• 沒有找到結果。

Capillary zone electrophoresis /electrospray mass spectrometry of priority phenols

N/A
N/A
Protected

Academic year: 2021

Share "Capillary zone electrophoresis /electrospray mass spectrometry of priority phenols"

Copied!
7
0
0

加載中.... (立即查看全文)

全文

(1)

JOURNAL OF

CHROMATOGRAPHY A

ELSEVIER

Journal of Chromatography A, 743 (1996) 315-321

Capillary zone electrophoresis / electrospray mass spectrometry of

priority phenols

Cheng-Yuan Tsai, Guor-Rong Her*

Department of Chemistv, National Taiwan University, Taipei, Taiwan Received 21 September 1995; revised 5 March 1996; accepted 5 March 1996

Abstract

Eleven priority phenols were separated by capillary zone electrophoresis (CZE) and detected on-line with electrospray ionization mass spectrometry (ESI-MS) in the negative-ion mode. Parameters critical to the coupling of CZE and MS were studied. The best result was obtained with 2-[N-cyclohexylaminolethanesulfonic acid (CHES) as the running buffer and a solution of water-2-propanol (20:80) containing 0.5% ammonia as the sheath liquid. With the use of field amplification as the preconcentration technique and the mass spectrometer operated in the selected ion monitoring mode, the detection limit of most phenols was found to be in the range of 50 ppb. This method has been successfully applied to the determination of pentachlorophenol in a paper clay sample.

Keywords: Mass spectrometry; Buffer composition; Phenols; Priority phenols

1. Introduction

Many phenolic compounds are of great environ-

mental concern because of their toxicity. Eleven

phenols including phenol, 2-nitrophenol (2-NP), 4-

nitrophenol (4-NP), 2,4-dinitrophenol (2,4-DNP), 2-

chlorophenol (2-CP), 2,4-dichlorophenol (2,4-DCP),

2,4_dimethylphenol (2,4-DMP), 4-chloro-3-methyl-

phenol (4-C-3-MP), 2-methyl-4,6_dinitrophenol (2-

M-4,6-DNP), 2,4,6-trichlorophenol (2,4,6-TCP) and

pentachlorophenol (PCP) were listed by the United

States Environmental Protection Agency as priority

pollutants. Conventional methods based on gas chro-

matography (GC) [l] or high-performance liquid

chromatography (HPLC) [2-41 have been routinely

used for the monitoring of these pollutants. Recently,

0021-9673/96/$15.00 0 1996 Elsevier Science B.V. All rights reserved PII SOO21-9673(96)00195-l

*Corresponding author

mainly due to its superior separation efficiency,

several groups have explored the potentials of

analyzing phenols by capillary electrophoresis (CE)

[S-S].

The primary detection method of CE is UV

absorption [9-l 11, although, electrochemical and

fluorescence detection have also been reported [ 12- 151. The sensitivity of these techniques is, in general,

adequate for many applications, however, the in-

formation provided by these techniques is often not enough for a high confidence identification. Further- more, while the retention times are often used for

compound identification, the retention times of CE

are not as reproducible as in GC and HPLC and thus there is a greater risk in the identification of analytes based on retention time.

Because of its low detection limit, high specificity

(2)

316 C.-Y. Tsai, G.-R. Her I J. Chromatogr. A 743 (1996) 315-321

tion, mass spectrometry (MS) has been considered as

one of the ideal devices for chromatographic de-

tection. The merits of using MS as the chromato- graphic detector is best demonstrated with the highly

successful GC-MS. The coupling of HPLC and CE

with MS is, in general, much more difficult than the

interfacing of GC with MS, however, the recent

developments in MS such as electrospray ionization

(ESI) and atmospheric pressure chemical ionization

(APCI) makes the coupling of HPLC and CE with

MS [16-201 much easier.

CE is a modern technique known for its superior separation efficiency per unit time. The much smaller loading capacity and higher separation efficiency of CE make fraction collection more difficult compared to HPLC. In this report, we described the on-line

coupling of capillary zone electrophoresis (CZE)

with ESI-MS in the analysis of priority phenols.

2. Experimental 2.1. Chemicals

Phenol was obtain from Aldrich (Milwaukee, WI,

USA). 2,4-Dinitrophenol (2,4-DNP) was obtain from

Janssen (Belgium). 2-Nitrophenol (2-NP), 4-nitro-

phenol (4-NP), 2-chlorophenol (2-CP), 2,4-dichloro-

phenol (2,4-DCP), 2,4_dimethylphenol (2,4-DMP),

4-chloro-3-methylphenol (4-C-3-MP), 2-methyl-4,6-

dinitrophenol (2-M-4,6-DNP), 2,4,6-trichlorophenol

(2,4,6-TCP) and pentachlorophenol (PCP) were pur-

chased from Fluka (Switzerland). 2-[N-Cyclohexyl-

amino lethanesulfonic acid] (CHES) was purchased

from Sigma (St, Louis, MO, USA). Sodium tetra-

borate anhydrous and ammonium hydroxide were

purchased from Janssen (Belgium). Deionized (18

Ma) water (Milli-Q water system, Millipore, Bed-

ford, MA, USA) was used in the preparation of the

samples and buffer solution. 2-Propanol was HPLC

grade from J.T. Baker (Phillipsburg, NJ, USA). 2.2. Apparatus

The CZE system was made in-house using CZE

lOOOR (Spellman, Plainview, NY, USA) high-volt-

age power supply. CE columns were fused-silica

capillary (Polymicro Technologies, Phoenix, AZ,

USA) 50-/*m I.D., 375pm O.D. and 100 cm length

(90 cm to detector). A small area of the polyimide coating was burned off to form a window for UV detection. The detector (UV-C, Rainin, Emeryville,

CA, USA) wavelength was set at 210 nm. Before

use, the new capillary column was washed with 1 M

NaOH followed by water and running buffer. The

capillary was then equilibrated with the running

buffer under an electric field 200 V cm ’ for 20 min. To avoid absorption, the capillary column needs to be rinsed with sodium hydroxide between analysis. In normal operation, the sample was injected electro- kinetically at 10 kV for 10 s. The injection volume was calculated to be near 10 nl. In field amplification

for improving the sensitivity of this approach,

phenols prepared in water were introduced hydro-

dynamically into the column (15 cm; 7 min). High

voltage with reversed polarity (-20 kV) was then

applied to the column to push water out while

retaining the analytes on the column. When the

current reaches about 95% of the original value

(about 45 s) the polarity of the electrodes was

switched back to normal configuration (20 kV). A VG Platform single quadrupole mass spectrome-

ter (Fisons Instruments/VG Bio Tech, Altrincham,

UK) equipped with a CE interface was used for this study. The interface (Fig. 1) utilizes a triaxial flow arrangement whereby the CZE eluent is mixed with a suitable make-up solution (sheath liquid) at the probe tip and then nebulized using N, gas. In the negative- ion mode, a potential about -3.5 kV was applied to the probe tip. When the high voltage used for the

CZE separation was maintained at +20 kV, the

overall potential across the separation capillary was about 23.5 kV. A syringe pump (Model H-74900-00, Cole-Parmer, Niles, IL, USA) was used to deliver the sheath flow at a flow-rate of 10 pllmin. Nitrogen gas at a flow-rate of 0.5 l/min was used as the nebulizing gas. The warm (80°C) bath gas (nitrogen) at a flow-rate of 1.6 to 2.5 l/min aided dissolution of

the electrospray droplets. Mass spectral data were

collected at either full scan or selected-ion moni- toring (SIM) mode (0.2 s dwell time, 0.2 mass unit span).

3. Results and discussion

Both continuous flow fast atom bombardment

(3)

C.-Y. Tsai, G.-R. Her I J. Chromatogr. A 743 (1996) 315-321 317

Skimmrr Anne

System

Nebuliter Gas (N?) Fig. 1. The triaxial CZE-ES.1 interface

with CE [ 1 S-221. In comparison with FAB, ES1 has

the advantage of operating at atmospheric pressure

so that hydrodynamic flow due to the vacuum in the

separation capillary does not occur. Although CF-

FAB and ES1 can provide comparable sensitivity for different types of analytes, the coupling of CZE to

ES1 has gained considerable popularity in view of

the ease of operation and simplicity of the interface design.

CZE and micellar electrokinetic chromatography

(MEKC) are the two most popular CE techniques.

CZE is amenable to broad compound classes with

applications limited only by the necessity for solu- bility in the buffer and a non zero net electrophoretic mobility. Separation is based on the differential rate of migration of ionic species in an electric field [23-251. In contrast to CZE, MEKC is often used for

the separation of water insoluble and noncharged

compounds. Both CZE and MEKC have been suc-

cessfully used in the separation of priority phenols

[5-81. In this work, CZE instead of MEKC was

chosen as the separation technique because, most

likely due to the high concentration of surfactants, the sensitivity of CZE/ESI was much better than that of MEKC / ESI.

Sodium tetraborate at pH 9.9 has been used by

Whang et al. [8] as the running buffer in the CZE

analysis of phenols with indirect fluorescence de-

tection. Although the system produced a baseline

resolved electropherogram, borate buffer was not a

good choice for the CZE/ESI-MS operation. Both

sensitivity and ion current stability were not suitable

for the CZE/ESI-MS operation. Moreover, white

solid deposit was observed in the source due to the poor volatility of borate buffer. This solid deposit slowly blocked the orifice of the mass analyzer and thus reduced the sensitivity. A more volatile buffer, CHES, was then tested. The long term sensitivity and ion current stability are superior to borate buffer.

(4)

C.-Y. Tsai, G.-R. Her I .I. Chronuztogr. A 743 (1996) 315-321 318 (a) 6.1 13.3 20 26 min. (b) 10 20 min.

Fig. 2. Electropherograms of the eleven priority phenols. (a) sodium tetraborate buffer (pH 10.0). (b) CHES buffer (pH 10.0). Peaks were assigned as 1=2,4-DMP; 2=phenol; 3=4-C-3-MP; 4=PCP; 5=2,4,6-TCP; 6=2,4-DCP; 7=2-M-4,6-DNP; 8=2-CP; 9=2,4-DNP; lO= 4-NP; 11=2-NP.

resolution is not as good as with the borate as the running buffer (Fig. 2a), however, all eleven phenols are still baseline resolved (Fig. 2b). The elution order of the phenols was found to be the same as with the

25.000 20.000

E

‘3 15.000

z

‘;:

E

.s 10.000 E 0.000

borate buffer. The resolution and retention time may be affected by pH of the running buffer. Fig. 3 shows

the pH dependence of migration time for phenols

between pH 9 to pH 10.5 with 20 mk2 CHES buffer.

With pH above 9, all eleven phenols could be

separated with CZE. The migration times increased

as pH of the buffer increased. It is interesting to see

that when pH of the solution decreased to below

9.75, the elution order of 2-CP and 2-M-4,6-DNP

was reversed in comparison with higher pH.

Make-up solution (sheath liquid) performs two

functions, one is to supplement the CZE flow to a level suitable for ES1 operation and the other is to make electrical contact between the CZE eluent and probe tip. Sheath liquid was reported critical to the

performance of the CZE/ESI interface [19]. Sheath

liquids of different composition have been tested for

sensitivity and stability under negative ES1 con-

ditions. In the experiment, a 10 ppm 4-NP solution

was electroosmotically infused through the silica

capillary, and then mixed with sheath liquid (10

,ul/min) at the tip of the CE probe. The results

showed that the addition of ammonia improved the

sensitivity of phenols and 2-propanol produced a

more stable signal than methanol and ethanol (Fig. 4). The best result was obtained with a solution of

20:80 water-2-propanol containing 0.5% ammonia.

One major function of the sheath liquid is to

supplement the CZE flow to a level suitable for

electrospray operation. With the use of a sheath

9 9.5 9.75 10 10.5 pH Value

(5)

C.-Y. Tsai, G.-R. Her / J. Chromatogr. A 743 (1996) 315-321 319 l.Oe+j 8 6.0ei5 $ j.Oe+j 5 d < 4.0ec5 2 3.0e+s Z.Oe+j 1 Oe75 0 Oe+O

I)75125 2.propP"ol i wa,cr 2)80/20 2.propanal/water 3) 84116 2-propanol/water 4) SO/20 methrnol/warer+02% ammonia 5) 80/20 ahanolwater +a 2 % amnlO",l 6) 80/20 2-prapanol/water+ 0.2 % ammo",a

c 7) 8Ol20 2-propanoliwaw+O 4 % ammonll 8)80/20 2.propanol/water+O 5 % ammonia 9)SOlZO 2-propanollwawr*O6 % amm,,,,~

I 2 3 4 5 6 7 S 9

Sheath Liquid Composition

Fig. 4. Ion abundance as a function of ES1 sheath liquid composition, The 4-nitrophenol (10 ppm) was electroosmotically infused (20 kV) through the silica capillary.

liquid of 20:80 water-2-propanol containing 0.5%

ammonia, a flow of 8 to 12 ,ullmin was found to be best suited for ES1 operation.

Sheath liquid was chosen because of the optimi- zation in sensitivity and stability, therefore, sheath liquid would be an ideal choice as the running buffer in CZE. Unfortunately, good separation could not be

obtained with the sheath liquid (20180 water/2-

propanol containing 0.5% ammonia) as the running

buffer.

Sheath liquid was passed between the separation and the stainless capillary column (Fig. 1) and then mixed with the running buffer at the tip of the probe.

The relative position between these two columns

affected the sensitivity of the CZE/ESI interface.

The best sensitivity was obtained with the separation

column protruding slightly (about 0.2 mm) beyond

the tip of the stainless capillary column (Fig. 5). If the silica tubing protruded too far to make a good electrical contact, the sensitivity dropped very quick- ly.

Phenols showed very little fragmentation under

negative ES1 conditions. The negative ES1 mass

spectra of phenols are characterized with molecular

ion and background ions from buffer and solvent

(Fig. 6a and b). For the following reasons, the mass

spectrometer was operated at the SIM mode in the

CZE/MS analysis of low level phenols. First, the

sensitivity of SIM was shown to be about 10 times better than full scan. Secondly, background subtract-

L

L

Adjncenr Outside Outside (O.?mmi (O.Smm) Capillary Tip Posirmn

Inside (OSmm)

Fig. 5. Ion abundance as a function of capillary tip position. The 4-nitrophenol (10 ppm) was electroosmotically infused (20 kV) through the silica capillary. The flow-rate of sheath liquid [water- 2-propanol (20:80) containing 0.5% ammonia] was 10 pllmin.

ing routines, which were used to remove background

ions from the spectra of high concentration of

phenols (Fig. 6c and d), were not quite successful for

low levels (ppm-ppb) of phenols. The small vari-

ation in background ion current, not a serious

problem for high concentration samples, appeared to affect low level phenols greatly. Finally, as men- tioned earlier, there is little fragmentation in the

spectra, the lack of fragment ions makes little

difference between full scan and SIM.

The on-line CZE/MS analysis of a 10 ppm

mixture of phenols is given in Fig. 7. All phenols

gave single ion electropherograms with good peak

shape and signal-to-noise ratio. Although the same

amount of samples were injected, phenol, 2-CP, 2-

NP, and 4-C-3-MP appeared to have poorer SIN ratio (Fig. 7). Direct injection of phenols into the ES1 showed that the responses of phenol, 2-CP, 2-NP, and

4-C-3-MP were poorer than others.

The separation efficiency, represented by the

number of theoretical plates, N, was calculated from

the peak half-width for each phenol. For CZE-UV,

the range was from 50 900 for phenol to 182 000 for

(6)

320 C.-Y. Tsai, G.-R. Her 1 J. Chromatogr. A 743 (1996) 315-321

(4

206

Fig 6. Electrospray mass spectrum of (a) 4-C-3-MP and (b) 2,4,6-TCP obtained in the full scan mode under CZE-MS operation. The ions at m/z 141 and 195 are the [M-H]- ion of 4-C-3-MP and 2,4,6-TCP. The m/z 206 ion is the [M-H]- ion of CHES buffer. (c) and (d) were obtained after background subtrac- tion.

as good as in CZE-UV. The number of theoretical

plates drooped to 22 200 for phenol to 100 339 for

2-M-4,6-DNF’. The loss of separation efficiency is

most likely due to the dead volume of the CZE-MS

interface. The repeatability in migration time and

signal area were found to be 4% and 17% respective- ly.

The method was used to the determination of

priority phenols in a paper clay sample. The sample

was extracted with acidified methanol and then

analyzed directly by CZE-MS. Only one peak was

4-m 2-NP miz I38 m/z 138 1784 LL 18.30 2,4-DhQ m/z 183 17.15

n

2-cP m/z 127 16.23 h 2-MA,6-DNP m/z 197 “.“..J5 2,4-DCP m/z 161 15.72 I

1

2.4.6~TCP , m/z 195

4

15.41

I

1

PCP m/z 265 13.oB 1 1 4-c-3-M? ml2 141 phenol 2,4-DMP ml2 121 II.45 h 5.00 w.oa 1J.W 20.w l5.cQ”

Fig. 7. Selected ion electropherograms of the eleven priority phenols (IO ppm concentration). The peak at 15.41 min of the m/z

197 mass electropherogram is the [M+2] isotope peak of 2,4,6- TCP.

observed in the total ion current electropherogram (Fig. 8). This peak was assigned to pentachloro- phenol based on the molecular weight and isotopic ratio of the peak. After the identification, the com-

pound was quantified by CZE-MS and CZE-UV,

Our results showed that while the dynamic range of these two approaches were similar_(two orders), the

calibration curve from CZE-UV had a better linear

correlation than CZE-MS (r=0.999 vs. 0.984). The

concentration of pentachlorophenol in the paper clay

sample was determined to be 0.3 mg/g and 0.34

mg/g by CZE-UV and CZE-MS respectively.

The detection limit of this CZE-MS technique is

in the low ppm range. This sensitivity is not

(7)

C.-Y. Tsai, G.-R. Her I J. Chromatogr. A 743 (1996) 315-321 321

I

I bo 10’00 IJ’W 7.0 00

2J’OOfl

Fig. 8. Total and selected mass electropherograms of the paper clay sample.

samples without preconcentration. In order to im-

prove the sensitivity of this approach, the technique of field amplification was adopted. Phenols in water

were injected hydrodynamically (15 cm, 7 min) and

then stacked under reversed polarity (-20 kV) for 45

s. Except for 2,4-DMP, phenol and 4-C-3-MP, the

detection limit of this modified approach was found to be in the range of 50 ppb.

4. Conclusion

The combination of CZE and MS provided an

excellent analytical tool for the separation and

identification of priority phenols. This study showed that the position of the probe tip, the composition of running buffer and sheath liquid were all critical to

the performance of the CZE-ES1 interface. Although

it is possible to obtain useful mass electropherogram with 50 ppb level of most phenols, further improve- ment in sensitivity is needed for the detection of low ppb level of phenols.

This work was supported by National Research

Council of the Republic of China.

References

[7] C.P. Ong, C.L. Ng, N.C. Chong, H.K. Lee and S.F.Y. Li, J. Chromatogr., 516 (1990) 263.

Y.C. Chao and C.W. Whang, J. Chromatogr. A, 663 (1994) 229. 181 191 1101 1111 [I21 [I31 [I41 [ISI [I61 [I71 1181 [I91 PO1 Pll 1221 ~31 ~241 WI

F.E.P. Mikkers, F.M. Everaerts and Th.P.E.M. Verheggen, J. Chromatogr., 169 (1979) 11.

S. Terabe, K. Otsuka K. Ichikawa, A. Tsuchiya and T. Ando, Anal. Chem., 56 (1984) 111.

K. Ganzler, K.S. Greve, A.S. Cohen, B.L. Karger, A. Gunman and N.C. Cooke, Anal. Chem., 64 (1992) 2665. Y.E. Yik and S.F.Y. Li, Trends Anal. Chem., 11 (1992) 325. R.A. Wallingford and A.G. Ewing, Anal. Chem., 59 (1987)

1762.

S. Nie, R. Dadoo and R.N. Zare, Anal. Chem., 65 (1993) 3571.

E.S. Yeung, D. Wang, W. Li and R.W. Giese, J. Chromatogr., 608 (1992) 73.

G. Hopfgartner, T. Wachs, K. Bean and J. Henion, Anal. Chem., 65 (1993) 439.

S. Jun, U. Yoshihisa, K. Akihiro and K. Ikunoshin, Anal. Biochem., 207 (1992) 1.

R.D. Smith, J.A. Olivares, N.T. Nguyen and H.R. Udseth, Anal. Chem., 60 (1988) 426.

M.A. Moseley, J.W. Jorgenson, J. Shabanowitz, D.F. Hunt and K.B. Tomer, J. Am. Sot. Mass Spectrom., 3 (1992) 289. S. Pleasance, P. Thibault and J. Kelly, J. Chromatogr., 591 (1992) 325.

M.A. Moseley, L.J. Deterding, K.B. Tomer and J.W. Jorgen- son, Anal. Chem., 63 (1991) 109.

M.A. Moseley, L.J. Deterding, K.B. Tomer and J.W. Jorgen- son, J. Chromatogr., 516 (1990) 167.

H.H. Lauer and D. McManigill, Anal. Chem., 58 (1986) 166. J.W. Jorgenson and K.D. Lukacs, Anal. Chem., 53 (1981) 1298.

J.S. Green and J.W. Jorgenson, J. Chromatogr., 352 (1986) 337.

[l] R.A. Booth and J.N. Lester, J. Chromatogr. Sci., 32 (1994) 259.

[2] N.G. Buckman, J.O. Hill, R.J. Magee and M.J. McCormick, J. Chromatogr., 284 (1984) 441.

[3] F.P. Bigley and R.L. Grob, J. Chromatogr., 350 (1985) 407. [4] H.K. Lee, S.F.Y. Li and Y.H. Tay, J. Chromatogr., 438

(1988) 429.

[5] K. Otska, S. Terabe and T. Ando, J. Chromatogr., 348 (1985) 39.

[6] C.D. Gaitonde and PV. Pathak, J. Chromatogr., 514 (1990) 389.

數據

Fig.  3.  Effect  of  pH  on  elution  time  and  order.
Fig.  4.  Ion  abundance  as  a  function  of  ES1  sheath  liquid  composition,  The  4-nitrophenol  (10  ppm)  was  electroosmotically  infused  (20  kV)  through  the  silica  capillary
Fig.  7.  Selected  ion  electropherograms  of  the  eleven  priority  phenols  (IO  ppm  concentration)
Fig.  8.  Total  and  selected  mass  electropherograms  of  the  paper  clay  sample

參考文獻

相關文件

EQUIPAMENTO SOCIAL A CARGO DO INSTITUTO DE ACÇÃO SOCIAL, Nº DE UTENTES E PESSOAL SOCIAL SERVICE FACILITIES OF SOCIAL WELFARE BUREAU, NUMBER OF USERS AND STAFF. ᑇؾ N

Reading Task 6: Genre Structure and Language Features. • Now let’s look at how language features (e.g. sentence patterns) are connected to the structure

By correcting for the speed of individual test takers, it is possible to reveal systematic differences between the items in a test, which were modeled by item discrimination and

(a) The magnitude of the gravitational force exerted by the planet on an object of mass m at its surface is given by F = GmM / R 2 , where M is the mass of the planet and R is

• X-ray variability correlates with mass transfer rate: H/He disk stability model predicts various states. • X-ray bursts (active and

Mass-to-light ratios of old stellar populations are uncertain; (1) the initial mass function (IMF) of star formation is poorly known; it may vary with radius in an individual galaxy

To improve the convergence of difference methods, one way is selected difference-equations in such that their local truncation errors are O(h p ) for as large a value of p as

It is useful to augment the description of devices and services with annotations that are not captured in the UPnP Template Language. To a lesser extent, there is value in