• 沒有找到結果。

Fast and sensitive diagnosis of thalassemia by capillary Electrophoresis

N/A
N/A
Protected

Academic year: 2021

Share "Fast and sensitive diagnosis of thalassemia by capillary Electrophoresis"

Copied!
7
0
0

加載中.... (立即查看全文)

全文

(1)

O R I G I N A L P A P E R

Po-Ling Chang Æ I-Ting Kuo Æ Tai-Chia Chiu Huan-Tsung Chang

Fast and sensitive diagnosis of thalassemia by capillary electrophoresis

Received: 19 October 2003 / Revised: 15 March 2004 / Accepted: 31 March 2004 / Published online: 30 April 2004  Springer-Verlag 2004

Abstract This paper demonstrates the diagnosis of b-thalassemia by capillary electrophoresis in conjunction with laser-induced fluorescence using poly(ethylene oxide) (PEO) solutions in the presence of electroosmotic flow (EOF). During the electrophoretic separation, PEO solution entered a capillary from the anodic vial by EOF. The separation of a mixture of the polymerase chain reaction (PCR) products (330 and 334 base pairs) from a healthy person and a b-thalassemia patient was accomplished within 15 min at 15 kV using 1.5% PEO containing 2 M urea at 30C. The electropherogram patterns instead of migration times were used to diag-nose b-thalassemia, with an accuracy of 100% for the analyses of 11 blood samples from suspected patients. After injecting a large volume of the mixture to the capillary filled with 800 mM Tris-borate buffer (pH 10.0), the DNA fragments stacked due to increases in viscosity and sieving when migrating into 1.5% PEO solution. As a result of improved sensitivity, only 15 PCR cycles were required when using 500 ng of DNA templates. The results shown in this study indicate the potential of this simple, rapid, and cost-effective method for the diagnosis of b-thalassemia.

Keywords Capillary electrophoresis Æ DNA Æ Laser-induced fluorescence Æ Polymer solutions Æ Thalassemia

Abbreviations CE: Capillary electrophoresis Æ EOF: Electroosmotic flow Æ EtBr: Ethidium bromide Æ LIF: Laser-induced fluorescence Æ PCR: Polymerase chain reaction Æ PEO: Poly(ethylene oxide) Æ TRIS: Tris(hy-droxymethyl)aminomethane Æ TB: TRIS-borate

Introduction

Thalassemia results from quantitative reductions in glo-bin chain synthesis [1]. Those with decreased a-gloglo-bin chain production are called a-thalassemias, whereas those with diminished b-globin chains are termed b-tha-lassemias. The manifestations of thalassemias range from mild anemia with microcytosis to fatal serve anemia (Hb Barts hydrops fetails or b-thalassemia major). b-thalas-semia major presents in infancy and requires life-long transfusion therapy and/or bone marrow transplantation [2], and thus its diagnosis is of considerable importance. Slab gel electrophoresis has become one of the most popular tools for diagnosis of b-thalassemia [3, 4]. Compared to slab gel electrophoresis for DNA analysis, capillary electrophoresis (CE) provides the advantages of rapidity, use of minute amounts of samples, and easy automation [5, 6]. With such advantages, numerous CE techniques have been tested and validated for DNA sequencing, genotyping, mutation analysis, and forensic human identification [7–9]. The success of CE in DNA separation stems in part from the use of replaceable polymer solutions that provide the capability of self coating (dynamic coating) and single-base resolution [10–12]. Dynamic coating of polymer molecules on the capillary wall is a simple approach to minimizing the variation of electroosmotic flow (EOF) and DNA adsorption, leading to reproducibility and better resolution.

Temperature and electric field strength induce dif-ferential changes in the conformation (mobility) of various sizes and sequences of DNA, and thus temper-ature gradient and voltage programming in CE have shown powerful for the analyses of single-point muta-tions [13–15]. The applicamuta-tions of temperature gradient-CE and gradient-CE-using strongly acidic, isoelectric buffers to the analysis of point mutants [16] and screening b-39 mutations in thalassemia [17] have been demonstrated. Adding denaturant reagents such as urea and glycerol to the background electrolytes to improve resolving power

P.-L. Chang Æ I-T. Kuo Æ T.-C. Chiu Æ H.-T. Chang (&) Department of Chemistry, National Taiwan University, Section 4, Roosevelt Road, Taipei, Taiwan, R.O.C E-mail: changht@ntu.edu.tw

Tel.: +886-2-23621963 Fax: +886-2-23621963 DOI 10.1007/s00216-004-2627-9

(2)

of CE for DNA has also been recognized [18–20]. Recently, we have demonstrated DNA separation in the presence of EOF under discontinuous conditions [21– 23]. Polymer solutions enter the capillaries filled with Tris-borate (TB) buffer by EOF after injecting DNA samples. In the course of separation, DNA fragments possessing negatively charges enter PEO (neutral mole-cules) and they slow down as a result of sieving and an increase in viscosity. Since DNA migrates against EOF, large DNA fragments with small electrophoretic mobilities are detected earlier at the cathode end. In the presence of EOF, we have also developed on-line con-centration and separation techniques for the analysis of DNA [24–26]. Unlike other conventional on-line con-centration methods [27–32], stacking in our approaches is due to sieving and increases in viscosity when DNA migrating into PEO solution.

With the capability of single-base resolution and improved sensitivity (up to 450-fold), these recently developed on-line concentration techniques have estab-lished high potentials for clinical applications. In this study, one of the on-line concentration CE techniques was applied to the analyses of a mixture containing the polymerase chain reaction (PCR) products from a healthy person and a suspected thalassemia patient [24]. Our experimental goal also aimed at evaluating the effects of temperature and urea on resolution and speed for the diagnosis of thalassemia.

Materials and methods

Equipment

The basic design of the separation system has been pre-viously described [21]. Briefly, a high-voltage power sup-ply (Gamma High Voltage Research Inc., Ormond Beach, FL, USA) was used to drive electrophoresis. The entire CE system was enclosed in a black box with a high-voltage interlock. The high-voltage end of the separation system was put in a laboratory-made plexiglass box for safety. A 4.0-mW He-Ne laser with 543.6-nm output from Uni-phase (Mantence, CA, USA) was used for excitation. The light was collected with a 20X objective (numerical aperture=0.25). One RG 610 cutoff filter was used to block scattered light before the emitted light reached the phototube (Hamamatsu R928). The amplified currents were transferred directly through a 10-kW resistor to a 24-bit A/D interface at 10 Hz (Borwin, JMBS Develop-ments, Le Fontanil, France) and stored in a PC. Bare fused-silica capillaries (Polymicro Technologies, Phoenix, AZ, USA) with 75 lm i.d. and 365 lm o.d. were used for DNA separations without any further coating process.

Chemicals

All chemicals for preparing buffer and PEO (M. W.

WI, USA). Ethidium bromide (EtBr) that is cheap and weakly fluorescent was obtained from Molecular Probes (Eugene, OR, USA). TB buffers were prepared from Tris adjusted with boric acid to pH 9.0 and 10.0, respectively. In this manuscript, the molarity of TB buffer refers to that of Tris. Buffers of 800 mM TB (pH 10.0) and 200 mM TB (pH 9.0) were used to fill capillaries and prepare 1.5% PEO (with/without urea), respectively [21]. QIAamp DNA blood mini kit was purchased from QIAGEN (Hilden, Germany).

DNA extraction and PCR products

The blood samples were collected from patients that possibly suffer from thalassemia (lifelong transfusion) in Mackey Memorial Hospital. Human genomic DNA from buffy coat was extracted using QIAamp DNA blood mini kit according to the manufacturer’s instruc-tions. Amplification of the DNA samples was conducted as suggested by Chang’s group [33]. Briefly, a certain amount of human genomic DNA was mixed with 100 ng of the primer, 200 lM dNTP, and 2.5 U Taq polymer-ase in 100-ll PCR reaction buffer consisted of 10 mM Tris-HCl (pH 8.3), 50 mM KCl, 1.5 mM MgCl, and 0.01% gelatin. The settings for conducting PCR were: initial incubation at 94C for 10 min; cycling at 94 C for 1.1 min, at 55C for 1.1 min, and at 72 C for 2.1 min; and stored at 4C after the PCR cyclings (15 and 30) were finished. The times for 15 and 30-cycles PCR reactions were 90.0 and 170.0 min, respectively.

Stacking and separation

Prior to analyses, capillaries were treated with 0.5 M NaOH overnight. At least three times capillary-volume 800 mM TB (pH 10.0) was used to flush NaOH out of the capillary. DNA samples were prepared by mixing equal volumes of the PCR products either from a thy person and a b-thalassemia patient, from two heal-thy people, or from two b-thalassemia patients. Hydrodynamic injection (30-cm height; 12–360 s) was applied to inject DNA samples into the capillary filled with 800 mM TB (pH 10.0). During separation (and stacking), PEO entered the capillary by EOF and acted as sieving (and stacking if large-volume injection was applied) matrices. Owing to self-adsorption capability of PEO, EOF gradually decreased in the course of elec-trophoretic separation. For reproducibility, the capillary was equilibrated with 0.5 M NaOH at 1 kV for 10 min after each run.

To investigate the urea dependence, buffers consisting of 200 mM TB (pH 9.0) and urea (1–7 M) were used to prepare 1.5% PEO. After sample injection, the DNA fragments migrated into PEO solution, wherein they encountered urea that entered the capillary by EOF from the anodic end. For investigating the temperature

(3)

from the inlet to the point of 15 cm from the outlet, was surrounded by a teflon tube, wherein water flowed through at a rate of 5 mL/min. Using a temperature controller (YIH DER Instruments Co. Ltd., R.O.C.), the system temperature was stable (±1C) in the range of 20–80C.

Results and discussion

Analysis of PCR products

Owing to a 4-bp difference between the PCR products from a healthy person (334 bp) and a b-thalassemia patient (330 bp), a technique providing high-resolving power is essential for diagnosis of b-thalassemia. In a traditional method for b-thalassemia, enzymatic diges-tion of the PCR products is conducted prior to the separation by slab gel electrophoresis, in which only the PCR products from b-thalassemia patients can be di-gested [33]. However this method is slow and is difficult for automation, and thus it is not suitable for high-throughput diagnosis of b-thalassemia. With the advantages of simplicity, sensitivity, high-resolving power, and ease of automation [21–26], CE-LIF using PEO solutions should be suitable for fast diagnosis of b-thalassemia. To test our hypothesis, the CE-LIF technique was applied to the separations of three dif-ferent mixtures of the PCR products, which were from one healthy person and a suspected b-thalassemia patient, from two different healthy persons, and two suspected b-thalassemia patients. Electropherogram (a) in Fig. 1 presents two peaks corresponding to the PCR products from a healthy person and a b-thalassemia patient, respectively. We note that the first peak corre-sponds to the 334-bp fragment since DNA migrated

against EOF. Electropherograms (b) and (c) in Fig. 1 show only one peak when separating the mixtures of the PCR products from two healthy persons and two b-thalassemia patients, respectively. The relative standard deviation (RSD) values (n=3) of the migration times for the two peaks are 2.4 and 2.5%, respectively. We note that the time (t0in Table 1) at which the baseline shifted

can be used as an internal standard if the correction of the variation of the migration times for the 334- and 330-bp fragments is needed [21]. When PEO migrated through the detection window, the shift occurred as a result of the change in refractive index and a greater quantum yield of EtBr in PEO. Based on the results shown in electropherogram (b) and (c), the migration times for the PCR products could not be reliably used to tell the samples obtained from two healthy people or suspected b-thalassemia patients. It is also impossible to tell samples whether from healthy person or b-thalas-semia patients if we only separate the PCR product from a person’s blood, based on the migration time data. Fortunately, two partially resolved peaks were always detected for the mixtures containing the PCR products from a healthy person and a suspected patient although their differential migration time is so small. In other words, one should easily detect b-thalassemia from the electropherogram patterns regardless a small difference and variation of migration times for the two peaks around 16.0 min (more details see Sect. 3.3).

Effect of urea and temperature on DNA analysis

To further speed up the separation and optimize reso-lution, we conducted the separations at various tem-peratures and urea concentrations. DNA conformation and the viscosity of the background electrolyte are

Fluorescence Intensity

334 14 19 330 (b) (c) (a) 16.5 50 mV

Time (min)

Fig. 1a–c Electropherograms of separating three mixtures each containing two PCR products. Blood samples from a healthy person and a b-thalassemia patient in (a), two healthy persons in (b), and two b-thalassemia patients in (c). The PCR products of 5.0 ng/lL DNA templates were amplified by 30 cycles. Capillary: 50 cm in total length and 40 cm in effective length, filled with 800 mM TB, pH 10.0. PEO containing 5 lg/ml EtBr was prepared in 200 mM TB, pH 9.0. Hydrodynamic injections were conducted at 30-cm height for 60 s and separations were conducted at 25C and 15 kV in the presence of EOF using 1.5% PEO

(4)

temperature dependent, and thus the electrophoretic mobility of DNA should be different at various tem-peratures. Figure 2 presents the analysis was faster at higher temperature as a result of decreases in viscosity (increases in EOF). For example, in the absence of urea, the separation time and the bulk EOF mobility were 16.40 min and 1.84·10)4cm2V)1s)1 at 20C, while they were 11.51 min and 2.70·10)4cm2V)1s)1 at 60 C, respectively. Next, we turned our attention to exploring the effect of urea on DNA separation at constant temperature. Although our measurements shows that the effect of urea on the viscosity of 1.5% PEO solutions is negligible (less than 3.0% in the pres-ence of 0 and 7 M urea), which is in good agreement with a reported result [18], Fig. 2 does show the analysis time was slightly shorter in the presence of urea at a

constant temperature. This is mainly due to a suppres-sion of PEO adsorption on the capillary wall, presum-ably because hydrogen bonding between the hydroxyl groups in PEO and the Si-OH on the capillary surface reduced in the presence of urea [34–39]. Our reasoning is supported by the fact that the bulk EOF mobility values were 1.84·10)4 and 1.93·10)4cm2V)1s)1 in the ab-sence and preab-sence of 2 M urea at 20C. It is important to note that the reproducibility is better in the presence of urea, partially because of reduced adsorption of PEO (more constant EOF) on the capillary wall. For example, the RSD values of the migration time for the 330-bp fragment were 2.5 and 1.5% in the absence and presence of 2 M urea, respectively.

Temperature and urea also play some roles in deter-mining the fluorescence intensity and resolution for the

0 M 2 M 0 M 2 M 0 M 2 M 0 M 2 M 0 M 2 M 16.37 15.25 15.27 11.11 11.83 13.10 14.02 11.49 12.56 14.15 [Urea] Temp. 250 mV 20˚C Fluorescence Intensity 30˚C 40˚C 50˚C 60˚C

Fig. 2 Effects of temperature and urea on the separation of a mixture of the PCR products from a healthy person and a b-thalassemia patient. The migration times for the 334-bp fragment were shown in the electropherograms. Other conditions were the same as in Fig. 1

Table 1 Effects of urea and temperature on migration time and resolution of the two PCR products from a healthy person and a b-thalassemia patient

Temperature (C)

t0(min) Migration time (min) Resolutiona

0 M 1 M 2 M 0 M 1 M 2 M 0 M 1 M 2 M 20 12.06 (0.75)b 11.82 (0.62) 11.50 (0.56) 16.37c/16.40d 15.62/15.65 15.25/15.28 0.9 0.9 0.7 30 11.16 (0.58) 11.18 (0.55) 10.42 (0.50) 15.27/15.31 14.90/14.93 14.02/14.04 0.9 1.0 0.8 40 10.31 (0.59) 9.80 (0.49) 9.80 (0.46) 14.15/14.19 13.15/13.17 13.10/13.13 1.0 1.1 0.8 50 9.05 (0.49) 9.02 (0.48) 8.82 (0.45) 12.56/12.59 12.15/12.18 11.83/11.85 1.0 1.1 0.8 60 8.23 (0.52) 8.60 (0.49) 8.23 (0.46) 11.49/11.51 11.61/11.63 11.11/11.13 1.0 1.1 0.8 70 7.89 (0.67) 7.88 (0.57) 7.47 (0.54) 11.14/11.17 10.74/10.77 10.21/10.23 1.1 1.1 0.9 80 7.67 (0.71) 7.33 (0.58) 7.28 (0.50) 11.10/11.13 10.08/10.10 10.10/10.12 1.1 1.0 0.9 a

Rs=0.589 (t2- t1)/w1/2av; t1and t2are the migration times for the

334- and 330-bp fragments, respectively; w1/2avis the average width

at the half heights for the two peaks corresponding to the 334- and 330-bp fragments, respectively

b

RSD (n=3; %)

c334-bp fragment d330-bp fragment

(5)

two DNA fragments. At a constant urea concentration, the fluorescence intensities (peak heights) of the EtBr-DNA complexes slightly varied (less than 10%) in the range of 20–35C, while decreased with increasing temperature in the range of 35–80C. For example, the fluorescence intensities of the EtBr-DNA complexes at 80 C were about threefold lower than those at 30 C. Unlikely, the fluorescence monotonically decreased with increasing temperature from 20–80 C under static conditions (measured with a fluorometer), resulting from dynamic quenching. At high temperature, the changes in DNA conformation (denaturation) also contribute to the loss of fluorescence intensity because EtBr binds more strongly to ds DNA than to ss DNA

[40, 41]. A differential temperature dependence of fluo-rescence in CE from that in the static state suggested another mechanism involved in CE. In the absence of urea, narrower and more symmetric peak profiles were observed at high temperature, which is a result of weaker DNA adsorption and small diffusion (shorter migration times). Thus we suggest that quenching, stability of intercalated DNA, and peak profiles affect the sensitivity of the two DNA fragments in CE. We note that the resolving power for the two peaks increases with increasing temperature in the range of 30–70C (Table 1) as a result of narrower peak profiles. The RSD values (n=3) for the resolution were less than 6.0%. Overall, we concluded that the separation conducted at 30–40C using solutions consisting of 1.5% PEO and 1–2 M urea is proper. Hereafter, the separations were conducted at 30C in the presence of 2 M urea. Using this condition, the resolution for the two DNA frag-ments was 0.8, with a RSD of 3.6% in five consecutive runs.

Diagnosis of b-thalassemia

Figures 1 and 2 suggests that the electropherogram pattern is useful for diagnosing b-thalassemia. One peak and two partially resolved peaks around 14.0 min shown in the electropherograms (30C; 2 M urea) indicate that the patients do not and do suffer from b-thalassemia, respectively. To further validate this method for the diagnosis of b-thalassemia, we collected bloods from one healthy person and eleven patients who may suffer from b-thalassemia. The electropherogram patterns of sepa-rating 11 mixtures each consisting of the PCR products from the healthy person and one of the suspected

Table 2 Diagnosis of b-thalassemia by CE in the presence of EOF using 1.5% PEO containing 2 M urea at 30C

Sample Total peaks Codon 41-42 ()4 nt) frame shift This work Traditional methodb

H1 + P1a 2 Positive Positive H1 + P2 2 Positive Positive H1 + P3 2 Positive Positive H1 + P4 1 Negative Negative H1 + P5 2 Positive Positive H1 + P6 2 Positive Positive H1 + P7 2 Positive Positive H1 + P8 2 Positive Positive H1 + P9 1 Negative Negative H1 + P10 2 Positive Positive H1 + P11 1 Negative Negative a

H1 means the PCR product was from a healthy person and P1– P11 mean the PCR products were from patients 1 to 11, respec-tively bRef. [33] Fluorescence Intensity 1 mV 17.16 17.18 Time (min)

Fig. 3 Separations of a large-volume mixture of the 15-cycle PCR products from a healthy person and a b-thalassemia patient using 1.5% PEO containing 2 M urea at 30C. The concentrations of the DNA templates were 5.0 ng/ll. Hydrodynamic injection was conducted at 30-cm height for 180 s. Other conditions were the same as in Fig. 1

(6)

patients were used to read the diagnostic results. The sensitivity of this method tabulated in Table 2 is 100% when compared to those diagnosed by a traditional method [33], indicating the potential of this method for fast diagnosis of b-thalassemia. Unlike most traditional methods, the variation of migration time does not affect our diagnostic results because the reading relies on the electropherogram patterns in our method. Without conducting enzyme digestion prior to separation at an electric field of 375 V cm, this new method is simple, low costly, and fast.

On-line concentration

The total analysis time (PCR, injection, and separa-tion time) increases with increasing PCR cycles, and thus low PCR cycles are favorable for fast diagnosis of b-thalassemia. However, a highly sensitive detection method is required for small amounts of PCR prod-ucts. For the past years, our efforts have made to develop on-line CE concentration and separation techniques for the analyses of trace amounts of DNA [24–26]. These techniques allow the analyses of DNA prepared in high- and low-conductivity media without tedious sample pretreatment. Together with the result shown in Table 2, we believed that our on-line CE concentration and separation techniques should be highly suitable for fast diagnosis of b-thalassemia. To demonstrate the beauty of our method, a large volume (ca. 132 nl) of the mixture of the PCR products from a healthy person and b-thalassemia patient after 15-cycle amplification was injected. Figure 3 presents the separation was accomplished within 17.2 min, with two well separated peaks. The RSD values (n=3) of the migration times for the two peaks were 3.8%. Due to relatively stronger PEO adsorption at low ionic strengths, the separation time was slightly longer when compared to that shown in Fig. 2 (30C; 2 M urea). It is important to point out that the total analysis time was shortened from 184.3 min (30 cycles) to 110.2 min (15 cycles) as shown in Table 3, revealing the potential of this method for diagnosing b-thalassemia from a great number of blood samples, especially when using a capillary array electrophoresis (CAE) system [42, 43]. When conducting DNA separation in a CAE system, the migration times for the DNA fragments

mainly due to the differential capillary properties, the variation of the electric field strength in different capillaries, and so on. To correct the variation, an internal standard is usually added to the sample. By applying our method, an internal standard is not re-quired since the diagnosis is according to the electro-pherogram patterns, not migration times, and the shift in the baseline is useful to correct the migration times. Thus our suggested method should hold great poten-tial for high-throughput diagnosis of b-thalassemia using a CAE system.

Concluding remarks

We have progressed in the investigation of diagnosis of b-thalassemia by CE in the presence of EOF using 1.5% PEO. From the electropherogram pattern of separating a mixture of the PCR products from a healthy person and a suspected b-thalassemia patient, one can easily and accurately diagnose whether the patient suffers from b-thalassemia. Together with applying the on-line CE concentration technique, we have demonstrated the analysis of a mixtures of 15-cycle PCR products from a healthy person and a b-thalassemia patient. Because electropherogram patterns instead of migration times are used to diagnose b-thalassemia, the variation of migration times if obtained in different runs is not problematic. With the advantages of sensitivity, rapid-ity, simplicrapid-ity, and low costs, the proposed method has great potential for screening b-thalassemia from a large number of samples when using a CAE system.

Acknowledgements This work was supported by the National Sci-ence Council of the Republic of China under contract number NSC 92-2113-M002-048.

References

1. Clarke GM, Higgins TN (2000) Clin Chem 46:1284–1290 2. Chern SR, Chen CP (2000) Int J Gynecol Obstet 69:103–106 3. Guillemette C, Millikan RC, Newman B, Housman D (2000)

J Cancer Res 60:950–956

4. Huang CS, Chang PF, Huang MJ, Chen ES, Chen WC (2002) Gastroenterology 123:127–133

5. Heller C (2001) Electrophoresis 22:629–643 6. Issaq HJ (2000) Electrophoresis 21:1921–1939 7. Hu S, Dovichi NJ (2002) Anal Chem 74:2833–2850

8. Tian H, Brody LC, Mao D, Lander JP (2000) Anal Chem Table 3 Summary of the analyses of PCR products amplified after 30 and 15 cycles

Methoda Template (ng) Injection volume (nl) Time (min)b Advantages PCR Injection Separation Total

A 500 7.5 170.0 0.2 14.1 184.3 Simple

B 500 132 90.0 3.0 17.2 110.2 Fast

a

Conditions for methods A and B were as in Fig. 2 (30C; 2 M urea) and Fig. 3, respectively

b

Total analysis time was about 320 min using a conventional method (Ref. [33])

(7)

9. Hahn M, Wilhelm J, Pingoud A (2001) Electrophoresis 22:2691–2700

10. Zhou H, Miller AW, Sosic Z, Buchholz B, Barron AE, Kotler L, Karger BL (2000) Anal Chem 72:1045–1052

11. Horvath J, Dolnı´k V (2001) Electrophoresis 22:644–655 12. Xue G, Pang HM, Yeung ES (2001) J Chromatogr A 914:245–

256

13. Tian H, Brody LC, Mao D, Landers JP (2000) Anal Chem 72:5483–5492

14. Yeung ES, Gao Q (2000) Anal Chem 72:2499–2506

15. Tian H, Brody LC, Fan S, Huang Z, Landers JP (2001) Clin Chem 47:173–185

16. Gelfi C, Righetti PG, Travi M, Fattore S (1997) Electrophoresis 18:724–731

17. Gelfi C, Vigano´ A, Carta P, Manchia P, Cossu GF, Righetti PG (2000) Electrophoresis 21:780–784

18. Otim O (2001) Biopolymers 58:329–334

19. Liang D, Song L, Chen Z, Chu B (2001) J Chromatogr A 931:163–173

20. Liang D, Song L, Quesada MA, Tian Z, Studier FW, Chu B (2000) Electrophoresis 21:3600–3608

21. Chen HS, Chang HT (1999) Anal Chem 71:2033–2036 22. Huang MF, Hsu CE, Tseng WL, Lin YC, Chang HT (2001)

Electrophoresis 22:2281–2290

23. Hsieh MM, Chang PL, Chang HT (2002) Electrophoresis 23:2388–2393

24. Huang CC, Hsieh MM, Chiu TC, Chang HT (2001) Electro-phoresis 22:4328–4332

25. Tseng WL, Hsieh MM, Wang SJ, Huang CC, Lin YC, Chang PL, Chang HT (2001) J Chromatogr A 927:179–190

26. Kuo IT, Chiu TC, Chang HT (2003) Electrophoresis 24:3339– 3347

27. Chien RL, Helmer JC (1991) Anal Chem 63:1354–1361 28. Auriola S, Ja¨a¨skela¨inen I, Regina M, Urtti A (1996) Anal

Chem 68:3907–3911

29. Guttman A, Szo¨ko¨ E´ (1996) J Chromatogr A 744:321-324 30. Barme´ I, Bruin GJM, Paulus A, Ehrat M (1998)

Electropho-resis 19:1445–1451

31. Beckers JL, Bocˇek P (2000) Electrophoresis 21:2747–2767 32. Xiong Y, Park SR, Swerdlow H (1998) Anal Chem 70:3605–

3611

33. Chang JG, Chen PH, Chiou SS, Lee LS, Perng LI, Liu TC (1992) Blood 80:2092–2096

34. Mathur S, Moudgil BM (1997) J Colloid Interface Sci 196:92– 98

35. Rubio J, Kitchener JA (1976) J Colloid Interface Sci 57:132– 141

36. Kaupp S, Bubert H, Baur L, Nelson G, Wa¨tzig H (2000) J Chromatogr A 894:73–77

37. Tseng WL, Hsieh MM, Wang SJ, Chang HT (2000) J Chro-matogr A 894:219–230

38. Zhang Z, Zhu Y, Shi Y (2001) Biophys Chem 89:145–162 39. Nooren IMA, Rietveld AWM, Melacini G, Sauer RT, Kaptein

R, Boelens R (1999) Biochemistry 38:6035–6042

40. Nock T, Dove J, McCord B, Mao D (2001) Electrophoresis 22:755–762

41. Schell J, Wulfert M, Riesner D (1999) Electrophoresis 20:2864– 2869

42. Gao Q, Yeung ES (2000) Anal Chem 72:2499–2506

43. Shortreed MR, Li H, Huang WH, Yeung ES (2000) 72:4640– 4645

數據

Table 1 Effects of urea and temperature on migration time and resolution of the two PCR products from a healthy person and a b-thalassemia patient
Table 2 Diagnosis of b-thalassemia by CE in the presence of EOF using 1.5% PEO containing 2 M urea at 30 C

參考文獻

相關文件

por dissolução ou tratamento químico de polímeros orgânicos naturais (por exemplo, celulose), para produzir polímeros tais como raiom cuproamónio (cupro) ou raiom viscose, ou

The calculation of derivatives of complicated functions involving products, quotients, or powers can often be simplified by taking logarithms. The method used in the next example

Normalization by the number of reads in the sample, or by calculating a Z score, should be performed on the reported read counts before comparisons among samples. For genes with

 Promote project learning, mathematical modeling, and problem-based learning to strengthen the ability to integrate and apply knowledge and skills, and make. calculated

Now, nearly all of the current flows through wire S since it has a much lower resistance than the light bulb. The light bulb does not glow because the current flowing through it

In view of the large quantity of information that can be obtained on the Internet and from the social media, while teachers need to develop skills in selecting suitable

Wang, Solving pseudomonotone variational inequalities and pseudocon- vex optimization problems using the projection neural network, IEEE Transactions on Neural Networks 17

Define instead the imaginary.. potential, magnetic field, lattice…) Dirac-BdG Hamiltonian:. with small, and matrix