• 沒有找到結果。

Determination of mephenokalone in human plasma sample by high-performance liquid chromatography-fluroescence detection

N/A
N/A
Protected

Academic year: 2021

Share "Determination of mephenokalone in human plasma sample by high-performance liquid chromatography-fluroescence detection"

Copied!
7
0
0

加載中.... (立即查看全文)

全文

(1)

www.elsevier.com / locate / chromb

Determination of mephenoxalone in human plasma sample by

high-performance liquid chromatography–fluorescence detection

a,c a b c ,

*

Yow-Shieng Uang

, I.-Kei Chen , Li-Hsuan Wang , Kuang-Yang Hsu

a

Protech Pharmaservices Corporation, 6Fl., No. 380, Sung Chiang Road, Taipei 104, Taiwan b

Department of Pharmacy Service, Taipei Medical University Hospital, No. 252, Wu-Hsing Street, Taipei 110, Taiwan c

School of Pharmacy, Taipei Medical University, 250 Wu-Hsing Street, Taipei 110, Taiwan Received 28 November 2000; received in revised form 20 March 2001; accepted 4 April 2001

Abstract

A simple and sensitive high-performance liquid chromatographic method involving fluorescence detection was developed for the determination of mephenoxalone in human plasma. A Cosmosil 5C18-MS column (250 mm34.6 mm I.D., 5 mm) was used as stationary phase and the mobile phase consisted of water–acetic acid–acetonitrile (200:1:300) at a flow-rate of 1.0 ml / min. The fluorescence absorbance was monitored at 280 nm for excitation wavelength and 310 nm for emission wavelength. Temperature control was kept at 408C for the column. The limit of quantitation achieved was 10 ng / ml, and the standard curve was found to be linear in the concentration range of 10–10 000 ng / ml. Under these analytical conditions, a sufficient mephenoxalone plasma concentration profile could be obtained for pharmacokinetic study.  2001 Elsevier Science B.V. All rights reserved.

Keywords: Mephenoxalone

1. Introduction metric or fluorimetric analysis, measurement of

mephenoxalone substance by differential scanning

Mephenoxalone, as Fig. 1 shows, is a tranquilizer calorimetry, and measurement of the hydrolytic

pro-of the propanediol type. It has neuropharmacologic duct of mephenoxalone by a potentiometric method

activity in laboratory animals. Clinical studies indi-cate that mephenoxalone has a therapeutic effect on anxiety disorders and serves as a muscle relaxant in the treatment of muscle spasm in human subjects [1]. There are several assay methods for the determi-nation of mephenoxalone concentration in tablets. These methods involve direct measurement of mephenoxalone in solution by UV

spectrophoto-*Corresponding author. Tel.: 1886-2-2736-1661 ext. 684; fax: Fig. 1. Structures of mephenoxalone and internal standard

(es-1886-2-2737-4214. tradiol).

0378-4347 / 01 / $ – see front matter  2001 Elsevier Science B.V. All rights reserved. P I I : S 0 3 7 8 - 4 3 4 7 ( 0 1 ) 0 0 2 1 6 - X

(2)

[2–4]. None of these methods contain a separation filtered and degassed before use. The column

tem-procedure. Only the method described by Jancik et perature was set at 408C.

al. was applied to the determination of

mephenox-alone in biological samples [3]. For pharmacokinetic 2.3. Solution preparation

research, it is important to have both sufficient assay

specificity and sensitivity for mephenoxalone in A stock solution (1 mg / ml) of mephenoxalone

biological samples. However, these reported methods was prepared by dissolving 10 mg of mephenoxalone

are considered to have insufficient specificity and in 10 ml of acetonitrile–water (1:1, v / v) solution and

sensitivity for the determination of mephenoxalone stored at 2808C. Working solutions were also

pre-concentration in biological samples due to an inher- pared in the acetonitrile–water (1:1, v / v) solution at

ent interference by endogenous substances or metab- concentrations of 0.1, 1, 10 and 100 mg / ml and

olites. stored at 48C.

Therefore, the purpose of this study is to report a

simple, sensitive and reproducible high-performance 2.4. Sample preparation

liquid chromatography (HPLC) method for the

de-termination of mephenoxalone concentration in The sample preparation consisted of a single

human plasma. liquid–liquid extraction procedure. A 40-ml volume

of estradiol solution [2 ng / ml in acetonitrile–water (3:1, v / v) solution] as an internal standard was

2. Experimental added to 0.2 ml of plasma sample. A 3-ml volume of

hexane–dichloromethane mixing solution (7:3, v / v)

2.1. Chemicals and reagents as extraction solvent was added to the sample and

vortex-mixed for 30 s. After centrifugation for 10

Mephenoxalone and estradiol were obtained from min at 1945 g, the supernatant was transferred to a

Chemagis (Ramat-Hovav, Israel) and Sigma (St. clean tube and evaporated to dryness under a stream

Louis, MO, USA), respectively. HPLC-grade aceto- of nitrogen at ambient temperature. Then, 200 ml of

nitrile and analytical-grade acetic acid were pur- mobile phase was added and vortex-mixed for 30 s.

chased from BDH (Poole, UK) and E. Merck Finally, 20 ml was injected onto the HPLC system.

(Darmstadt, Germany), respectively. Hexane and

dichloromethane were GR grade and purchased from 2.5. Freeze and thaw stability

E. Merck. All other chemicals were analytically

graded and used without further purification. Freeze and thaw stability for mephenoxalone in

plasma samples was studied in three cycles with two

2.2. Apparatus and chromatographic conditions concentrations (40 and 8000 ng / ml) in six batch

plasmas. Samples were frozen at 2808C for at least

The HPLC system consisted of a Hitachi Model 2 h and then thawed to room temperature for another

L-7100 pump, a Model L-7200 autosampler, a Model at least 2 h.

L-7480 fluorescence detector, and a Model L-7000

HPLC system manager as data processor (Hitachi, 2.6. Standard curve

Japan), and a Model Super CO-150 column oven

(Enshine Scientific, Taiwan). Separation was A standard curve was prepared by adding 20 ml of

achieved on a Cosmosil 5C18-MS column (250 0.1 mg / ml, and 4, 10 and 20 ml of 1, 10 and 100

mm34.6 mm I.D., 5 mm; Nacalai Tesque, Japan). mg / ml of mephenoxalone working solution,

respec-The column eluent was monitored with a fluores- tively, to 0.2 ml of blank plasma and prepared

cence detector (280 nm for excitation wavelength according to the aforementioned sample preparation

and 310 nm for emission wavelength). The procedure. The concentrations used were 10, 20, 50,

mobile phase was water–acetic acid–acetonitrile 100, 200, 500, 1000, 2000, 5000 and 10 000 ng / ml.

(3)

was plotted against the concentration of mephenox- during the study. No strenuous activity was allowed.

alone added. Linearity was determined by weighted Water was limited from 1 h pre-dose and until 2 h

2

linear regression analysis (1 /x ). The concentration post-dose of each study phase. The subjects were

of mephenoxalone in the test samples was calculated required to fast for 12 h prior to each study session

using the regression parameters obtained from the until 4 h post-dose.

standard curve. After oral ingestion of one 200 mg of Dorsiflex

tablet (Syntex Pharm, Allschwil / Basel, Switzerland),

2.7. Accuracy and precision blood samples (10 ml each) were collected with

coded vacutainers (green top, containing heparin as

Four different concentrations of mephenoxalone anticoagulant) at the following times: 0 (pre-drug),

(10, 40, 600 and 8000 ng / ml) were added to drug- 0.33, 0.67, 1, 1.33, 1.67, 2, 2.5, 3, 4, 6, 9, 12, 14 and

free plasma and the concentrations were determined 23 h. The blood samples were centrifuged for 10 min

using the corresponding standard curves. The accuracy at 1945 g, immediately. The plasma for each sample

of the method was shown as relative error and was separated and aliquotted into polypropylene

calculated based on the difference between the mean tubes stored at 2808C in a freezer until analysis.

calculated and added concentrations, while precision was evaluated by calculating the within- and

be-tween-run relative standard deviations (RSDs). 3. Results and discussion

2.8. Recovery Mephenoxalone contains a 2-methoxyphenoxy

functional group in its molecule as shown in Fig. 1.

Triplicate human plasma samples at five con- This functional group makes it

fluorescence-absorb-centrations (10, 40, 600, 8000 and 10 000 ng / ml) ing. For analysis of biological samples, applying

were prepared. The internal standard solution was fluorescence detection would be a good choice for its

added to reconstitute the residues of plasma samples specificity. In the development of our analytical

after sample preparation. Solutions of mephenox- method, HPLC involving fluorescence detection was

alone were spiked into the internal standard solution, used. The monitoring wavelengths were set at 280

representing 100% recovery, at the same volume as nm for excitation and 310 nm for emission. Typical

those spiked into plasma samples and prepared in chromatograms are shown in Fig. 2. No significant

triplicate. After analyzing by HPLC, the absolute endogenous peak co-eluted with the mephenoxalone

recovery for mephenoxalone in human plasma after as shown in the corresponding chromatogram of

preparation procedure was determined by comparing drug-free plasma. In addition, the baseline is very

the mean peak height ratio of mephenoxalone in smooth and no interference peak is shown in the

human plasma (called plasma sample) to that of chromatogram except for the front peak. The

re-mephenoxalone in internal standard solution (called tention times for mephenoxalone and internal

stan-solvent sample). dard are 3.1 and 4.7 min, respectively. The analytical

time was only 8 min for a sample analysis.

2.9. Pharmacokinetic study The standard curve for mephenoxalone was made

using 10 spiking plasma samples over a

concen-Four healthy volunteers were selected to enroll in tration range of 10–10 000 ng / ml. In order to cover

the study. Complete physical examinations, blood the concentration range and to increase the sensitivity

chemistry, urinalysis and hematological evaluations, without sacrificing precision and accuracy, it was

and medical history were obtained from the subjects necessary and advisable to use weighted linear

prior to and after to the study, together with chest regression analysis. A linear relationship was

ob-X-rays and EKGs. Written informed consent were tained in the concentration range. Corresponding

2

also obtained from each prospective subject prior to correlation coefficients (r ) were over 0.99 from

the study. No smoking, alcohol and caffeine con- each of six different standard curves in one run

(4)

Fig. 2. Typical chromatograms of (A) drug-free plasma, and (B) a plasma sample taken 2 h after oral administration of mephenoxalone to a healthy volunteer. Peaks: 15mephenoxalone; 25internal standard.

separate runs (between-run). The regression parame- in plasma over a relatively wide range of

con-ters (n56) were y50.00100x20.00024 for within- centrations.

run and y50.00097x20.00060 ( y: peak height ratio; The criteria for the determination of the limit of

x: spiking concentration). The relative standard de- quantitation in plasma is based on signal-to-noise,

viations of the slope for six different standard curves the reproducibility of the response and the variability

were 0.9% for within-run and 4.4% for between-run. of the back-calculated concentration. Spiked plasma

The within-run precision and accuracy of standard samples with a final concentration of 10 ng / ml were

curve were evaluated with each of standard curve prepared and analyzed and were found to have a

concentration point after back-calculation. As Table signal-to-noise ratio of $5 for both within-run and

1 shows, RSDs ranged from 2.8 to 6.6% and relative between-run validation (n56). The mean peak height

error from 23.1 to 6.0%. The between-run precision ratio and standard deviation were 0.009453 and

and accuracy of standard curve were also evaluated 0.000771 for within-run validation, and 0.009045

with each standard curve concentration point after and 0.000581 for between-run validation,

respective-back-calculation. As Table 2 shows, RSDs ranged ly. The mean peak height ratio was greater than three

from 1.6 to 5.9% and relative error from 21.4 to standard deviations and the relative standard

devia-2.5%. These results indicate that the standard curve tions of the peak height ratio were only 8.2 and 6.4%

had a good linearity after weighted linear regression for within-run and between-run validations,

respec-analysis and was reproducible. It also shows that the tively. The relative standard deviation and relative

(5)

Table 1

Reproducibility of the standard curve (n56)

Spiking plasma Within-run Between-run

concentration

a a b

(ng / ml) Concentration RSD Relative Concentration RSD Relative error b

calculated (%) error calculated (%) (%)

(mean6SD) (%) (mean6SD) (ng / ml) (ng / ml) 10 9.6960.45 4.6 23.1 9.9560.31 3.1 20.5 20 21.261.4 6.6 6 20.561.2 5.9 2.5 50 50.461.8 3.6 0.8 49.861.3 2.6 20.4 100 97.763.8 3.9 22.3 99.062.5 2.5 21 200 199612 6 20.5 20366 3 1.5 500 498621 4.2 20.4 49368 1.6 21.4 1000 1004628 2.8 0.4 1006621 2.1 0.6 2000 2005674 3.7 0.3 2030656 2.8 1.5 5000 49646192 3.9 20.7 49546139 2.8 20.9 10 000 98826304 3.1 21.2 10 0396245 2.4 0.4 a RSD5100%3(SD/ mean). b

Relative error5100%3(concentration calculated2spiking plasma concentration) / spiking plasma concentration.

and 23.1% for within-run validation, and 3.1 and between 1.8 and 5.8%. The accuracy for within-run

20.5% for between-run validation, respectively, as and between-run was from 21.5 to 1.5% and from

Table 1 shows. These results indicate that the limit 20.6 to 3.0%, respectively, over the concentrations

of quantitation is 10 ng / ml and that the variability is examined. These results show that the method has

less than 20%, which shows that the limit of both good reproducibility and accuracy.

quantitation has acceptable accuracy, precision and To obtain good extraction efficiency and remove

reproducibility. interfering peak, a hexane–dichloromethane (7:3, v /

Within- and between-run accuracy and precision v) mixture was used as extraction solvent. The result

were examined by performing replicate analyses of indicated that no significant endogenous peak was

plasma samples (n56) to which four different known coeluted with the mephenoxalone as shown in the

concentrations of mephenoxalone had been added. corresponding chromatogram of drug-free plasma

As Table 2 shows, the within-run precision was (Fig. 2A). The absolute recovery for mephenoxalone

between 1.5 and 4.0% over the concentrations ex- in plasma after the extraction procedure was found to

amined. In addition, between-run precision was be 68.6%, as shown in Table 3. It indicates that the

Table 2

Precision and accuracy (analysis with spiking plasma samples at four different concentrations)

Spiking plasma Within-run Between-run

concentration

a a

(ng / ml) Concentration RSD Relative Concentration RSD Relative

b b

measured (%) error measured (%) error

(mean6SD) (n56) (%) (mean6SD) (n512) (%) (ng / ml) (ng / ml) 10 10.160.4 4.0 1.0 10.360.6 5.8 3.0 40 39.461.0 2.5 21.5 39.861.1 2.8 20.5 600 609613 2.1 1.5 601611 1.8 0.2 8000 81136119 1.5 1.4 79496141 1.8 20.6

(6)

Table 3 Recovery

a

Spiking plasma Peak height ratio (mean6SD) Recovery

concentration (%)

(ng / ml) Water sample Plasma sample

(n53) (n53) 10 0.0041260.00019 0.0027560.00002 66.9 40 0.0160560.00030 0.0111460.00021 69.4 600 0.2596260.00216 0.1801860.00119 69.4 8000 3.5167060.01455 2.4101860.01360 68.5 10 000 4.4797760.00735 3.0839060.03153 68.8 a

Recovery5100%3(peak height ratio of plasma sample / peak height ratio of water sample).

hexane–dichloromethane (7:3, v / v) mixture is a stability in plasma sample after three freeze and thaw

good medium for extraction of mephenoxalone from cycles.

plasma sample. The procedure was applied to a pharmacokinetic

The freeze and thaw stability study proposes a study in which mephenoxalone was orally

adminis-short-term stability of compound in plasma sample in tered to four healthy volunteers. Typical plasma

the thaw procedure. Furthermore, some plasma pro- concentration–time profiles are shown in Fig. 3. The

tein could coagulate after freezing and precipitating plasma concentrations of mephenoxalone were in the

after thaw. Owing to the fact that this compound standard curve range and remained above the 10

could bind to plasma protein, the phenomenon of ng / ml quantitation limit for the entire sampling

plasma protein coagulation in frozen condition and period. The pharmacokinetic parameters obtained

precipitation at room temperature after thaw would were described as follows. The value of area under

result in concentration loss of the compound. As the plasma concentration–time curve from time 0 to

Table 4 shows, the differences of mephenoxalone the last sampling time (AUC0 – t) was 15 94267096

concentrations between initial and each cycle were (ng / ml3h), and area under the plasma

concentra-less than 10% for both concentrations (40 and 8000 tion–time curve from time 0 to time infinite

ng / ml). In addition, there are no statistical signifi- (AUC0 – `) was 16 27467340 (ng / ml3h). The

ob-cant differences between cycles by two-way analysis served maximum plasma concentration (Cmax) was

of variance (ANOVA) analysis for both concentra- 28496535 (ng / ml), time to observed maximum

tions. It indicates that mephenoxalone has a good plasma concentration (Tmax) was 2.1760.58 h, and

elimination half-life was 2.7761.21 h. In addition, the plasma concentration–time profile of

mephenox-Table 4 alone for the entire sampling period can depict about

Freeze and thaw stability 90% of absorption according to the ratio of AUC 0 – t Cycle Concentration (ng / ml) divided by AUC0 – `.

These results demonstrate that this method is

40 8000

simple, sensitive, reproducible and accurate and

Initial 35.562.1 78826119

meets the requirement of the report of the conference

1 37.462.2 80846163 on Analytical Methods Validation: Bioavailability, a

(5.4%) (2.6%)

Bioequivalence and Pharmacokinetic studies [5]. As

2 38.062.2 80816115 shown by the data obtained during a pharmacokinetic (7.0%) (2.5%) study in which this particular method was applied, it 3 38.162.5 80146251 is concluded that the method described here offers

(7.3%) (1.7%) the opportunity to derive pharmacokinetic parameters a

(7)

Fig. 3. Plasma concentration–time profile of mephenoxalone after oral administration of Dorsiflex tablet to four healthy volunteers. Data are shown as mean6SD.

[4] J. Dohnal, Z. Volkova, K. Vytras, J. Pharm. Biomed. Anal. 7

References

(1989) 755.

[5] V.P. Shan, K.K. Midha, S. Dighe, I.J. McGilveray, J.P. [1] J.A. Morrison, Arch. Int. Pharmacodyn. Ther. 157 (1965) Skelly, A. Yacobi, T. Layloff, C.T. Viswanathan, C.E. Cook,

385. R.D. McDowall, K.A. Pittman, S. Spector, Pharm. Res. 9

[2] N. Erk, J. Pharm. Biomed. Anal. 21 (1999) 429. (1992) 588. [3] F. Jancik, J. Dohnal, B. Kakac, Z. Kollnerova, J. Pospisilova,

數據

Fig. 2. Typical chromatograms of (A) drug-free plasma, and (B) a plasma sample taken 2 h after oral administration of mephenoxalone to a healthy volunteer
Table 1 shows. These results indicate that the limit 20.6 to 3.0%, respectively, over the concentrations
Table 4 shows, the differences of mephenoxalone the last sampling time (AUC 0 – t ) was 15 94267096
Fig. 3. Plasma concentration–time profile of mephenoxalone after oral administration of Dorsiflex tablet to four healthy volunteers

參考文獻

相關文件

(In Section 7.5 we will be able to use Newton's Law of Cooling to find an equation for T as a function of time.) By measuring the slope of the tangent, estimate the rate of change

11[] If a and b are fixed numbers, find parametric equations for the curve that consists of all possible positions of the point P in the figure, using the angle (J as the

0 allow students sufficient time to gain confidence and the skills of studying in English, allow time for students to get through the language barrier, by going through

220V 50 Hz single phase A.C., variable stroke control, electrical components and cabling conformed to the latest B.S.S., earthing through 3 core supply cable.. and 2,300 r.p.m.,

3: Calculated ratio of dynamic structure factor S(k, ω) to static structure factor S(k) for "-Ge at T = 1250K for several values of k, plotted as a function of ω, calculated

A Boolean function described by an algebraic expression consists of binary variables, the constant 0 and 1, and the logic operation symbols.. For a given value of the binary

• Since the term structure has been upward sloping about 80% of the time, the theory would imply that investors have expected interest rates to rise 80% of the time.. • Riskless

Two cross pieces at bottom of the stand to make a firm base with stays fixed diagonally to posts. Sliding metal buckles for adjustment of height. Measures accumulated split times.