameter of 10µm and the second PUF (PUF2) as the impaction substrate.

(4)

88 C.-J. TSAI ET AL.

FIG. 1. Schematic diagram of the novel personal 3-stage dust sampler. (a) Top view of the inlet top disc. (b) Schematic diagram of the present 3-stage sampler, PUF1, PUF2, and PUF3.

The respirable impactor has a cutoff aerodynamic diameter of 4µm and the third PUF (PUF3) as the substrate. The after filter (AF) was used to collect respirable particles. The impactors of the sampler were designed with the square root of the cutoff Stokes number,√Stk50, of 0.39, which was determined

accord-ing to Huang et al. (2005). Stk50is defined as

Stk50 = ρpd2

pa50U C

9µDo [3]

where d2

pa50is the cutoff aerodynamic diameter (m), U is the gas

velocity at nozzle (m/s) and Dois the nozzle diameter (m). The

ratio of the jet-to-plate distance was designed to be 1.0 for these two impactors each of which has six nozzles with the diameter of 3.5 and 2.0 mm for the second and third stages, respectively. The porosity and thickness of all three PUFs are 100 ppi (pores per inch) and 6 mm, respectively. The diameter of the first and second foams is 47 mm while the third foam is 22 mm.

Particles collected on the AF represent the respirable dust. The sum of particles collected on the PUF3 and the AF represents the thoracic dust while the sum of all three PUFs and the AF represents the inhalable dust. That is, the mass concentration of the respirable (CR), thoracic (CT), and inhalable (CI) dusts of

the present three-stage sampler can be calculated as

CR = MAF

Qt , [4]

CT = MPUF3+ MAF

Qt , [5]

CI = MPUF1+ MPUF2+ MPUF3+ MAF

Qt , [6]

where Q is the flow rate (3.2 L/min), t is the sampling time (min) and MPUF1, MPUF2, MPUF3, and MAFare particle mass (µg) on

PUF1, PUF2, PUF3, and the after filter, respectively.

METHODS

Calibration in the Laboratory

In this study, both monodisperse liquid oleic acid (OA) parti-cles tagged with uranine dye and standard fluorescent polymer microspheres (Duke Scientific, Palo Alto, CA) between 1.8– 20.2 and 1.9–27.7µm, respectively, were used to calibrate the present three-stage sampler at a calm air chamber (wind speed

<0.01 m/sec). As shown in Figure 2, the chamber has an

in-ner diameter of 50 cm and a small fan was used for particle mixing. Before calibrating the three-stage sampler, the concen-tration uniformity of the generated particles in the chamber was examined first by five reference samplers, each consisted of a vertical tube and a filter holder designed based on the criterion of Agarwal and Liu (1980). The five reference samplers were located uniformly in the central part of the chamber to cover the test zone, which is a circular region of 12 cm in diameter. OA particles of 20.2µm and fluorescent polymer microspheres at 27µm in aerodynamic diameter were generated by a vibrating orifice monodisperse aerosol generator (VOMAG, TSI Model 3450) and a small scale powder disperser (TSI Model 3433), respectively, to conduct the uniformity test. The sampling flow rate was 2 L/min through each sampler and the sampling time for OA particles and fluorescent microspheres was 0.5 and 5 hours, respectively. After sampling, the glass fiber filters were soaked in 0.001 N NaOH (for OA particles) or xylene (for fluorescent mi-crospheres) (Glenny et al. 1993) and were ultra-sonicated for 30 minutes, allowing the fluorescence material to dissolve in the so-lutions. The uranine mass of particles collected on the filters was measured by a fluorometer (Turner Designs Model 10-AU-005, Cincinnati, OH, USA). Results showed that the sampled particle concentrations of the five samplers differed within 10%. That is, the chamber is uniform across the test zone of the chamber.

Calibration of the three-stage sampler was conducted inside the test zone in which a reference sampler was collocated with

(5)

FIG. 2. Experimental setup for calibrating the 3-stage sampler.

a three-stage sampler as shown in Figure 2. An Aerodynamic Particle Sizer (APS, TSI Model 3321) was used for checking the size of particles produced by the VOMAG. OA particles from 1.8–20.2 µm in aerodynamic diameter were generated by the VOMAG; ammonium sulfate ((NH4)2SO4) particles from

0.8-7.2 were generated by an atomizer (Sono-Tek 8700, Sono-Tek Inc., N.Y.); fluorescent microspheres from 1.9–5µm and 6.2– 27.7µm in aerodynamic diameter were generated by another atomizer (Retec X-70/N, Cavitron Corp., Portland, OR) and a powder disperser, respectively. After sampling, particle uranine mass on the three PUFs and the after filter of the three-stage sampler was determined in the same way as that conducted in the chamber uniformity test. To determine particle wall loss, cotton swabs were used to recover deposited particles at the in-ner surfaces, which were also dissolved in 0.001 N NaOH (for OA particles) or xylene (for fluorescent microspheres) for fur-ther determination of fluorescence concentration. It was found that most particle loss occurred at the bottom of the annular slot mainly due to gravitational settling, while wall loss was negli-gible at other parts of the sampler (<0.5% for all sizes particles tested).

The RF, TF, and IF of the present three-stage sampler are related to the aspiration efficiencies,ηasp, which is defined as

ηasp= Ct

Cr e f, [7]

where Cref is the uranine concentration of the reference

sam-pler and Ct is the sum of CPUF1, CPUF2, CPUF3, and CAF as

well as the uranine concentration of the deposited particles on the bottom disc of the annular inlet, Cs, due to gravitational

settling. CPUF1, CPUF2, CPUF3, and CAF are the uranine

con-centrations of PUF1, PUF2, PUF3, and the after filter, respec-tively.ηaspis related to the inlet efficiency,ηinlet, which can be written as

ηinlet= ηaspηtrans, [8] whereηtransis the inlet transmission efficiency and can be written as

ηtrans=Ct− Cs

Ct , [9]

(6)

90 C.-J. TSAI ET AL.

As mentioned earlier, Cscan be obtained from the experiment. ηtransis then determined accordingly. Combining Equations [7– 9],ηinletis found to be Ct−Cs

Cref which can be obtained from the

experiment directly. By connecting the outlet tube of the refer-ence sampler to the APS, Crefcan be read from the APS. Ct−Cs

is obtained when the annular inlet of the three-stage sampler is connected to APS. Onceηi nletandηtransis obtained,ηaspis then determined according to Equation [8].

The theoretical transmission efficiency,ηtrans,theo, due to grav-itational settling was also calculated in the present study, which is equal to 1 minus theoretical settling efficiency,ηs,theo. As a particle enters the annular inlet of the present sampler, it starts to settle toward the bottom plate of the annular inlet with its terminal settling velocity, Vts (m/s). ηs,theo can be calculated

as

ηs,theo= h

h , [10]

where h is the distance between the top and bottom plates andh is the particle settling distance in z direction inside the annular inlet. During a time interval, dt, particle settling distance, dh, in z direction can be calculated as

dh= Vtsdt, [11]

where dt depends on the gas horizontal speed, u(r ), which is a function of the radial position, r (r = 0 representing the center of the inlet bottom plate), and can be calculated as

dt= dr

u(r ), [12]

where the horizontal gas speed, u(r ), will increase with decreas-ing radial position and can be calculated as

u(r )= Q

2πrh. [13] Combining Equations [11–13], the following equation for dh can be obtained:

dh =2πrhVts

Q dr [14]

Integrating Equation [14] for r from R2to R1,ηtrans,theois then

obtained as

ηtrans,theo= 1 − ηs,theo = 1 − πVts

Q



R12− R22 [15] where R1and R2are the radius of the top disc, 2.5 cm, and inner

radius of the bottom plate, 0.9 cm, respectively.

The RF, TF, and IF of the present three-stage sampler were determined experimentally as RF= CAF Cref = CAF Ct × Ct Cref, [16] TF= CPUF3+ CAF Cref = CPUF3+ CAF Ct × Ct Cref, [17] IF= CPUF1+ CPUF2+ CPUF3+ CAF

Cr e f

= CPUF1+ CPUF2+ CPUF3+ CAF Ct

× Ct

Cref = ηinlet. [18]

Field Sampling

In field sampling study, two three-stage samplers, two IOSH cyclones and two Respicons were mounted on a life-sized rotat-ing mannequin simultaneously as shown in Figure 3 to conduct the tests in three factories, including a lead powder, a lead-acid battery, and a casting factory. The height of the samplers was 1.5 m from the ground. The sampling locations were near a ball mill in the lead powder factory, the plate cutting area of the lead-acid battery plant, and the pouring area of the casting factory. The wind speed in all three sampling sites was below 0.1 m/s. To ensure uniform particle concentration, the mannequin was rotated at the speed of 1 turn per 10 minutes. Sampling lasted for 4–6 hours and at least six effective samples were obtained in each workplace. Equations [4–6] were used to determine the three dust concentrations measured by the three-stage sampler.

It was found that the environmental condition in the weighing room and the condition chamber (24-h conditioning before and after sampling) must be controlled carefully to obtain accurate weighing data for PUFs since they absorb water vapor readily. In this study, temperature was controlled between 22± 0.4◦C, and relative humidity (RH) was kept at 35± 3% in the weighing room and condition chamber. The precision of weighing was determined to be 5µg by repeated weighing for at least 5 times.

RESULTS AND DISCUSSION Laboratory Test

Figure 4 shows the Inhalable fraction (or inlet efficiency) of the present three-stage sampler for particles from 0.8 to 27µm. The current sampler agrees with the inhalable criterion (solid curve) very well for particles from 0.8 to 17µm with a maximum deviation of 8%. The deviation increases with increasing particle diameter with a maximum of 28% for 27 µm particles. This difference is mainly caused by low transmission efficiency due to gravitational settling in the inlet. Similar inhalable fraction for liquid and solid particles indicates there is no solid particle bounce in the present sampler.

(7)

FIG. 3. Schematic diagram of the manikin and the location of samplers.

Under calm wind condition, the inhalable fraction of the present sampler shows a similar decreasing trend with increas-ing diameter as that of the Respicon in Feather and Chen (2003) (wind speed<0.015 m/sec), and Koch et al. (1999) (wind speed

<0.2 m/sec).

FIG. 4. Comparison of the inhalable fraction between the experimental data of the 3-stage sampler and the criteria.

The experimental transmission efficiency is shown in Figure 5, together with the theoretical transmission efficiency. Good agreement in Figure 5 for both flow rates of 1.6 and 3.2 L/min indicates the deposition efficiency of the annular inlet can be predicted theoretically. According to Equation [15],ηtrans can

FIG. 5. Comparison of the theoretical and experimental inlet transmission efficiencies of the 3-stage sampler.

(8)

92 C.-J. TSAI ET AL.

FIG. 6. Aspiration efficiency of the 3-stage sampler.

be increased when the flow rate or R2(inner radius of the

bot-tom plate) is increased or R1 (the radius of the top plate) is

decreased. That is, Equation [15] can be used to design a better annular inlet which can increase theηtransas well as the IF for large particles to match with the inhalable, curve better in the future.

The experimental aspiration efficiency,ηasp, versus aerody-namic diameter is calculated from Equation [7] and shown in Figure 6. It is seen theηaspis close to 100% for particles smaller than 3µm while ηaspis about 80–90% for particles greater than 3

µm. There are no significant differences for ηaspbetween liquid and solid particles.

Figure 7 shows the comparison of the respirable and thoracic fractions between the experimental data of the three-stage sam-pler and the criteria of the ISO/CEN/ACGIH (shown as curves). The experimental respirable and thoracic dust fractions were calculated by Equations [16] and [17], respectively. For thoracic dust, the present sampler oversamples slightly with a maximum deviation of 8% for particles smaller than 5µm while it un-dersamples the dust with a maximum deviation of 10% when particles are larger than 10µm. The thoracic cutoff size of the sampler is 9.6µm in aerodynamic diameter, which is close to criterion value of 10µm. For respirable dust, the present data almost fall on the respirable curve with the cutoff aerodynamic diameter of 4µm. In addition, there are no significant differ-ences of sampling efficiencies between solid and liquid parti-cles. That is, the present sampler using PUFs as the impaction substrates matches both thoracic and respirable criteria simulta-neously very well.

FIG. 7. Comparison of the thoracic and respirable fractions between the ex-perimental data of the 3-stage sampler and the criteria.

Field Sampling

Before comparing particle concentrations between the Respi-con, three-stage sampler and the IOSH cyclone, two collocated identical samplers were compared first to show the precision of the tests. Figure 8a shows the comparison of the thoracic dust concentrations sampled by the two collocated Respicons. It is seen all data at the three workplaces are almost fall on the 1:1 line. Similar results are observed in Figure 8b which shows the respirable dust concentrations sampled by two collo-cated three-stage samplers. Concentrations of other dust frac-tions for any of two identical samplers show similar agree-ment. Furthermore, linear regression analysis of the data for all three workplaces showed no significant difference (p> 0.05) between the two collocated samplers for all three dusts. That is, the precision of the present field comparison study is very good.

Figure 9 shows the comparison of respirable dust concen-trations sampled by the IOSH cyclone, the present three-stage sampler and the Respicon. The three-stage sampler and IOSH cyclone collected similar respirable dust concentrations with less than 10 % difference, but they were 20 and 31% higher than the Respicon, respectively. For comparison, the study of Rando et al. (2005) showed the SKC cyclone had 19% lower respirable dust concentration of the Respicon.

The comparison of the thoracic dust concentrations between the present three-stage sampler and the Respicon is shown in Figure 10. It shows the three-stage sampler sampled lower tho-racic concentrations than the Respicon with a difference of about 22%. For comparison, the study of Rando et al. (2005) showed

(9)

FIG. 8. Comparison of two collocated samplers: (a) the thoracic dust concen-trations sampled by the Respicon, (b) the respirable dust concenconcen-trations sampled by the present 3-stage sampler.

the GK 2.69 cyclone sampled 32% lower thoracic dust than the Respicon.

While the differences between the reference samplers and the Respicon in Rando et al. (2005) need further explanation, the reason why the present sampler undersampled the thoracic dust by 22% compared to the Respicon maybe due to higher tho-racic sampling efficiency of the Respicon as shown in Figure 6 of Feather and Chen (2003) or Figure 4 of Koch et al. (1999). Fur-thermore, Figure 7 shows the current sampler having a slightly

FIG. 9. Comparison of the respirable dust concentration between the 3-stage sampler, the IOSH cyclone, and the Respicon.

sharper and lower thoracic sampling efficiency than the criterion curve may also lead to undersampling than Respicon.

The 20% oversampling of the respirable dust by the present sampler compared to that of the Respicon perhaps is due to slightly lower respirable sampling efficiency of the Respicon shown in Figure 4 of Koch et al. (1999).

FIG. 10. Comparison of the thoracic dust concentration between the 3-stage sampler and the Respicon.

(10)

94 C.-J. TSAI ET AL.

FIG. 11. Comparison of the inhalable dust concentration between the 3-stage sampler and the Respicon.

With and without applying the extrathoracic correction fac-tor of 1.5 for the Respicon, the comparison of the inhalable dust concentration between the three-stage sampler and the Respicon is shown in Figure 11. It is seen the three-stage sampler sampled slightly lower (within 5% difference) inhalable dust concen-tration than the uncorrected Respicon. However, the difference increased to about 24% compared to the corrected Respicon.

Although the present data show good agreement with the uncorrected Respicon for the inhalable dust under calm wind condition (<0.1 m/sec), Koch et al. (2002) found that an in-crease in sampling deficiency of the Respicon with increasing coarseness of the dust. It was suggested to correct the Respi-con’s extrathoracic dust by a factor of 1.8 to match with the CALTOOL reference sampler. Similarly, a correction factor of 1.5 for Respicon’s extrathoracic dust was also found to be neces-sary for accurate inhalable sampling in Tatum et al. (2002) using CIS inhalable sampler as a reference sampler, and Rando et al. (2005) using IOM as a reference sampler in field studies. That is, the present three-stage sampler may also undersample inhalable fraction due to similar sampling deficiency for large particles. Further improvement can be made in the future to design a better inlet which reduces this discrepancy in the inhalable efficiency.

CONCLUSIONS

This study designed and calibrated a novel three-stage sam-pler using monodisperse liquid and solid particles in the labo-ratory at a calm air chamber. The sampler used a single fixed sampling flow rate of 3.2 L/min. Results showed that using PUFs as the substrates to design the two impactors in this sampler, it is able to match with the respirable and thoracic sampling criteria

simultaneously. For the inhalable fraction, the present sampler showed a reasonable agreement with the inhalable convention, although an increase of sampling deficiency exists with increas-ing aerodynamic diameter for particles greater than 15µm. Sim-ilar inhalable, thoracic, and respirable fractions for liquid and solid particles indicated there was no solid particle bounce in the present sampler.

In field studies, two three-stage samplers, two IOSH cyclones, and two Respicons were mounted on a life-sized rotating man-nequin simultaneously to conduct the sampling in three factories. The three-stage sampler was found to sample similar respirable dust concentration as the IOSH cyclone while it oversampled the respirable dust than the Respicon. The three-stage sampler sampled lower thoracic concentrations than the Respicon, which maybe due to higher thoracic sampling efficiency of the Respi-con. For the inhalable dust, the present data showed good agree-ment with the uncorrected Respicon and the three-stage sampler. However, according to Feather and Chen (2003), the present three-stage sampler may also undersample inhalable fraction due to similar deficiency of the inhalable curve for large parti-cles. Further improvement can be made in the future to design a better inlet which reduces this discrepancy in the inhalable efficiency.

REFERENCES

Agarwal, J. K., and Liu, B. Y. H. (1980). A Criterion for Accurate Aerosol Sampling in Calm Air, Am. Ind. Hyg. Assoc. J. 41:191–197.

Aitken, R. J., and Donaldson, R. (1995). Development and Testing of Size-selective Personal Aerosol Sampling Devices based on Porous Foam Selec-tors, Abstract of Am. Ind. Hyg. Conf. Kansas, MO, May, 1995.

Dunkhorst, W., L¨odding, H., and Koch, W. (1995). A New Personal Mon-itor for Real-Time Measurement of the Inspirable, the Thoracic, and the Respirable Fraction of the Airborne Dust, J. Aerosol Sci. 26:S107– S108.

Feather, G. A., and Chen, B. T. (2003). Design and Use of a Setting Chamber for Sampler Evaluation under Calm-Air Conditions, Aerosol Sci. Technol. 37:261–270.

Glenny, R. W., Bernard, S., and Brinkley, M. (1993). Validation of Fluorescent-Labeled Microspheres for Measurement of Regional Organ Perfusion, J.

Ap-plied Physiol. 74:2585–2597.

Huang, C. H., Chang, C. S., Chang, S. H., Tsai, C. J., Shih, T. S., and Tang, D. T. (2005). Use of Porous Foam as a Substrate of an Impactor for Respirable Aerosol Sampling, J. Aerosol Sci. 36:1373–1386.

Kalatoor, S., Grinshpun, S. A., Willeke, K., and Baron, P. (1995). New Aerosol Sampler with Low Wind Sensitivity and Good Filter Collection Uniformity,

Atmos. Environ. 29:1105–1112.

Kavouras, I. G., and Koutrakis, P. (2001). Use of Polyurethane Foam as the Impaction Substrate/Collection Medium in Conventional Inertial Impactors,

Aerosol Sci. Technol. 34:46–56.

Kenny, L. C., and Gussman, R. A. (1997). Characterization and Modeling of a Family of Cyclone Aerosol Preseparators, J. Aerosol Sci. 28:677– 688.

Koch, W., Dunkhorst, W. L., and L¨odding, H. (1999). Design and Performance of a New Personal Aerosol Monitor, Aerosol Sci. Technol. 31:231–246. Koch, W., Dunkhorst, W., L¨odding, H., Thomassen, Y., Skaugset, N. P., Nikanov,

A., and Vincent, J. (2002). Evaluation of the Respicon as a Personal Inhalable Sampler in Industrial Environments, J. Environ. Monit. 4:657–662. Li, S. N., Lundgren, D. A., and Rovell-Rixx, D. (2000). Evaluation of Six

In-halable Aerosol Sampler, Am. Ind. Hyg. Assoc. J. 61:506–516.

(11)

Liu, B. Y. H., and Pui, D. Y. H. (1981). Aerosol Sampling Inlets and Inhalable Particles, Atmos. Environ. 15:589–600.

Mark, D., and Vincent, J. H. (1986). A New Personal Sampler for Airborne Total Dust in Workplaces, Ann. Occup. Hyg. 30:89–102.

Marple, V. A. (1978). Simulation of Respirable Penetration Characteristics by Inertial Impaction, J. Aerosol Sci. 9:125–134.

Maynard, A. D. (1999). Measurement of Aerosol Penetration through Six Per-sonal Thoracic Samplers under Calm Air Conditions, J. Aerosol Sci. 30:1227– 1242.

Rando, R., Poovey, H., Mokadam, D., Brisolara, J., and Glindmeyer, H. (2005). Field Performance of Industrial Wood Processing Dust, J. Occup. Environ.

Hyg. 2:219–226.

Seltzer, D. F., Bernaski, W. J., and Lynch, J. R. (1971). Evaluation of Size-Selective Presamplers II. Efficiency of 10-mm Nylon Cyclone, Am. Ind. Hyg.

Assoc. J. 32:441–446.

Tsai, C. J., Shih, T. S., and Sheu, R. N. (1997). Characteristics of Lead Aerosols in Different Work Environments, Am. Ind. Hyg. Assoc. J. 58:650–656. Tatum, V., Ray, A. E., and Rovell-Rixx, D. (2002). Performance of the RespiCon

(TM) Personal Aerosol Sampler in Forest Products Industry Workplaces, Am.

Ind. Hyg. Assoc. J. 63:311–316.

TSI (2001). Application Note ITI–078, http://www.tsi.com/documents/iti 078.pdf. TSI Incorporated, St. Paul, MN, USA.

Witschger, O., Fabries, J. F., and Gorner, P. (1997). Particle Entry Efficiency of an Annular Slot Aerosol Sampler, J. Aerosol Sci. 28:S679–S680.

數據

FIG. 1. Schematic diagram of the novel personal 3-stage dust sampler. (a) Top view of the inlet top disc
FIG. 2. Experimental setup for calibrating the 3-stage sampler.
Figure 4 shows the Inhalable fraction (or inlet efficiency) of the present three-stage sampler for particles from 0.8 to 27 µm
FIG. 3. Schematic diagram of the manikin and the location of samplers.
+4

參考文獻

相關文件

Ikeda, “Soft tissue chondroma of the hard palate: a case report,” Asian Journal of Oral and Maxillofacial Surgery, vol.. Merigo et al., “Soft tissue chondroma of the oral cavity:

Abu-Ghanem S, Yehuda M, Carmel NN, et al: Elective neck dissection vs observation in early-stage squamous cell carci- noma of the oral tongue with N0 clinically apparent lymph

13 According to Kim et al., 8 positive patch tests for dental materials were found in 70.5% of patients, mostly in oral lichen planus (75%), cheilitis (75%), and BMS (25%); the

Kristensen et al., “Prevalence of psoriatic arthritis in patients with psoriasis: a systematic review and meta-analysis of observational and clinical stud- ies, ” Journal of

Kurokawa et al., “Severe destruction of the temporomandibular joint with complete resorption of the condyle associated with synovitis, acne, pustulosis, hyper- ostosis, and

Jeejeebhoy FM, Zelop CM, Lipman S, et al; for the American Heart Association Emergency Cardiovascular Care Committee, Council on Cardiopulmonary, Critical Care, Perioperative

• label embedding: PLST, CPLST, FaIE, RAk EL, ECC-based [Tai et al., 2012; Chen et al., 2012; Lin et al., 2014; Tsoumakas et al., 2011; Ferng et al., 2013]. • cost-sensitivity: CFT,

Reading: Stankovic, et al., “Implications of Classical Scheduling Results for Real-Time Systems,” IEEE Computer, June 1995, pp.. Copyright: All rights reserved, Prof. Stankovic,