Proceedings - 19th International Conference

-

IEEE/EMBS Oct. 30

-

Nov. 2, 1997 Chicago, IL. USA

DESIGN AND IMPLEMENTATION

OF MICROSENSOR SYSTEM

Chii-Wann Lid, Shao-Cheng Wang', Oscal T.-C. Chen3, Sandy Wang3

'Center for Biomed Eng, Coll. Med., Natl. Taiwan Univ., Taipei, Taiwan, R.O.C. 'EE Dept., Natl. Taiwan Univ., Taipei, Taiwan, R.O.C.

'EE Dept., Natl. Chung Cheng Univ., Chia-Yi, Taiwan, R.O.C. 'E-mail: cwlin@cbme.mc.ntu.edu.tw

Abstract

The design, implementation and characterization of a microsensor system for electrochemical measurements are introduced. This microsensor system consists of an electrochemical cell, amplifier, switches for miniature applications. Currently, we use a single ship microprocessor (80C5 I), an ADC (ADC0804) and seven-segment display for portable instrument prototype. The proposed microsensor is implemented by using the UMC 0.51 CMOS technology with a die size of 1.0 mm x 1 .O mm. The electrode area is 600 U x 800 U. The unique design of a feedback voltage controlled resistor allows a flexible dynamic range for linear conversion of the induced current in various applications.

Introduction

The effective cost of a microsensor totally depends on its employed mass-production technologies. As the available methods, screen printing [l], thin film [2] and silicon technologies [3] have all been used in the past with some successful examples in the real world [4-61. Together they provide numerous novel devices for better performance and faster tum-around time in clinical diagnostic and therapies [7]. Even though there are still quality control issues to be addressed with standard Si-technology processes, it attracts a lot of focus for the possible biomedical applications ranging from physical sensors to chemical ones [4,8,9]. The potential small size, low cost, high precision, and good reproducibility can significantly contribute to the quality of health care with smart design for better performance [ 10,l I]. The same technologies can also meet the increasing demands in food industries, environmental monitoring and process control [12]. In this paper, we report an amperometric microsensor for possible medial applications. The flexible design of electrodes is explored with the amplifying and compensating circuitry. The unique architecture of this microsensor enables two-electrode or three-electrode configuration with a pass-transistor control and provides a flexible dynamic range adjustment for the induced current by a voltage-control resistor. The integrated device of microelectrode, current-to-voltage amplifier, voltage-control resistor, operational amplifier, and pass-transistor is designed, simulated and fabricated as the front-end smart device. The analog output is then digitized by an analog-to-digital converter (ADC0804) which interfaces to a single chip

controller (8051) for post processing and data reporting via seven segment display.

Methods

The designed microsensor consists of an electrochemical cell, a current-to-voltage amplifier, a voltage-controlled resistor, an operational amplifier, and four pass-transistor switches as shown in figure 1.

The induced current of amperometric method can be derived according to the following equation,

I, = n*F*A*D,*(d[O]/dx) n : valence number F : Faraday constant A :Area

Do : Diffusion coefficient [O]: Oxygen concentration where

With the constrains of total area on Si-substrate, the scale-down electrode area is about 0.48 mm2 which will induce 0-300

nA

current with glucose range from 0 to 12.5 mM.

The reference and workingkounter electrodes are implemented by using Ag and Pt, respectively. They are realized by using photolithography and vacuum etching on hletal-2 layer of standard CMOS technologies. The circuitry of a current-to-voltage amplifier, a voltage-control resistor, an operational amplifier, and four pass-transistor switches is also implemented on the same substrate to fabricate an integrated microsensor as the front-end device for measurement system.

The analog output of this front-end device is then fed into an ADC (ADC0804) for 8 bits digital conversion. The output data is interfaced to a low power consumption single chip controller (80C5 1) for further processes which include storage, calibration, and report. Three seven-segment displays are used to report the data for range from 0 to 5V with two decimal precision,

Results

The partial cross section of this microelectrode is shown in figure 2. A folded-cascade operational amplifier with a negative feedback voltage-controlled resistor is realized for the conversion of the induced currents to a reliable analog voltage output. At a control voltage of lV, the HSPICE

2267 (O-7803-4262-3/97/$10.00 (C) 1997 IEEE)

(2)

Proceedings

-

19th International Conference

- IEEE/EMBS Oct. 30

-

Nov. 2, 1997 Chicago, IL. USA

circuits simulation results in linear conversion of input current with sensitivity of 6.4 mVlnA as shown in figure 3. Its equivalent impedance is 5.8 M Ohm. The feedback voltage-controlled resistor has an impedance increase rate of 2 M O h " of its control voltage. The operation of two or three electrodes can be controlled through external registration. The proposed microsensor as shown in figure 4 is implemented by using the UMC 0.5 U CMOS technology. The total electrode area is 600 U x 800 U for the deposition of enzyme layer after its partial package and return to the laboratory. With the designed specifications, the induced current range can be effectively measured based on the two-electrode operation at a control voltage of 1 V. The maximum output voltage for blood glucose detection is about I .8 V. The single-chip controller with three seven-segment display circuit is shown as in figure 5 .

Discussions

A microsensor system is designed and realized by using 0.5 uM standard CMOS technology and a single chip controller for post-processing of data. There are a few technical issues which can be addressed for further development. First of all, post-modification of the microsensor with existing cleanroom technologies needs to put in more efforts to improve the yield rate. Secondly, self-calibration procedure is important for the performance of microsensor system and needs to integrate as a part of the hardware or software. Thirdly, the post data processing module needs to implement with low power consumption consideration to increase the durability of the microsensor system.

Conclusions

A unique architecture of microsensor is proposed and fabricated which enables two-electrode or three-electrode configuration with a pass-transistor control and provides a flexible dynamic range adjustment for the induced current by a voltage-control resistor. This integrated microsensor is designed, simulated and fabricated as the front-end smart device for biomedical applications using amperometric method.

Acknowledgments Council, Taiwan, R.O.C.

This work is supported in part by National Science

References

[ l ] R. Nagata, K. Yokoyama, S . A. Clark and I. Karube, "A glucose sensor fabricated by the screen printing technique." Biosensors & Bioelectronics 10,261-267 1995.

[2] 0. J. Prohaska, F. Olcaytug, F. Kohl, G. Urban, W. Chu, R. Vollmer and J. LaManna, Medical Applications of Miniaturized Chamber-type Electrochemical Sensors. In Microsensors and Catheter-Based Imaging Technology (SPIE

Proceedings v. 904). (Ed. A. I. West) pp. 13-16, SPIE, Bellingham, WA 1988.

[3] 0. T.-C. Chen, S. Wang, C. -W. Lin and Y. -C. Lu, A medical microsensor for blood glucose monitoring. ISCAS'97 Hong Kong (1 996). (Abstract)

[4] E. Kress-Rogers, Handbook of biosensors and electronic noses: medicine, food and the environment, CRC Press, 1997. [5] H. M. McConnell, J. C. Owicki, J. W. Parce,

D.

L. Miller, G. T. Baxter, H. G. Wada and S. Pitchford, "The cytosensor microphysiometer: biological applications of silicon technology." Sci. 257, 1906-1912 1992.

[6] F. Scheller and F. Schubert, Biosensors, Elsevier Science Publisher, Amsterdam 1992.

[7] J. H. Nichols, C. Howard, K. Loman, C. Miller,

D.

Nyberg and D. W. Chan, "Laboratory and bedside evaluation of portable glucose meters." Am J Clin Pathol 103, 244-251

1995.

[8] M. Adami, M. Martini and L. Piras, "Characterization and enzymatic application of a redox potential biosensor based on a silicon transducer." Biosensors & Bioelectronics [9] C. Nicolini, M. Lanzi, P. Accossato, A. Fanigliulo, F. Mattioli and A. Martelli, "A silicon-based biosensor for real-time toxicity testing in normal versus cancer liver cells." Biosensers & Bioelectronics 10, 723-733 1995.

[IO] U. Fischer, S. Alcock and A. P. F. Tumer, "Assessment of devices for in vivo monitoring of chemical ' species.:'

Biosensors & Bioelectronics 10, xxiii-xiv 1995.

[ l l ] I. Moser, G. Jobst, E. Aschauer, P. Svasek, M. Varahram, G. Urban, V. A. Zanin, G. Y. Tjoutrina, A. V. Zharikova and T. T. Berezov, "Miniaturized thin firm glutamate and glutamine biosensors." Biosensors &

Bioelectronics 10, 527-532 1995.

[12] R. Hintsche, M. Paeschke, U. Wollenberger, U. Schnakenberg, B. Wagner and T. Lisec, "Microelectrode arrays and application to biosensing devices." Biosensors &

Bioelectronics 9, 697-705 1994. 10,633-638 1995.

Figure 1. Schematic diagram of the proposed microsensor

(3)

Proceedings

-

19th International Conference

-

IEEE/EMBS Oct. 30

- Nov. 2, 1997 Chicago, IL.

USA

1

I

_._.-.. - 1

.. .

__

- . -

. ..

___

Figure 2. Partial cross section view of electrode area

Figure 3. The HSPICE simulation result of the current-to-voltage amplifier.

Figure 5. Schematic diagram of single-chip controller and ADC for microsensor applications.

Figure 4. The top-view of the microsensor system layout.

數據

Figure  1. Schematic diagram of the proposed microsensor
Figure  3.  The  HSPICE  simulation  result  of  the  current-to-voltage amplifier.

參考文獻

相關文件

• Tactics: the art of organizing an army, and using weapons or military units in combination against the enemy in military encounters.. • Operational art: a component of military

/** Class invariant: A Person always has a date of birth, and if the Person has a date of death, then the date of death is equal to or later than the date of birth. To be

Type case as pattern matching on values Type safe dynamic value (existential types).. How can we

• Information retrieval : Implementing and Evaluating Search Engines, by Stefan Büttcher, Charles L.A.

Variable gain amplifier use the same way to tend to logarithm and exponential by second-order Taylor’s polynomial.. The circuit is designed by 6 MOS and 2 circuit structure

We will design a simple and effective iterative algorithm to implement “The parallel Tower of Hanoi problem”, and prove its moving is an optimal

This design the quadrature voltage-controlled oscillator and measure center frequency, output power, phase noise and output waveform, these four parameters. In four parameters

The applications of an ERP cloud service to improve the effective management of a SCC and the related possible operational issues are then discussed1. Finally, a real case of an